Test Name:
Method:
Department:
Prepared by:
Date:
Method Performance Characteristic Summary
Accuracy/Method Comparison
Verify or establish the accuracy of the method.
Number of samples:_____________________
Reference method:______________________
Sample type:__________________________
Other information:______________________
Include a copy of regression statistics from a statistical
evaluation program.
Achieved performance - include summarized data for
reference and comparison methods. Brief description of
number of samples, sample type, how samples were
chosen, brief description of comparative method/name
of comparative method. This should include regression
statistics. Use of reference materials with known
concentrations is suggested in establishing or verifying
accuracy.
Analytical Precision
Verify or establish the precison of the method
Number of samples:____________________
Concentrations tested:__________________
Sample type:_________________________
Include a copy of statistics including mean, SD, and CV
from a statistical evaluation program.
Achieved performance - include summarized pertinent
data for all levels. Analytic precision can be established
by making measurements on a series of aliquots of the
same test sample within a specified period of time.
There should be a minimum of 10 data points for each
precision check. Precision should be checked at 2 or 3
concentrations whenever possible. Use concentrations
as close as possible to the medical decision level for
the analyte.
Analytic Sensitivity
Verify or establish the analytic sensitivity of the method.
Include a copy of the calculated mean and SD
statistics.
Include a brief description of how the data was
obtained. The method should be sensitive enough to
detect differences in the level of the analyte, especially
at the decision point level. Run material of known lowest
concentration such as linearity material to determine
analytic sensitivity at that range.
Method Validation Plan
Instructions:
Summarize the method performance characteristics in the table below. Attach all supporting documentation to
this plan. The number of samples and sample type is determined by the clinical laboratory and approved by the
medical director prior to initiation of the validation.
Master Series Webinar. © CAP 2013
Please note this is a sample form only.
Use is not required and will not guarantee that your facility is compliant.
VAL 002
49886-10
Interferences
Verify or establish interferences for the method. Include
brief description of the manufacturer information
reviewed and substances tested. Include data results
for interferences if tested.
Brief description of how data was obtained. May include
manufacturer information. Analytical system should be
evaluated for cross reactivity or interferences by
compounds known to be similar in chemical structure,
or known to be interfering substances. This
investigation should also focus on substances that are
unique to the population being served.
Reportable Range/AMR
Determine the analytical measurement range for the
method. Verify or establish AMR for the method.
Include copy of statistical evaluation from a statistical
evalaution program.
Achieved performance - include summarized results for
samples tested. Materials of appropriate matrix with
known target values must be used. In order for a test of
linearity, five analyte concentrations must be utilized.
Replicate samples must be run. (Minimum duplicate
analysis.)
Reference Intervals
Reporting of results with reference intervals.
Achieved performance - summary of data obtained.
Include a brief description of how data was obtained.
This may include the number of subjects, criteria for
inclusion, composition of patient pool, number of
assays, etc. The minimum number of values is 20; 100
values are recommended.
Other Method Performance Characteristics
Describe: Summary of assessment:
(Specimen stability studies) (description here)
(Specimen type comparisons)
(Carryover studies)
Include documentation of evaluation or manufacturers
literature summary review.
This validation study has been reviewed and the performance of the method is
considered acceptable for patient testing.
Laboratory Medical Director Approval:_____________________________________
Date:______________________________
Master Series Webinar. © CAP 2013
Please note this is a sample form only.
Use is not required and will not guarantee that your facility is compliant.
VAL 002
49886-10