BiPAP AVAPS
USER MANUAL
REF 1105933
1105957 R01
JJB 01/01/2013
Table of Contents
i
Chapter 1. Introduction....................................................................................................................................... 1
Package Contents ........................................................................................................................................... 1
Intended Use ....................................................................................................................................................2
Warnings and Cautions .................................................................................................................................2
Warnings ...................................................................................................................................................2
Cautions ....................................................................................................................................................5
Notes ..........................................................................................................................................................6
Contraindications ............................................................................................................................................6
Patient Precautions .........................................................................................................................................7
System Overview .............................................................................................................................................7
Control Buttons ......................................................................................................................................9
Available Therapy Modes ..................................................................................................................10
Available Therapy Features...............................................................................................................11
Symbols ............................................................................................................................................................12
How to Contact Philips Respironics ........................................................................................................13
Chapter 2. Device Setup ....................................................................................................................................15
Installing the Air Filter .................................................................................................................................15
Where to Place the Device .........................................................................................................................15
Connecting the Breathing Circuit ...........................................................................................................16
Supplying AC Power to the Device .........................................................................................................17
Display Symbols ...................................................................................................................................18
BiPAP AVAPS
user manual
Table of Contents
© 2013 Koninklijke Philips Electronics N.V. All rights reserved.
BiPAP AVAPS user manual
ii
Navigating the Device Screens ................................................................................................................21
Starting and Stopping the Device .................................................................................................21
Monitor Pressure Screen ...................................................................................................................22
Changing the Comfort Settings......................................................................................................23
Setup Screen ..........................................................................................................................................25
Viewing the Information Screen .....................................................................................................27
Viewing the Monitor Parameter Screen .......................................................................................28
Viewing Measured Parameters .......................................................................................................28
Chapter 3. Device Alarms ..................................................................................................................................29
Audible and Visual Alarm Indicators ......................................................................................................29
Alarm LED Indicators ..........................................................................................................................30
Alert Audible Indicators .....................................................................................................................30
Silencing an Alarm ...............................................................................................................................31
Alarm Message Screens .....................................................................................................................31
Alarm Summary Table .................................................................................................................................32
Troubleshooting ............................................................................................................................................37
Chapter 4. Accessories ........................................................................................................................................41
Adding a Humidier with or without Heated Tubing ......................................................................41
SD Card .............................................................................................................................................................41
Supplemental Oxygen ................................................................................................................................41
Shielded DC Cord ..........................................................................................................................................42
Carrying Case ..................................................................................................................................................42
Airline Travel ..........................................................................................................................................42
Chapter 5. Cleaning the Device .......................................................................................................................43
Cleaning or Replacing the Filters.............................................................................................................43
Cleaning the Tubing .....................................................................................................................................44
Ser vice ...............................................................................................................................................................44
Chapter 6. Specications ...................................................................................................................................45
Chapter 7. EMC Information .............................................................................................................................49
Limited Warranty ...................................................................................................................................................53
Chapter 1 Introduction
1
This chapter provides an overview of the device.
Package Contents
The BiPAP AVAPS system may include the following components. Some components
(e.g., humidier) are optional accessories that may not be packaged with the device.
Carrying Case
User and
Provider Manuals
Power Cord
AC Power Adapter
Flexible Tubing (1.8 m x 22 mm)
(optional 15 mm tubing is also available)
Secure Digital
(SD) Card
Ventilator
Humidier
Reusable Gray
Foam Filter
BiPAP AVAPS
user manual
1. Introduction
Package Contents
BiPAP AVAPS user manual
2
Intended Use
The BiPAP AVAPS device is intended to provide non-invasive ventilatory support to treat adult patients
weighing over 66 lbs (30 kg) and pediatric patients 7 years or older and weighing over 40 lbs (18 kg)
with Obstructive Sleep Apnea (OSA) and Respiratory Insuciency. This device may be used in the
hospital or home.
Warnings and Cautions
Caution: US federal law restricts this device to sale by or on the order of a physician.
Warnings
A warning indicates the possibility of injury to the user or operator.
Device Usage This device is not intended for life support. The device provides Positive
Pressure Ventilation and is indicated for assisted ventilation. The device
does not provide ventilation with guaranteed V
T
delivery. Patients requiring
ventilation at a predetermined V
T
are not candidates for Pressure Support
ventilation.
Personnel
Qualications
This manual serves as a reference. The instructions in this manual are not
intended to supersede your health care professional’s instructions regarding
the use of the device.
The prescription and other device settings should only be changed on the
order of the supervising physician.
The operator should read and understand this entire manual before using
the device.
Patient
Circuits
The device should be used only with masks and connectors recommended
by Philips Respironics or with those recommended by the health care
professional or respiratory therapist. A mask should not be used unless the
device is turned on and operating properly. The exhalation port(s) associated
with the mask should never be blocked.
Explanation of Warning: The device is intended to be used with special
masks or connectors that have exhalation ports to allow continuous ow of
air out of the mask. When the device is turned on and functioning properly,
new air from the device ushes the exhaled air out through the mask
exhalation port. However, when the device is not operating, enough fresh air
will not be provided through the mask, and exhaled air may be rebreathed.
Chapter 1 Introduction
3
Patient
Circuits
In the event of a power or device failure, audible and visual alarm signals will
activate. The device must be disconnected from the patient immediately. As
is the case with most ventilators with passive exhalation ports, when power
is lost, sucient air will not be provided through the circuit, and exhaled air
may be rebreathed.
At low EPAP pressures, the ow through the exhalation port may be inadequate
to clear all exhaled gas from the tubing. Some rebreathing may occur.
The device does not have an alarm to detect occlusion of the exhalation
port. Before each use, inspect the patient circuit to verify that the port is
not occluded. Occlusion or partial occlusion can reduce airow and result in
rebreathing of exhaled air.
Verify the operation of the Patient Disconnect alarm with any changes in the
patient circuit.
Oxygen When using oxygen with this system, the oxygen supply must comply with
local regulations for medical oxygen.
When using oxygen with this system, turn the device on before turning
on the oxygen. Turn the oxygen o before turning the device o. This will
prevent oxygen accumulation in the device. Explanation of the Warning:
When the device is not in operation and the oxygen ow is left on, oxygen
delivered into the tubing may accumulate within the device’s enclosure.
Oxygen accumulated in the device enclosure will create a risk of re.
When using oxygen with this system, a Philips Respironics Pressure Valve
(REF 302418) must be placed at the air outlet port. Failure to use the pressure
valve could result in a re hazard. Refer to the pressure valve instructions for
use for proper use.
Supplemental oxygen cannot be used with the heated tube accessory. The safety
pressure valve is not compatible with this set-up, and could result in a re hazard.
If administering xed-ow supplemental oxygen, the oxygen concentration may
not be constant. The inspired oxygen concentration will vary depending on the
pressure setting, patient breathing pattern, and leak rate. Substantial leaks around
the mask may reduce the inspired oxygen concentration to less than the expected
concentrations. Appropriate patient monitoring should be implemented.
Oxygen supports combustion. Oxygen should not be used while smoking or
in the presence of an open ame.
Do not connect the device to an unregulated or high pressure oxygen source.
Do not use the device near a source of toxic or harmful vapors.
BiPAP AVAPS user manual
4
Operating
Temperatures
Do not use this device if the room temperature is warmer than 95˚ F (35˚ C). If
the device is used at room temperatures warmer than 95˚ F, the temperature
of the airow may exceed 109˚ F (43˚ C). This could cause irritation or injury
to your airway.
Do not operate the device in direct sunlight or near a heating appliance
because these conditions can increase the temperature of the air coming out
of the device.
Bacteria Filter If the device is used by multiple persons (such as rental devices), Philips
Respironics recommends that a low-resistance, main ow bacteria lter (Part
Number 342077) be installed in-line between the device and the circuit
tubing to prevent device contamination.
Improperly
Functioning
Ventilator
If you notice any unexplained changes in the performance of the device, if it
is making unusual sounds, if it has been dropped or mishandled, if water is
spilled into the enclosure, or if the enclosure is cracked or broken, disconnect
the power cord and discontinue use. Contact your home care provider.
The use of accessories, transducers and cables other than those specied,
with the exception of transducers and cables sold by Philips Respironics
as replacement parts for internal components, may result in increased
Emissions or decreased Immunity.
This device should not be used adjacent to or stacked with other equipment
and that if adjacent or stacked use is necessary, the device should be observed
to verify normal operation in the conguration in which it will be used.
Operation of the device may be adversely aected by:
Electromagnetic elds exceeding the level of 3 V/m in the test
conditions of EN 60601-1-2
Operation of high frequency (diathermy) equipment
Debrillators, or short wave therapy equipment
Radiation (e.g., x-ray, CT scan)
Magnetic elds (e.g., MRI)
Mobile RF communication equipment
The use of accessories, transducers and cables other than
those specied, with the exception of transducers and cables
sold by Philips Respironics
Power Cord Be sure to route the power cord to the outlet in a way that will prevent the
cord from being tripped over or interfered with by chairs or other furniture.
This device is activated when the power cord is connected.
Chapter 1 Introduction
5
Maintenance Never operate the device if any of the parts are damaged or if it is not
working properly. Have any damaged parts replaced before continuing use.
Electrical cords, cables, and the power supply device should be periodically
inspected for damage or signs of wear. Replace any damaged parts before using.
Repairs and adjustments must be performed by Philips Respironics-
authorized service personnel only. Unauthorized service could cause injury,
invalidate the warranty, or result in costly device damage.
Cleaning To avoid electric shock, unplug the device before cleaning it.
Do not immerse the device in any uids or spray the device with water or
cleaners. Clean the device with a cloth dampened with an approved cleaner.
Humidier For safe operation, the humidier must always be positioned below the
breathing circuit connection at the mask and the air outlet on the device.
The humidier must be level for proper operation.
Cautions
A caution indicates the possibility of damage to the device.
Electrostatic
Discharge
(ESD)
Pins of connectors should not be touched. Connections should not be
made to these connectors unless ESD precautionary procedures are
used. Precautionary procedures include methods to prevent build-up of
electrostatic charge (e.g., air conditioning, humidication, conductive oor
coverings, non-synthetic clothing), discharging ones body to the frame of
the equipment or system or to earth or a large metal object, and bonding
oneself by means of a wrist strap to the equipment or system or to earth.
Before operating the device, ensure that the SD card cover is replaced
whenever any of the accessories such as the Link Module or modem are not
installed. Refer to the instructions that came with your accessory.
Do not use antistatic or conductive hoses or conductive patient tubing with
the device.
EMC
Information
All Medical Electrical Equipment needs special precautions regarding
EMC and needs to be installed and put into service according to the EMC
information provided in Chapter 7: EMC Information.
Condensation Condensation may damage the device. If the device has been exposed
to either very hot or very cold temperatures, allow it to adjust to room
temperature (operating temperature) before starting therapy.
Do not operate the device outside of the operating temperature range
shown in the Specications chapter.
BiPAP AVAPS user manual
6
Extension
Cords
Do not use extension cords with this device.
Device
Placement
Do not place the device in or on any container that can collect or hold water.
Do not place the device directly onto carpet, fabric, or other ammable
materials.
Do not plug the device into an outlet controlled by a wall switch.
Air Filter A properly installed, undamaged reusable foam inlet lter is required for
proper operation.
Operating the device with a dirty lter may keep the system from working
properly and may damage the device.
A dirty inlet lter may cause high operating temperatures that may aect
device performance. Regularly examine the inlet lter as needed for
integrity and cleanliness.
Never install a wet lter into the device. You must ensure sucient drying
time for the cleaned lter.
Cleaning Do not immerse the device in liquid or allow any liquid to enter the
enclosure or inlet lter.
Notes
t
Additional warnings, cautions and notes are located throughout this manual.
t Please see the “Limited Warranty section of this manual for information on warranty coverage.
Contraindications
The device is contraindicated on patients without a spontaneous respiratory drive. If any of the
following conditions apply to you, consult your physician before using the device:
t Inability to maintain a patent airway or adequately clear secretions
t At risk for aspiration of gastric contents
t Diagnosed with acute sinusitis or otitis media
t Allergy or hypersensitivity to the mask materials where the risk from allergic reaction
outweighs the benet of ventilatory assistance
t Epistaxis, causing pulmonary aspiration of blood
t Hypotension
When assessing the relative risks and benets of using this equipment, the health care professional
should understand that this device can deliver the pressure ranges indicated in the Control
Accuracy table in chapter 6. In the event of certain fault conditions, a maximum pressure of 40 cm
H
2
O is possible.
Chapter 1 Introduction
7
Patient Precautions
t Immediately report any unusual chest discomfort, shortness of breath, or severe
headache.
t If skin irritation or breakdown develops from the use of the mask, refer to the mask
instructions for appropriate action.
t The following are potential side eects of noninvasive positive pressure therapy:
Ear discomfort
Conjunctivitis
Skin abrasions due to noninvasive interfaces
Gastric distention (aerophagia)
System Overview
The BiPAP AVAPS device is intended to augment patient breathing
by supplying pressurized air through a patient circuit. It senses the
patient’s breathing eort by monitoring airow in the patient circuit
and adjusts its output to assist in inhalation and exhalation. This
therapy is known as Bi-level ventilation. Bi-level ventilation provides a
higher pressure, known as IPAP (Inspiratory Positive Airway Pressure),
when you inhale, and a lower pressure, known as EPAP (Expiratory
Positive Airway Pressure), when you exhale. The higher pressure
makes it easier for you to inhale, and the lower pressure makes it
easier for you to exhale. The device can also provide a single pressure
level, known as CPAP (Continuous Positive Airway Pressure).
When prescribed, the device can also provide features to help make your therapy more
comfortable. The ramp function allows you to lower the pressure when trying to fall asleep. The
air pressure will gradually increase until the prescription pressure is reached. Additionally, the Flex
comfort feature provides increased pressure relief during the expiratory phase of breathing, and the
AVAPS feature helps you maintain a target V
T
.
Several accessories are also available for use with the device. Contact your home care provider to
purchase any accessories not included with your system.
BiPAP AVAPS user manual
8
The gure below illustrates some of the device features, described in the table below.
Air Outlet Port
Power Inlet
Filter Area
SD Card (Accessory) Slot
Side Cover
Ta b
SD Card Cover
Feature Description
Air Outlet Port Connect the exible tubing here.
Note: Heated Tubing should only be connected to the Air Outlet Port of the
compatible System One Heated Humidier and not to the Air Outlet Port of
the therapy device.
SD Card
(Accessory) Slot
If applicable, insert the optional SD card here.
SD Card Cover If applicable, the optional accessories such as a Link Module or Modem can
be installed here. Refer to the instructions supplied with your accessory.
When not using an accessory, this cover must be in place on the device.
Power Inlet Connect the power supply cord here.
Filter Area A reusable, gray foam lter must be placed in the lter area to screen out
normal household dust and pollen.
Side Cover If using a humidier with the device, this side cover can be easily removed
with the release tab before attaching the humidier. Refer to the Humidier
Manual for more information.
Chapter 1 Introduction
9
Control Buttons
The gure below shows the display screen and primary control buttons on the device.
Humidier Icon
and Number Settings
Alarm Silence/
Indicator Button
LCD Display Screen
Control Wheel/
Push Button
Ramp Button
Feature Description
Display Screen Shows therapy settings, patient data, and other messages. The startup
screen is shown temporarily when the device is rst powered.
Humidier Icon This Icon lights up (dierent colors) when the optional humidier and/
or heated tube is attached and heat is being applied. White means classic
humidication is selected. Blue means System One humidication is
selected. Orange means the heated tube is attached. Please refer to the
humidier user manual for more information.
Humidier
Numbers
The humidier number settings are only visible when the humidier is
attached and therapy is active. You can use the control wheel to change
the number settings for the humidier. When the heated tube is being
used with the humidier, these numbers will control the heated tube
temperature setting.
Control Wheel/
Push Button
Turn the Wheel to toggle between options on the screen. Press the Wheel
to choose an option. Primary function is to turn airow on/o. Pressing the
Wheel also resets alarms.
Ramp Button When the airow is on, this button allows you to activate or restart the ramp
function. This button lights up when therapy is active or during specic alerts.
Alarm Silence/
Indicator Button
Silences the audible portion of the alarm for a period of time and indicates
an alarm condition.
BiPAP AVAPS user manual
10
Available Therapy Modes
The table below describes the therapy modes available on the device:
Therapy Modes Description
CPAP Continuous Positive Airway Pressure; CPAP maintains a constant level of
pressure throughout the breathing cycle.
S Spontaneous Pressure Support; A Bi-level therapy mode where breaths are
patient-triggered and patient-cycled. The device triggers to IPAP (Inspiratory
Positive Airway Pressure) in response to spontaneous inspiratory eort and
cycles to EPAP (Expiratory Positive Airway Pressure) during exhalation. The
device also cycles a patient-triggered breath if no patient exhalation eort is
detected for 3 seconds. The level of Pressure Support delivered is determined
by the dierence between the IPAP and EPAP settings (PS = IPAP - EPAP)
S/T Spontaneous/Timed Pressure Support; A Bi-level therapy mode where each
breath is patient-triggered and patient-cycled or machine-triggered and
machine-cycled. S/T mode is similar to S mode, except that the device also
triggers machine-triggered breaths based on a set breath rate and cycles
machine-cycled breaths based on a set inspiratory time if the patient does
not spontaneously breathe within a set time.
T Timed Pressure Support; A Bi-level therapy mode where breaths are
machine-triggered and machine-cycled. T mode provides mandatory
pressure assist with bi-level pressures. The patients breathing rate has
no eect on the machine rate or pressure levels. The trigger to IPAP is
determined by the breath rate setting, and the cycle time is determined by
the inspiratory time setting.
PC Pressure Control Pressure Support; A Bi-level therapy mode where each
breath is patient or machine-triggered and machine-cycled. PC mode
is similar to S/T mode, except that all breaths are machine-cycled. This
is a pressure-limited, machine or patient-triggered, time-cycled mode.
Therefore, the inspiratory time may be triggered by the patient or by the
device, but IPAP will be pressure-limited with a set cycle time determined by
the Inspiratory Time setting.
Chapter 1 Introduction
11
Available Therapy Features
If prescribed for you, the device provides the following therapy features.
AVAPS
If enabled, Average Volume Assured Pressure Support (AVAPS) is a feature available in the S, S/T,
PC, and T modes. It helps patients maintain a tidal volume (V
T
) equal to or greater than the target
tidal volume (Tidal Volume setting in the AVAPS) by automatically controlling the pressure support
(PS) provided to the patient. The AVAPS feature adjusts PS by varying the IPAP level between the
minimum (IPAP Min) and maximum (IPAP Max) settings. AVAPS averages V
T
and changes the PS
value gradually. This occurs over several minutes. The rate of change is slow, so that the patient is
not aware of breath to breath pressure changes.
As patient eort decreases, AVAPS automatically increases PS to maintain the target tidal volume.
The IPAP level will not rise above IPAP Max, even if the target tidal volume is not reached.
Conversely, as patient eort increases, AVAPS will reduce PS. IPAP will not fall below IPAP Min, even
if the target tidal volume is exceeded. If IPAP Max is reached and the target tidal volume is not
achieved, the Low Tidal Volume alarm activates.
Bi-Flex Comfort Feature
If enabled, the device provides a comfort feature called Bi-Flex in S mode only. The Bi-Flex
attribute adjusts therapy by inserting a small amount of pressure relief during the latter stages of
inspiration and during active exhalation (the beginning part of exhalation). Bi-Flex levels of 1, 2, or 3
progressively reect increased pressure relief that will take place at the end of inspiration and at the
beginning of expiration.
Ramp
If enabled, the device is equipped with a linear ramp function. The Ramp feature will reduce the
pressure and then gradually increase (ramp) the pressure to the prescription pressure setting so
patients can fall asleep more comfortably.
BiPAP AVAPS user manual
12
Rise Time
If enabled, the device provides a feature called Rise Time in S, S/T, T, and PC modes. Rise time is the
amount of time it takes the device to change from the expiratory pressure setting to the inspiratory
pressure setting. Rise time levels of 1, 2, 3, 4, 5, or 6 progressively reect slowed response of the
pressure increase that will take place at the beginning of inspiration. A setting of 1 is the fastest
rise time while a setting of 6 is the slowest. Providers should adjust the rise time to nd the most
comfortable setting for the patient. Rise time cannot be adjusted when Bi-Flex is enabled.
Symbols
The following symbols appear on the device and power supply.
Symbol Description
Consult accompanying instructions for use.
AC Power
DC Power
IP22
Drip Proof Equipment
Caution, consult accompanying documents.
ESD Warning Symbol
Class II (Double Insulated)
Type BF Applied Part
For indoor use only
Do not disassemble
For Airline Use. Complies with RTCA DO-160F section 21, category M.
Chapter 1 Introduction
13
How to Contact Philips Respironics
To have your device serviced, contact your home care provider. If you need to contact Philips
Respironics directly, call the Customer Service department at 1-724-387-4000 or 1-800-345-6443.
You can also use the following addresses:
Caution: U.S. federal law restricts this device to sale by or on the order of a
physician.
Use only with the standard 60W power supply 1091398.
(not for use with Heated Tubing)
Use only with the Heated Tubing compatible 80W power supply 1091399. (can
also be used when Heated Tubing is not in use)
BiPAP AVAPS user manual
14
Chapter 2 Device Setup
15
Installing the Air Filter
The device uses a gray foam lter that is washable and reusable. The reusable lter screens out
normal household dust and pollen. It must be in place at all times when the device is operating.
One reusable gray foam lter is supplied with your device. If your lter is not already installed when
you receive the device, you must install it before using the device. To install the gray foam lter,
insert it into the lter area.
Where to Place the Device
Place the device upright on a rm, at surface somewhere within easy reach of where you will use
it, at a level lower than your sleeping position. Make sure the lter area on the back of the device
is not blocked by bedding, curtains, or other items. Air must ow freely around the device for the
system to work properly. Make sure the device is away from any heating or cooling equipment
(e.g., forced air vents, radiators, or air conditioners).
BiPAP AVAPS
user manual
2. Device Setup
BiPAP AVAPS user manual
16
Connecting the Breathing Circuit
To use the system, you will need the following accessories in order to assemble the recommended circuit:
t Philips Respironics interface (nasal mask or full face mask) with integrated exhalation
port, or Philips Respironics interface with a separate exhalation device (such as the
Whisper Swivel II)
t Philips Respironics 1.83 m (6 ft.) 22 mm flexible tubing or the optional 15 mm
flexible tubing
t Philips Respironics headgear (for the mask)
Complete the following steps to connect your breathing circuit to the device:
1. Connect the exible tubing to the air outlet on the side of the device.
Note: If required, connect a bacteria lter to the device air outlet, and then connect the exible tubing to the outlet
of the bacteria lter.
Note: When using the bacteria lter, the device performance may be aected. However, the device will remain
functional and deliver therapy.
Note: When using the optional heated tubing, attach the heated tubing to the humidier’s modied air outlet port,
with the bacteria lter installed in-line, but at the opposing end of the tubing.
2. Connect the tubing to the mask. Refer to the instructions that came with your mask.
3. Attach the headgear to the mask if necessary. Refer to the instructions that came with your
headgear.
Warning:
The exhalation device (Whisper Swivel II) or exhalation port (on masks with an integrated exhalation port) is
designed to exhaust CO
2
from the patient circuit. Do not block or seal the ports on the exhalation device.
Warning:
If you are using a full face mask (a mask covering both your mouth and your nose), the mask must be equipped
with a safety (entrainment) valve. You must ensure that the entrainment valve is functioning properly.
Chapter 2 Device Setup
17
Supplying AC Power to the Device
Complete the following steps to supply AC power to the device:
1. Plug the socket end of the power cord (included) into the power supply (also included).
Important! When you are using Heated Tubing with the compatible System One Heated Humidier, you must use
the 80W power supply.
2. Plug the pronged end of the power cord into an electrical outlet that is not controlled by a
wall switch.
3. Plug the power supply cord’s connector into the power inlet on the back of the ventilator.
4. Ensure that all connections are secure.
Important! To remove AC power, disconnect the power supply cord from the electrical outlet.
Note: See Chapter 4 for instructions on using DC Power.
BiPAP AVAPS user manual
18
Display Symbols
The following symbols may display on the device in place of text if the display language selected by
your home care provider is “Icon.
Symbol Description
Activate Mode
Alarm
Alarm Silence
Apnea
AVAPS
AVAPS therapy feature
Back
Backlight
Flex therapy feature
Flex Lock
Blower Hours
BPM
Breaths Per Minute
Clear Patient Data (in progress)
Clear Patient Data Failed
Clear Patient Data Successfully
Comfort Setting
hPa/cmH
2
O
Humidier, Humidity Level
Information
IPAP Max
IPAP Min
Language
Chapter 2 Device Setup
19
Symbol Description
Leak
Machine Hours
Min Vent
Minute Ventilation
Mode
No
No Settings Available
O (disabled)
On (enabled)
Patient Disconnect
Provider Mode
Ramp Start Pressure
Ramp Time
Reinsert SD Card
Reset Therapy Hours
Reset Blower Hours
RR
Respiratory Rate
Rise Time
Rise Time Lock
SD Card Corrupted
SD Card Full
SD Card Inserted
BiPAP AVAPS user manual
20
Symbol Description
SD Card Inserted: Writing in Progress
SD Card Inserted: Writing Successful
SD Card Inserted: Prescription Accepted
SD Card Inserted: Prescription Rejected
SD Card is Write-Protected
SD Card Removed
Setup
Setup Parameter Display
SYSTEM
ONE
System One Humidication
System One Resistance
System One Resistance Lock
Therapy (Blower O)
Therapy (Blower On)
Therapy Hours
T
I
Timed Inspiration
Ventilator Inoperative
V
T
Tidal Volume
V
TE
Exhaled Tidal Volume
Tubing Type
Tubing Type Lock
Tube Temperature
Yes (Selection Conrmed)
Chapter 2 Device Setup
21
Navigating the Device Screens
Turn the Wheel to toggle between options and settings on the screen. Press the Wheel to choose an
option or setting that is highlighted. If you choose “Back or the icon on any screen, it will take
you back to the previous screen.
Note: The screens shown in this manual are examples only. Information on your device screens may be dierent
depending on your prescription settings.
Note: Your device will either display in icon mode or text mode. Examples will be shown in both modes.
Starting and Stopping the Device
1. Supply power to the device.
The rst screen to display will be the Philips Respironics logo, followed by the screen showing the
current software version number. The Blower hours screen (
in icon mode) will then appear,
which shows the total blower hours for the device in hours and minutes.
2. The Main Menu screen appears, shown below.
Therapy
Info
Comfort
Setup
Icon View - Blower O Text View
Icon View - Blower On
3. Put on your mask assembly.
Note: If you are having trouble with your mask, refer to the instructions supplied with the mask.
4. Turn the Wheel to toggle between the four options. Highlight Therapy or the icon. Press
the Wheel to turn on the airow and begin therapy. The Monitor Pressure screen will appear,
described in detail in the next section.
5. Verify that the device beeps and the alarm and ramp LEDs light up each time therapy is started.
If the device does not operate accordingly, contact your home care provider, as the alarm
system may not be fully functional.
6. Make sure that no air is leaking from your mask into your eyes. If necessary, adjust the mask
and headgear until the air leak stops. See the instructions provided with your mask for more
information.
Note: A small amount of leak is normal and acceptable. Correct large mask leaks or eye irritation from an air leak as
soon as possible.
BiPAP AVAPS user manual
22
7. If you are using the device in a bed with a headboard, try placing the tubing over the headboard. This
may reduce tension on the mask.
8. Press and hold the Wheel for approximately 2 seconds to turn o therapy and return to the
Main Menu.
9. Verify that the device beeps when therapy is stopped. If the device does not operate
accordingly, contact your home care provider, as the alarm system may not be fully functional.
Monitor Pressure Screen
From the Main Menu, if you select Therapy and then press the Wheel, the following Monitor
Pressure screen appears.
S/T
20.0 cm H
2
O
AVAPS
RR 15
The Monitor Pressure screen displays the following items:
t Pressure
t Therapy Mode (CPAP, S, S/T, T, or PC)
t Timed Breath Indicator ( )
t Icon Bar
t Measured Parameters
Note: The Ramp symbol will also appear on the display if Ramp is active.
Note: If an accessory is attached to the therapy device, additional symbols may appear on the Monitor Pressure
screen. Refer to the instructions provided with the accessory for more information.
Chapter 2 Device Setup
23
The top of the display shows a group of status symbols. The symbols will only appear if the
conditions described in the following table exist.
Symbol Description
The Provider Access symbol indicates the device is in Provider mode.
AVAPS The AVAPS symbol displays only when the AVAPS therapy feature is enabled by
the provider.
FLEX The Flex symbol displays only when the Bi-Flex therapy feature is enabled by
the provider.
The Apnea alarm symbol displays only when the Apnea alarm is enabled by
the provider.
The Patient Disconnect symbol displays only when the Patient Disconnect alarm
is enabled by the provider.
The bottom section of the display shows additional measured parameters which may include:
t Respiratory Rate (RR)
t Tidal Volume in milliliters (ml)
t Minute Ventilation (Min Vent) in liters per minute (lpm)
t Leak in lpm
Note: The measured parameters display one at a time on-screen.
Changing the Comfort Settings
Your device is equipped with optional Flex, Ramp Time, and Rise Time features that your health care
professional may prescribe for you. From the Main Menu, when you highlight the Comfort option
and press the Wheel, the Comfort Settings screen below appears.
10
4 cm H
2O
10
4 cm H
2O
Note: If no comfort settings are available, the Comfort Settings screen displays “No Settings Available.
Note: If your home care provider has locked a comfort setting, a lock symbol (
) appears next to the value. You
cannot adjust any settings that are locked.
BiPAP AVAPS user manual
24
Flex Setting
The Flex comfort setting allows you to adjust the level of pressure relief that you feel during
therapy. Your home care provider can enable, lock, or disable this feature. When your provider
enables Flex, a level will already be set for you on the device. If this is not comfortable, you can
increase or decrease the setting from 1 to 3. A setting of 1 provides a small amount of pressure
relief, with higher numbers providing additional relief.
Ramp Time Setting
This enables you to modify the Ramp time setting in 5 minute increments. The range for this setting
is 0 to 45 minutes.
Ramp Start Setting
The device is equipped with an optional ramp feature that your home care provider can enable or
disable. This feature reduces the air pressure when you are trying to fall asleep and then gradually
increases (ramps) the pressure until your prescription setting is reached, allowing you to fall asleep
more comfortably.
Note: If the ramp feature is disabled, nothing will happen when you press the Ramp button.
If ramp is enabled on your device, after you turn on the airow, you can press the Ramp button on
the top of the device. Use the Ramp button as often as you like during the night. When Ramp is
enabled, the Ramp icon ( ) appears on the Monitor Pressure screen.
The Ramp Start pressure setting can be increased or decreased from 4 in increments of 1 to the
CPAP setting (if in CPAP therapy mode) or the EPAP setting (for all other therapy modes).
Rise Time Setting
Rise time is the time it takes for the device to change from EPAP to IPAP. If rise time is prescribed
for you, you can adjust the rise time from 1 to 6 to nd the setting that provides you with the most
comfort. A setting of 1 is the fastest rise time, while 6 is the slowest.
Note: If Flex is enabled, the rise time setting will be xed at 3.
Language
T
his feature allows you to choose which language to display on the interface when in Text
mode. You can also turn o (0) text mode which means the device will display the “Icon
Mode” on the interface.
Note: Both “Icon Mode” and English “Text Mode” are shown throughout this guide for your reference.
Chapter 2 Device Setup
25
Setup Screen ( )
From the Main Menu, highlight “Setup or the icon and press the wheel. The following Setup
screen will appear. The user can change settings in the Setup menu.
o
Setup
Back
Back light on
hPa/cmH
2
O cmH
2
O
Humidier 3
cmH
2
O
Icon View
Text View
Heated Tube Humi o
Humidity Level 4
Tube Temperature 3
System One humidi on
Tubing Type 15
System One Resist X1
Back
3
4
3
on
SYSTEM
ONE
15
X1
on
Note: The screen will only show a few lines at a time. As you rotate the wheel to toggle over dierent options
the screen will slide up and down accordingly. If the text is too long to completely t on the screen, it will scroll
horizontally across the screen when highlighted.
The following options appear on the Setup screen:
Heated Tube Humidication
This setting will only display if you are using the heated tube. You can enable (1) or disable
(0) this feature.
Humidity Level
This setting will only display if you are using the heated tube. This setting allows you to
choose the desired humidity setting for the humidier: 1, 2 or 3. This setting can only be
changed from the Setup screen.
Tube Temperature
This setting will only display if you are using the heated tube. This setting allows you to
choose the desired temperature for the heated tube: 0, 1, 2, 3, 4 or 5. If you choose zero (0),
this will turn o both the humidier and the heated tube.
Note: When using Heated Tubing, the control wheel can also be used to change this setting.
BiPAP AVAPS user manual
26
SYSTEM ONE Humidication
System One humidity control maintains a consistent mask humidity by monitoring and
adjusting for changes in room temperature and room humidity. You can enable (1) or
disable (0) this feature. If the System One humidity control has been disabled, the classic
style of basic temperature controlled heated humidication will be used. This will only
display if the humidier is attached.
Note: The System One Humidication option is only available if the Heated Tubing is removed or has been disabled.
Humidier
This setting allows you to view and choose the desired humidity setting. This will only
display if the humidier is attached. Please refer to the humidier manual if using a
humidier.
Note: The Humidier option is only available if the Heated Tubing is removed or has been disabled.
Tubing Type Setting
This setting allows you to select the correct size diameter tubing that you are using with the device.
You can choose either (22) for the Philips Respironics 22 mm tubing, or (15) for the optional Philips
Respironics 15 mm tubing. When using Heated Tubing, the device will automatically change this
setting to the appropriate tubing type (15H) and you will not be able to change it..
Note: If the Heated Tubing is removed, the device will default back to the previous tubing type setting.
Warning:
If you are using the optional Respironics 15 mm tubing, the device tubing type setting must be set to
15. If your device does not have the tubing type setting, you must use the Respironics 22 mm tubing selection.
SYSTEM ONE Resistance
This setting allows you to adjust the level of air pressure relief based on the specic Respironics
mask. Each Respironics mask may have a “System One resistance control setting. Contact your
home care provider if you cannot nd this resistance setting for your mask. If your provider has
locked the resistance setting into place, you can view the setting but cannot change it, and the
screen will display a lock symbol (
) next to the setting. If your provider has disabled resistance,
you will not see this setting.
Backlight
You can enable or disable the button LED backlight on the device.
hPa/cmH
2
O
You can select either hPa or cmH
2
O as the default unit of measure on the device.
Chapter 2 Device Setup
27
Humidier Preheat
When using a humidier, the device can preheat the water tank for up to 30 minutes prior to
starting therapy.
In order to activate the preheat mode, the blower must be o and a humidier must be attached.
From the device Home screen, highlight “Therapy or the
icon, then press and hold down the
control wheel for 5 seconds. You will hear a single beep and the device will now be in preheat
mode. The humidier icon ( ) will illuminate during this time.
During the 30 minute preheat, you will still be able to use the control wheel to select other menu
options from the Home screen. If you press the wheel while “Therapy or the
icon is highlighted
on the Home screen, preheat mode will end and the blower will turn on” to begin therapy. The
humidier number selected in the setup menu (0, 1, 2, 3, 4, or 5) will now take eect.
Viewing the Information Screen
From the Main Menu, when you highlight the Info option and press the Wheel, the Information
screen below appears. You cannot change settings on the Information screen.
Info
Back
Monitor Parameters
Blower Hours 890.1
Therapy Hours 890.1
Machine Hours 902.0
890.1
890.1
902.0
Icon View
Text View
Note: The Information screen is only for reference. Your home care provider may periodically ask you for this
information.
Note: If an accessory is attached to the therapy device, additional items may appear on the Information screen.
Refer to the instructions provided with the accessory for more information.
The following items appear on the Information screen:
t Monitor Parameters - Displays the available parameters.
t Therapy Hours - The device displays the total number of hours that the blower has been
on and patient breathing has been detected.
t Blower Hours - Displays the total number of hours that the blower has been on. It can
be reset by your home care provider. This setting allows the provider to track device
usage between patients.
t Machine Hours - Displays the total number of hours that the blower has been on. This
cannot be reset by the home care provider.
BiPAP AVAPS user manual
28
Viewing the Monitor Parameters Screen
There are two ways to access the Monitor Parameters screen:
t From the Monitor Pressure screen, press the Alarm Silence and Ramp keys
simultaneously for two seconds.
t From the Information screen, select the Monitor Parameters setting.
The parameters displayed in this screen are described in the following table. A sample screen is
shown below.
cmH
2
O
4.0
6
MinVent
6
Vte
200
RR
10
Press the Wheel to exit the Monitor Parameters screen and return to the previous screen.
Note: The information displayed in the last box shown on the sample screen will vary depending on what accessory
is attached to the therapy device. This box will be empty (as shown here) if no accessory is attached. Please refer to
the instructions included with your accessory for more information.
Viewing Measured Parameters
Several measured parameters can be viewed on-screen. The following table describes each
measured parameter. The measured parameters that display on the Monitor Pressure screen only
appear one at a time. The Setup Parameter Display setting on the Setup screen allows you to
choose which measured parameters you want displayed. The parameters below appear on both the
Monitor Pressure and Monitor Parameters screens.
Parameter Description
Pressure Displays the current patient pressure.
Leak (
)
The estimated leak is the average leak value for the last 6 breaths. The display is
updated at the end of each breath.
Respiratory Rate (RR) This is the average of the previous 6 breaths. If the mode supports machine-triggered
breaths, this display will be the total breathing rate (spontaneous breaths + machine
breaths). The display is updated at the end of each breath.
Minute Ventilation
(MinVent)
The estimated Exhaled Minute Ventilation is based on the average of the last 6
breaths. The display is updated at the end of each breath.
Exhaled Tidal
Volume (Vte)
The estimated Exhaled Tidal Volume is obtained by the integration of patient ow.
The display is updated at the end of each breath.
Note: If an accessory is attached to the therapy device, additional parameters may appear on-screen. Please refer to
the instructions included with your accessory for more information.
Chapter 3 Device Alarms
29
BiPAP AVAPS
user manual
3. Device Alarms
This chapter describes the ventilator alarms and what you should do if an alarm occurs.
There are three types of alarms:
t High Priority – Require immediate response by the operator
t Medium Priority – Require prompt response by the operator
t Low Priority – Require operator awareness. These alarms alert you to a change in the
ventilator status.
Additionally, the ventilator also displays informational messages and conrmation alerts that notify
you of conditions that need attention but are not alarm conditions.
Audible and Visual Alarm Indicators
When an alarm condition occurs:
t The alarm LED indicator on the Alarm Silence/Indicator button lights
t The audible alarm sounds
t A message appears on the screen describing the type of alarm
Note: If multiple alarms occur at the same time, only the highest priority alarm will be active. The precedence is in
the following order: high priority, medium priority, then low priority. When multiple priority alarms are active, the
highest priority LED light displays and the highest priority audible indicator sounds. On the display screen, the last
highest priority alarm displays.
Note: Informational messages are a lower precedence than alarms and will not display on the screen if any alarm is active.
BiPAP AVAPS user manual
30
Alarm LED Indicators
The Alarm Silence/Indicator button lights up as follows whenever an alarm is detected:
t Red Flashing Indicator – High priority alarm is detected.
t Yellow Flashing Indicator – Medium priority alarm is detected.
t Yellow Solid Indicator – Low priority alarm is detected.
The Alarm Silence/Indicator button does not light up when informational messages display.
Alert Audible Indicators
An audible indicator sounds whenever a power failure or a high, medium, or low priority alarm
is detected. Additionally, an audible indicator sounds for informational messages and to conrm
that certain actions have occurred (for example, when an SD card is inserted or removed from the
device).
t Ventilator Inoperative – When a ventilator inoperative alarm occurs, a continuous
audible indicator sounds. The alarm descriptions later in this chapter display this
indicator as:
t Power Failure – When a power failure occurs, a series of beeps sounds in a 1 beep
pattern, repeating one second on, then one second o. The alarm
descriptions later in
this chapter display this indicator as: tt
t High Priority – When a high priority alarm is active, a series of beeps sounds in the following
pattern, which is repeated twice: 3 beeps, a pause, and then 2 more beeps. This indicator
continues until the cause of the alarm is corrected or the audible alarm is silenced. The
alarm descriptions later in this chapter display this indicator as:
tttttttttt
t Medium Priority – When a medium priority alarm is active, a series of beeps sounds
in a 3-beep pattern. This pattern repeats until the cause of the alarm is corrected or
the audible alarm is silenced. The alarm descriptions later in this chapter display this
indicator as:
ttt
t Low Priority – When a low priority alarm is active, a series of beeps sounds in a 2-beep
pattern. This pattern repeats until the cause of the alarm is corrected or the audible
alarm is silenced. The alarm descriptions later in this chapter display this indicator as:
tt
t Informational Messages and Conrmation Audible Indicators – When an informational
message appears on screen, a brief, 1- beep audible indicator sounds. Additionally,
when the device detects that a certain action has been completed (for example, when
an SD card is inserted or removed from the device) a brief, 1- beep audible indicator
sounds. The descriptions later in this chapter display this indicator as:
t
Chapter 3 Device Alarms
31
Silencing an Alarm
You can silence an alarm by pressing the Alarm Silence/Indicator button. This will silence the alarm
for one minute. An icon will appear on the screen when the alarm is silenced ( ). If another
alarm occurs while the silence period is active, the audible alarm portion of the new alarm will not
sound until the silence period expires. When the silence period expires, the alarms audible alarm
is reactivated. Touching the Alarm Silence/Indicator button while the silence period is active will
restart the silence period.
Alarm Message Screens
When an alarm message is activated, an alarm screen is displayed, showing the text or icon specic
to the most recent, highest priority alarm.
Pressing the Control Wheel will reset the alarm and remove the alarm screen from the display.
Resetting the alarm allows you to return to the previous screen. If multiple alarms occur during the
same period of time, the alarm screen will display the higher priority alarm (higher priority alarms
take precedence over lower priority alarms).
Note: Pressing the Control Wheel resets all alarms.
Note: If the alarm pop-up is present, you cannot see the Monitor Pressure screen.
BiPAP AVAPS user manual
32
Alarm Summary Table
The following table summarizes all of the high, medium, and low priority alarms and informational
messages.
Alarm Priority Audible
Indicator
Visual Indicators Device
Action
User Action
Loss of
Power
High
tt
Red ashing button; Blank
screen
Shuts down Remove your mask. Check
your power connections.
Make sure there is power at
the outlet or power source.
Restore power to the device. If
the alarm continues, contact
your home care provider for
service.
Ventilator
Inoperative
High
Red solid button;
Ventilator Inoperative
- or - (if Icon option is
selected)
Shuts down Remove your mask. Press the
Alarm Silence/Indicator button
to silence the alarm. Contact
your home care provider for
service.
Low Pressure
Alarm
High
ttttt
Red ashing button;
Low Pressure
- or - (if Icon option is
selected)
cm
H2O
Operates This could be caused by an
excessive leak or blockage or
a device malfunction. Press
the Alarm Silence/Indicator
button to silence the alarm.
Remove your mask. Check for
the following: dirty inlet lters,
blocked air intake, excessive
leak in the patient circuit. If the
alarm continues, contact your
home care provider.
High
Pressure
High
ttttt
Red ashing button
High Pressure
- or - (if Icon option is
selected)
cm
H2O
Operates;
If the alarm
continues for
10 seconds,
the alarm
escalates to
a Ventilator
Inoperative
alarm
This may be caused by a
malfunctioning device. Press
the Alarm Silence/Indicator
button to silence the alarm.
Remove your mask Remove
power from the device.
Restore power. If the alarm
continues, contact your home
care provider for service.
Chapter 3 Device Alarms
33
Alarm Priority Audible
Indicator
Visual Indicators Device
Action
User Action
Apnea High
ttttt
Red ashing button
Apnea
- or - (if Icon option is
selected)
Operates This alarm is generated when
an apnea event occurs during
therapy. Press the Alarm
Silence/Indicator button to
silence the alarm. Report
the alarm to your home care
provider. Continue using your
device.
Low Minute
Ventilation
High
ttttt
Red ashing button
Low Minute Vent
- or - (if Icon option is
selected)
MinVent
Operates This alarm occurs when the
calculated minute ventilation
is less than or equal to the
alarm setting. Press the Alarm
Silence/Indicator button to
silence the alarm. Report
the alarm to your home care
provider. Continue using your
device.
Low Tidal
Volume
High
ttttt
Red ashing button
Low Tidal Volume
- or - (if Icon option is
selected)
V
TE
Operates Only enabled if AVAPS therapy
feature is enabled; This alarm
occurs when the device is
unable to reach the target
tidal volume setting. Press the
Alarm Silence/Indicator button
to silence the alarm. Report
the alarm to your home care
provider. Continue using your
device.
Patient
Disconnect
High
ttttt
Red ashing button
Patient Disconnect
- or - (if Icon option is
selected)
Operates This alarm occurs when the
patient circuit is disconnected
or has a large leak. Press
the Alarm Silence/Indicator
button to silence the alarm.
Reconnect the patient circuit
or x the leak. If the alarm
continues, contact your home
care provider for service.
BiPAP AVAPS user manual
34
Alarm Priority Audible
Indicator
Visual Indicators Device
Action
User Action
Low Input
Voltage
Medium
ttt
Yellow ashing button
Low Voltage
- or - (if Icon option is
selected)
Ventilator
Operates;
Humidier
shuts down
The alarm is caused when
input power at the device,
either from an AC outlet
or battery, falls below the
acceptable limit for 10
seconds.
Press the Alarm Silence/
Indicator button to silence the
alarm. If the device is plugged
into a wall outlet, unplug the
device and then plug it back
in. If the alarm continues to
occur, contact your home care
provider for service.
If you are using a battery,
replace the battery or plug the
device into an AC outlet. If the
alarm continues, contact your
home care provider for service.
SD Card
Corrupted
Low
tt
Solid yellow button
SD card corrupted
- or - (if Icon option is
selected)
Operates This alarm occurs when a
problem exists with the
SD card. The data may be
corrupted. Press the Alarm
Silence/Indicator button to
silence the alarm. Contact your
home care provider with any
questions.
SD Card Full Low
tt
Solid yellow button
SD card full
- or - (if Icon option is
selected)
Operates This alarm occurs when the
SD card is full. Press the Alarm
Silence/Indicator button to
silence the alarm. Remove the
SD card and replace it.
Chapter 3 Device Alarms
35
Alarm Priority Audible
Indicator
Visual Indicators Device
Action
User Action
SD Card:
Remove and
Reinsert
Low
tt
Reinsert SD Card
- or - (if Icon option is
selected)
Operates This alarm occurs when the
device cannot read the SD
card. The card may be inserted
incorrectly. Remove the SD
card and reinsert. If the alert
continues to occur, replace the
SD card or contact your home
care provider.
SD Card:
Prescription
Accepted
Info
t
SD card inserted:
prescription
accepted
- or - (if Icon option is
selected)
Operates This info message will be
present for 30 seconds or until
the user acknowledges it. No
action needed.
SD Card:
Prescription
Rejected
Info
t
SD card inserted:
prescription
rejected
- or - (if Icon option is
selected)
Operates This message occurs when
the prescription is missing
or incorrect. It is present for
30 seconds or until the user
acknowledges it. Contact your
home care provider for the
correct prescription.
SD Card
Inserted
Info
t
SD card inserted
- or - (if Icon option is
selected)
Operates This message occurs when
the SD card is inserted into
the device. It is present for
30 seconds or until the user
acknowledges. No action is
needed.
SD Card
Removed
Info
t
SD card removed
- or - (if Icon option is
selected)
Operates This message occurs when
the SD card is removed from
the device. It is present for
30 seconds or until the user
acknowledges it. No action is
needed.
BiPAP AVAPS user manual
36
Alarm Priority Audible
Indicator
Visual Indicators Device
Action
User Action
Check Power
Alert
Info None The following symbol Shuts down The power supply voltage is
incorrect. Make sure that you
are using the correct power
supply with your device. If
the alert continues to occur,
contact your home care
provider.
Humidier
Alert
Info None Humidier LED icon will
ash on the device.
Only
displayed
when
both the
humidier
and therapy
is on.
Humidier failure. Alert is
present for 12 minutes or
until the condition is xed.
Turn o airow and reconnect
the humidier to the device
according to the humidier
instructions. If the alert
continues to occur, contact
your home care provider.
Power
Supply Alert
Info None Humidier LED icon will
ash for 30 seconds.
Only
displayed
when
incorrect
power
supply is
used with
the heated
tube.
Using wrong power supply.
Alert is present for 30 seconds
or until the condition is xed.
You must use the 80W power
supply when using the heated
tube. If the alert continues to
occur, contact your home care
provider.
Heated Tube
Error Alert
Info None Humidier LED icon will
slowly ash for 30 seconds.
Alert
present for
30 seconds
or until
condition is
xed.
Tubing may be overheating
or malfunctioning. Alert is
present for 30 seconds or until
the condition is xed. Turn
o airow and reconnect the
heated tubing to the humidier
according to the humidier
instructions. If the alert
continues to occur, contact
your home care provider.
Humidier
Failure
Info None Flashing humidier LED
icon
Device
operate;
Humidier
shuts down
Alert is present for 12 minutes
or until the condition is xed.
Turn o airow and reconnect
the humidier to the device
according to the humidier
instructions. If the alert
continues to occur, contact
your home care provider.
Chapter 3 Device Alarms
37
Troubleshooting
The table below lists some of the problems you may experience with your device and possible
solutions to those problems.
Problem Why It Happened What To Do
Nothing happens
when you apply
power to the device.
The backlight on the
buttons does not
light.
There is no power
at the outlet or
the device is
unplugged.
If you are using AC power, check the outlet and
verify that the device is properly plugged in. Make
sure there is power available at the outlet. Make
sure the AC power cord is connected correctly to
the power supply and the power supply cord is
securely connected to the devices power inlet. If
the problem continues to occur, contact your home
care provider. Return both the device and power
supply to your provider so they can determine if the
problem is with the device or power supply.
If you are using DC power, make sure your DC
power cord and battery adapter cable connections
are secure. Check your battery. It may need
recharged or replaced. If the problem persists,
check the DC cord’s fuse following the instructions
supplied with your DC cord. The fuse may need to
be replaced. If the problem still occurs, contact your
home care provider.
The airow does not
turn on.
There may be a
problem with the
blower.
Make sure the device is powered correctly.
Make sure “Therapy or
is highlighted when
pressing the control Wheel to start airow. If the
airow does not turn on, there may be a problem
with your device. Contact your home care provider
for assistance.
The device display is
erratic.
The device has
been dropped or
mishandled, or
the device is in
an area with high
Electromagnetic
Interference (EMI)
emissions.
Unplug the device. Reapply power to the device. If
the problem continues, relocate the device to an
area with lower EMI emissions (away from electronic
equipment such as cellular phones, cordless phones,
computers, TVs, electronic games, hair dryers, etc.).
If the problem still occurs, contact your home care
provider for assistance.
BiPAP AVAPS user manual
38
Problem Why It Happened What To Do
The Device will not
turn o.
The correct blower
o sequence was
not followed.
Select Therapy to go back to the Monitor Pressure
screen. Push and hold the knob for 2 seconds.
The Ramp feature
does not work when
you press the Ramp
button.
Your home care
provider did not
prescribe Ramp
for you, or your
pressure is already
set to the minimum
setting.
If Ramp has not been prescribed for you, the Ramp
feature will not work.
If your provider has enabled Ramp but the feature
still does not work, check the pressure setting on
your Monitor Pressure screen. If the pressure is set
to the minimum setting, or the starting pressure
is the same as the prescribed pressure, the Ramp
feature will not work.
The airow is much
warmer than usual.
The air lters may
be dirty.
The device may be
operating in direct
sunlight or near a
heater.
Clean or replace the air lters.
The temperature of the air may vary somewhat
based on your room temperature. Make sure the
device is properly ventilated. Keep it away from
bedding or curtains that could block the ow of air
around the device. Make sure the device is away
from direct sunlight and heating equipment.
If using the humidier with the device, check
the humidier settings. Refer to the humidier
instructions to make sure the humidier is working
properly.
The airow pressure
feels too high or too
low.
The Tubing type
setting may be
incorrect.
Make sure the Tubing type setting (22 or 15)
matches the tubing that you are using (Philips
Respironics 22 or 15 mm tubing).
If you are using the Heated Tubing, this setting will
be 15H and you cannot change it.
Chapter 3 Device Alarms
39
Problem Why It Happened What To Do
Tube Temperature is
turned on in “Setup
screen but Heated
Tubing is not warm.
Tube Temperature
is turned on in
“Setup screen but
Humidier LED does
not stay orange.
Incorrect power
supply is being
used (60W is used
instead of 80W).
Heated Tubing is
attached incorrectly
or damaged.
Make sure the 80W power supply is being used.
This can be conrmed by looking at the power
supply for the 60W or 80W symbols. This can also
be checked by looking at the “Humidier” settings
under the “Setup screen.
Inspect Heated Tubing for damage and reconnect.
If the problem continues, contact your home care
provider.
The mask feels
uncomfortable
to wear, there
is signicant air
leakage around
the mask, or you
experiences other
mask-related issues.
This could be due to
improper headgear
adjustment or
improper mask
tting.
Make sure you are properly tted with the correct
size mask. If the problem continues, contact your
home care provider to be tted with a dierent
mask.
You have a runny
nose.
This may be caused
by a nasal reaction
to the airow.
Contact your home care provider.
BiPAP AVAPS user manual
40
Chapter 4 Accessories
41
BiPAP AVAPS
user manual
4. Accessories
There are several accessories available for your BiPAP AVAPS system, such as a humidier. Contact
your home care provider for additional information on the available accessories. When using the
optional accessories, always follow the instructions enclosed with the accessories.
Adding a Humidier with or without Heated Tubing
You can use the heated humidier and the heated tube with your device. They are available from
your home care provider. A humidier and heated tube may reduce nasal dryness and irritation by
adding moisture to the airow.
Note: Refer to the humidier’s instructions for complete setup information.
SD Card
The system comes with an SD card inserted in the SD card slot on the back of the device to record
information for the home care provider. Your home care provider may ask you to periodically re-
move the SD card and send it to them for evaluation.
Note: The SD card does not need to be installed for the device to work properly. The SD card records device usage
information for your home care provider. You can refer to the Device Alarms chapter of this manual for more
information on the SD card. Contact your home care provider if you have any questions about the SD card.
Supplemental Oxygen
Oxygen may be added anywhere in the patient circuit provided that a pressure valve is used.
Please note the warnings in Chapter 1 when using oxygen with the device.
BiPAP AVAPS user manual
42
Shielded DC Cord
The Philips Respironics Shielded DC Cord can be used to operate this device in a stationary recre-
ational vehicle, boat, or motor home. The Philips Respironics DC Battery Adapter Cable, when used
with the Shielded DC Cord, enables the device to be operated from a 12 VDC free-standing battery.
Refer to the instructions supplied with the Shielded DC Cord and adapter cable for information on
how to operate the device using DC power.
Caution:
When DC power is obtained from a vehicle battery, the device should not be used while the vehicle’s engine is
running. The device may not work properly if connected while the vehicles engine is running.
Caution:
Only use a Philips Respironics Shielded DC Cord and Battery Adapter Cable. Use of any other system may cause
damage to the device or vehicle.
Carrying Case
When traveling, the carrying case is for carry-on luggage only. The carrying case will not protect this
system if it is put through checked baggage.
For your convenience at security stations, there is a note on the bottom of the device stating that
it is medical equipment. It may be helpful to bring this manual along with you to help security per-
sonnel understand the BiPAP AVAPS device.
If you are traveling to a country with a line voltage dierent than the one you are currently using,
a dierent power cord or an international plug adapter may be required to make your power cord
compatible with the power outlets of the country to which you are traveling. Contact your home
care provider for additional information.
Note: If you are using a humidier with the device, the humidier should be emptied before traveling.
Airline Travel
The device is suitable for use on airlines when it is operating from an AC or DC power source.
Note: The device is not suitable for airline use with any modems or humidiers installed.
Chapter 5 Cleaning the Device
43
BiPAP AVAPS
user manual
5. Cleaning the Device
Follow the instructions below to clean the device. If you are using the device on multiple users,
complete the following steps before each new user.
Warning: If you are using the device on multiple users, discard and replace the bacteria lter each time the
device is used on a dierent person.
1. Unplug the device before cleaning.
2. Clean the outside of the device only. Use a cloth with one of the following cleaning agents to
clean the exterior of the device.
t Mild detergent
t 70% Isopropyl Alcohol
t DisCide Towelettes
t 10% Chlorine bleach solution
3. Allow the device to dry completely before plugging in the power cord.
4. Inspect the device and all circuit parts for damage after cleaning. Replace any damaged parts.
Cleaning or Replacing the Filters
Under normal usage, you should clean the gray foam lter at least once every two weeks and
replace it with a new one every six months.
1. If the device is operating, stop the airow. Disconnect the device from the power source.
2. Remove the lter from the enclosure by gently squeezing the lter in the center and pulling it
away from the device.
3. Examine the lter for cleanliness and integrity.
BiPAP AVAPS user manual
44
4. Wash the gray foam lter in warm water with a mild detergent. Rinse thoroughly to remove all
detergent residue. Allow the lter to air dry completely before reinstalling it. If the foam lter is
torn, replace it. (Only Philips Respironics-supplied lters should be used as replacement lters.)
5. Reinstall the lter.
Cleaning the Tubing
Clean the exible tubing before rst use and daily. Disconnect the exible tubing from the device.
For the 15 or 22 mm exible tubing, gently wash the tubing in a solution of warm water and a mild
detergent. Rinse thoroughly. Air dry.
Note: Refer to the humidier manual for the instructions on how to clean the heated tube.
Service
The device does not require routine servicing.
Chapter 6 Specications
45
BiPAP AVAPS
user manual
6. Specications
Environmental
Operating Storage
Temperature 41° F to 95° F (5° C to 35° C) -4° F to 140° F (-20° C to 60° C)
Relative Humidity 15 to 95% (non-condensing) 15 to 95% (non-condensing)
Atmospheric
Pressure
101 kPa to 77 kPa
(0-7500 ft / 0-2286 m)
N/A
Physical
Dimensions: 7” L x 5.5” W x 4” H (18 cm x 14 cm x 10 cm)
Weight: Approximately 3 lbs (1.36 kg)
Standards Compliance
This device is designed to conform to the following standards:
t IEC 60601-1: Medical Electrical Equipment – Part 1: General Requirements for Basic
Safety and Essential Performance
t IEC 60601-1-2: Medical Electrical Equipment – Part 1-2: General Requirements for Basic
Safety and Essential Performance – Collateral standard: Electromagnetic compatibility -
Requirements and tests
t IEC 60601-1-8: Medical Electrical Equipment – Part 1-8: General Requirements for Basic
Safety and Essential Performance – Collateral standard: General Requirements, tests and
guidance for alarm systems in medical electrical equipment and medical electrical systems
t IEC 60601-1-11: Medical Electrical Equipment – Part 1-11: General Requirements for Basic
Safety and Essential Performance – Collateral standard: Requirements for medical electrical
equipment and medical electrical systems used in the home healthcare environment
t ISO 10651-6: Lung Ventilators for Medical Use – Particular Requirements for Basic Safety
and Essential Performance, Part 6. Home care ventilatory support devices
t RTCA DO-160F Section 21, Category M; Emission of Radio Frequency Energy
BiPAP AVAPS user manual
46
Electrical
AC Power Consumption (with 60W power supply): 100 to 240 VAC, 50/60 Hz, 2.1 A
AC Power Consumption (with 80W power supply): 100 to 240 VAC, 50/60 Hz, 2.0 A
DC Power Consumption: 12 VDC, 5.0 - 6.67 A
Fuses: There are no user-replaceable fuses.
Type of Protection Against Electric Shock: Class II
Degree of Protection Against Electric Shock: Type BF Applied Part
Degree of Ingress Protection: Device Drip Proof
(Device and AC power supply): IP22
Mode of Operation: Continuous
Pressure
Pressure Increments: 4.0 to 30.0 cm H
2
O (in 1.0 cm H
2
O increments)
Flex Therapy Feature: O, 1, 2, 3
Control Accuracy
Parameter Range Accuracy
IPAP 4 – 30 cm H
2
O ± 2.5 cm H
2
O*
EPAP 4 – 30 cm H
2
O ± 2.5 cm H
2
O*
CPAP 4 – 20 cm H
2
O ± 2.5 cm H
2
O*
Breath rate 0 to 30 BPM greater of ± 1 BPM or ±10% of setting
Inspiration time 0.5 to 3 seconds ± (10% of setting + 0.1 second).
*Pressure measured at the patient connection port with or without the humidier (no patient ow,
with Whisper Swivel II).
Chapter 6 Specications
47
Displayed Parameter Accuracy
Parameter Accuracy Resolution Range
Estimated Leak Rate ±(5+15% of reading) LPM 1 LPM 0 to 200 LPM
Exhaled Tidal Volume ±(25+15% of reading) ml 5 ml 0 to 2000 ml
Respiratory Rate Greater of ±1 BPM or ±10% of reading 1 BPM 0 to 60 BPM
Exhaled Minute Ventilation ±(1+15% of reading) LPM 1 LPM 0 to 99 LPM
Spontaneous Breathing During Power Failure Conditions
Patient Flow (LPM) Expiratory Resistance (cm H
2
O)
Passive Circuit
Inspiratory Resistance (cm H
2
O)
Passive Circuit
30 <1.0 <1.0
60 <2.8 <2.8
Noise
Minimum Alarm Sound Level: 45 dB(A)
Disposal
Dispose of the device in accordance with local regulations.
BiPAP AVAPS user manual
48
Chapter 7 EMC Information
49
Guidance and Manufacturer’s Declaration - Electromagnetic Emissions – This device is intended for
use in the electromagnetic environment specied below. The user of this device should make sure
it is used in such an environment.
Emissions Test Compliance Electromagnetic Environment - Guidance
RF emissions
CISPR 11
Group 1 The device uses RF energy only for its internal function. Therefore, its RF emissions
are very low and are not likely to cause any interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class B The device is suitable for use in all establishments, including domestic establishments
and those directly connected to the public low-voltage power supply network.
Harmonic emissions
IEC 61000-3-2
Class A
9ROWDJHÁXFWXDWLRQV)OLFNHU
emissions
IEC 61000-3-3
Complies
BiPAP AVAPS
user manual
7. EMC Information
BiPAP AVAPS user manual
50
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity – This device is intended for
use in the electromagnetic environment specied below. The user of this device should make sure
it is used in such an environment.
Immunity Test IEC 60601 Test
Level
Compliance Level Electromagnetic Environment -
Guidance
Electrostatic
Discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, concrete or
FHUDPLFWLOH,IÁRRUVDUHFRYHUHGZLWK
synthetic material, the relative humidity
should be at least 30%.
Electrical fast
7UDQVLHQWEXUVW
IEC 61000-4-4
±2 kV for power supply lines
±1 kV for input-output lines
±2 kV for supply mains
N9IRULQSXWRXWSXWOLQHV
Mains power quality should be that of a
typical home or hospital environment.
Surge
IEC 61000-4-5
±1 kV differential mode
±2 kV common mode
±1 kV differential mode
±2 kV for common mode
Mains power quality should be that of a
typical home or hospital environment.
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
<5% U
T
(>95% dip in U
T
) for
0.5 cycle
40% U
T
(60% dip in U
T
) for
5 cycles
70% U
T
(30% dip in
U
T
) for 25 cycles
<5% U
T
(>95% dip in U
T
) for
5 sec
<5% U
T
(>95% dip in U
T
) for
0.5 cycle
40% U
T
(60% dip in U
T
) for 5 cycles
70% U
T
(30% dip in U
T
) for
25 cycles
<5% U
T
(>95% dip in U
T
) for
5 sec
Mains power quality should be that of a
typical home or hospital environment.
If the user of the device requires
continued operation during power mains
interruptions, it is recommended that the
device be powered from an uninterruptible
power supply or a battery.
3RZHUIUHTXHQF\+]
PDJQHWLFÀHOG
IEC 61000-4-8
$P $P 3RZHUIUHTXHQF\PDJQHWLFÀHOGVVKRXOGEH
at levels characteristic of a typical location
in a typical hospital or home environment.
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.
Chapter 7 EMC Information
51
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity – This device is intended for
use in the electromagnetic environment specied below. The user of this device should make sure
it is used in such an environment.
Immunity Test IEC 60601 Test Level Compliance Level Electromagnetic Environment -Guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
N+]WR0+]
9P
0+]WR*+]
3 Vrms
9P
Portable and mobile RF communications equipment should be
used no closer to any part of the device, including cables, than
the recommended separation distance calculated from the
equation applicable to the frequency of the transmitter.
Recommended separation distance
d = 1.2
d = 1.2 0+]WR0+]
d = 2.3
0+]WR*+]
where P is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer and d is
the recommended separation distance in meters (m).
)LHOGVWUHQJWKVIURPÀ[HG5)WUDQVPLWWHUVDVGHWHUPLQHGE\DQ
electromagnetic site survey
a
, should be less than the compliance
level in each frequency range.
b
Interference may occur in the vicinity of equipment marked with
the following symbol:
127($W0+]DQG0+]WKHKLJKHUIUHTXHQF\UDQJHDSSOLHV
127(7KHVHJXLGHOLQHVPD\QRWDSSO\LQDOOVLWXDWLRQV(OHFWURPDJQHWLFSURSDJDWLRQLVDIIHFWHGE\DEVRUSWLRQDQGUHÁHFWLRQIURP
structures, objects, and people.
D )LHOGVWUHQJWKVIURPÀ[HGWUDQVPLWWHUVVXFKDVEDVHVWDWLRQVIRUUDGLRFHOOXODUFRUGOHVVWHOHSKRQHVDQGODQGPRELOHUDGLRVDPDWHXU
radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
HQYLURQPHQWGXHWRÀ[HG5)WUDQVPLWWHUVDQHOHFWURPDJQHWLFVLWHVXUYH\VKRXOGEHFRQVLGHUHG,IWKHPHDVXUHGÀHOGVWUHQJWKLQWKH
location in which the device is used exceeds the applicable RF compliance level above, the device should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the device.
E 2YHUWKHIUHTXHQF\UDQJHN+]WR0+]WKHÀHOGVWUHQJWKVVKRXOGEHOHVVWKDQ9P
BiPAP AVAPS user manual
52
Recommended Separation Distances between Portable and Mobile RF Communications Equipment
and This Device: The device is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of this device can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and this device as recommended below, according to
the maximum output power of the communications equipment.
Rated Maximum Power
Output of Transmitter
W
Separation Distance According to Frequency of Transmitter
m
150 kHz to 80 MHz
d = 1.2
80 MHz to 800 MHz
d = 1.2
800 MHz to 2.5 GHz
d = 2.3
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated
using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts
(W) according to the transmitter manufacturer.
1RWH$W0+]DQG0+]WKHVHSDUDWLRQGLVWDQFHIRUWKHKLJKHUIUHTXHQF\UDQJHDSSOLHV
1RWH7KHVHJXLGHOLQHVPD\QRWDSSO\LQDOOVLWXDWLRQV(OHFWURPDJQHWLFSURSDJDWLRQLVDIIHFWHGE\DEVRUSWLRQDQGUHÁHFWLRQIURPVWUXFWXUHV
objects, and people.
Limited Warranty
53
Respironics, Inc. warrants that the system shall be free from defects of workmanship and materials and will
perform in accordance with the product specications for a period of two (2) years from the date of sale by
Respironics, Inc. to the dealer. If the product fails to perform in accordance with the product specications,
Respironics, Inc. will repair or replace – at its option – the defective material or part. Respironics, Inc. will pay
customary freight charges from Respironics, Inc. to the dealer location only. This warranty does not cover
damage caused by accident, misuse, abuse, alteration, water ingress, and other defects not related to material
or workmanship. The Respironics, Inc. Service department shall examine any devices returned for service, and
Respironics, Inc. reserves the right to charge an evaluation fee for any returned device as to which no problem
is found after investigation by Respironics, Inc. Service.
This warranty is non-transferable by unauthorized distributers of Respironics, Inc. products and reserves
the right to charge dealers for warranty service of failed product not purchased directly from Respironics or
authorized distributers.
Respironics, Inc. disclaims all liability for economic loss, loss of prots, overhead, or consequential damages
which may be claimed to arise from any sale or use of this product. Some states do not allow the exclusion or
limitation of incidental or consequential damages, so the above limitation or exclusion may not apply to you.
This warranty is given in lieu of all other express warranties. In addition, any implied warranties – including any
warranty of merchantability or tness for the particular purpose – are limited to two years. Some states do not
allow limitations on how long an implied warranty lasts, so the above limitation may not apply to you. This
warranty gives you specic legal rights, and you may also have other rights which vary from state to state.
To exercise your rights under this warranty, contact your local authorized Respironics, Inc. dealer or contact
Respironics, Inc. at:
1001 Murry Ridge Lane
Murrysville, Pennsylvania 15668-8550
1-724-387-4000
BiPAP AVAPS
user manual
Limited Warranty
BiPAP AVAPS user manual
54