Page1of86
UserManual
Page2of86
TableofContents
TableofContents............................................................................................................................................2
GeneralInformation............................................................................................................................................5
QuidelContactInformation.............................................................................................................................5
MainMenuStructure......................................................................................................................................6
LabelsandSymbols.........................................................................................................................................7
Warnings,PrecautionsandLimitations...........................................................................................................9
Introduction.......................................................................................................................................................10
HowtoUseThisGuide..................................................................................................................................10
Description............................................................................................................................... .....................10
WhatistheQuidel
TriageMeterPro?.......................................................................................................10
Unpacking..................................................................................................................................................11
Installation.........................................................................................................................................................15
UnpackMeter............................................................................................................................... .................15
PowerMeterOn............................................................................................................................... .........15
InstallPaper............................................................................................................................... ................16
InsertSupervisorCodeChipModule.........................................................................................................17
ToChangeMeterSettings.........................................................................................................................17
ChangeClock.............................................................................................................................................19
SetUserID.................................................................................................................................................20
ChangeReference
Ranges/ThresholdsforTests:......................................................................................20
ChangeTestSettings.................................................................................................................................20
ChangeQCParameters..............................................................................................................................21
ChangeBypassSettings.............................................................................................................................21
SetParameters............................................................................................................................... ...............23
Purpose............................................................................................................................... .......................23
ListofProgrammableParameters.............................................................................................................23
SupervisorAccess......................................................................................................................................23
HowtoSetParameters:BasicInstructions...................................................................................................24
MeterSettings...........................................................................................................................................
25
IDSettings..................................................................................................................................................25
Asymbolsglossarycanbefoundatquidel.com/glossary.
Page3of86
InstallCodeChip............................................................................................................................................37
Purpose............................................................................................................................... .......................37
TypesofCODECHIPmodules.................................................................................................................... 37
InstallationofReagent,QCSampleorQCDeviceCODECHIPModules...................................................37
Operation..........................................................................................................................................................39
AccesstotheQuidelTriageMeterPro..........................................................................................................39
RunTest............................................................................................................................... ..........................39
Purpose............................................................................................................................... .......................39
Options............................................................................................................................... .......................
39
RecallResults............................................................................................................................... ..................49
Purpose............................................................................................................................... .......................49
Function.....................................................................................................................................................49
LastRecord............................................................................................................................... .................49
PatientResults...........................................................................................................................................50
PatientID...................................................................................................................................................50
TestDeviceLotNumber(L/N)...................................................................................................................51
UserID............................................................................................................................... ........................52
FromDate..................................................................................................................................................52
TillDate......................................................................................................................................................53
QCResults............................................................................................................................... ...................
54
QCSampleLotNumber(L/N).................................................................................................................... 55
DeviceLotNumber(L/N)........................................................................................................................... 55
UserID............................................................................................................................... ........................56
FromDate..................................................................................................................................................57
TillDate......................................................................................................................................................57
ReagentLotsQC............................................................................................................................... ......58
ArchiveLastTest............................................................................................................................... .........58
Misc.TestResults............................................................................................................................... .......59
Misc.TestID..............................................................................................................................................59
TestDeviceLotNumber(L/N)...................................................................................................................60
UserID............................................................................................................................... ........................61
FromDate..................................................................................................................................................61
TillDate......................................................................................................................................................62
PrintAllResults............................................................................................................................... ...........63
PatientResults...........................................................................................................................................63
QCSampleResults.....................................................................................................................................63
Page4of86
QCDeviceResults......................................................................................................................................64
Misc.TestResults............................................................................................................................... .......64
UploadResultstoLIS............................................................................................................................... ..64
DeleteResults................................................................................................................................................64
Purpose............................................................................................................................... .......................64
TheQuidelTriageMeterProMemory.......................................................................................................65
DeletePatientResult............................................................................................................................... ..65
DisplayBacklight............................................................................................................................................66
AlphaNumericMode....................................................................................................................................67
ServiceandMaintenance
Procedures............................................................................................................... 68
PaperReplacement............................................................................................................................... ........68
ChangingPaperBetweenFunctions..........................................................................................................68
ChangingPaperWhilePrinting..................................................................................................................69
Cleaning............................................................................................................................... ..........................70
CheckBatteryLevel............................................................................................................................... ........70
BatteryReplacement............................................................................................................................... ..70
TotalQualityAssurance............................................................................................................................... ..71
Introduction...............................................................................................................................................71
QualityAssuranceFunctions.....................................................................................................................72
InternalDeviceQuality
Assurance............................................................................................................72
MeterQualityAssurance...........................................................................................................................73
SoftwareQualityAssurance......................................................................................................................73
QualityControlSamples............................................................................................................................74
NotesRegardingSoftwareControls.......................................................................................................... 74
Appendices............................................................................................................................... .........................76
AppendixASpecifications.............................................................................................................................76
MeterSymptoms............................................................................................................................... ........77
AppendixBTroubleshooting......................................................................................................................... 78
MeterMessageorSymptom..................................................................................................................... 78
AppendixCReturn
andDisposalProcedures................................................................................................80
ReturnProcedure............................................................................................................................... .......80
DisposalProcedure............................................................................................................................... .....80
AppendixDGlossary............................................................................................................................... .......80
AppendixESampleLogSheets...................................................................................................................... 82
Page5of86
GeneralInformation
QuidelContactInformation
ContactQuidelTechnicalSupportfrom7:00a.m.to5:00p.m.PacificTime
Tel: 800.874.1517(intheU.S.);
858.552.1100(outsidetheU.S.)
Fax: 858.552.7905
Email: [email protected]orcontactyourlocaldistributor
Website: quidel.com
IfoutsidetheU.S.contactlocaldistributororoneoftheTechnicalSupportcenterslistedbelow.Youmay
alsocontactusatquidel.com.
Region
Phone EͲmailAddress
EuropeandMiddleEast +44.161.483.9032 [email protected]
AsiaPacific +61.7.3363.7711 [email protected]
Africa,RussiaandCIS +972.8.9429.683 [email protected]
LatinAmerica +57.2.6618797 [email protected]
Canada +1.613.271.1144 [email protected]

Page6of86
MainMenuStructure

RunTe st RecallRes ults
InstallNew
CODECHIP
Set
Parameters*
Delete
Results*
QC
Sample
Misc.
Test
QC
Device
Patient
Sample
Patient
Samples
Misc.
Test
Results
QC
Results
Last
Record
PrintAll
Results
Upload
Results
toLIS
x QCSampleResult s
x QCDeviceResults
x Reag ent LotsQC
x ArchiveLastTest
UserID RangesClock
Meter
Settings
Test
Settings
QC
Parameters
Bypass
x NewUserID
x UpdateUserID
x Delet eUserID
x UserIDList
x QCDeviceFreq.
x QCSampleFreq.
x NumberofControls
x NumQua ntStdDevs
Display
Settings
Comm
Settings
ID
Settings
x CharactersUserID
x CharactersPat.ID
x CharactersAux.ID
x Aux.IDEnable/Disable
x CharactersMisc.ID
x Bau dRate
x Pat.ResultApproval
x LISEnable/Disable
x AutoUpload
x Language
x PrintMode
x AutoPowerͲOff
x Displa yContras t
x PrinterContras t
*SupervisorCODECHIPOnly
Page7of86
LabelsandSymbols
Label/Symbol Description
Manufacturer
Invitrodiagnosticmedicaldevice
Consultinstructionsforuse
AuthorizedrepresentativeintheEuropeanCommunity
Temperaturelimitation
Humiditylimitation
Wasteelectricalandelectronicequip m ent(WEEE)
SerialNumber
CatalogNumber
Warning/Caution
Important

Page8of86
CEmark
Batchcode
Dateofmanufacture
UseͲbydate
Intendeduse
Upperlimitoftemperature
DonotreͲuse
ChinaRoHS
Patientnumber
Testdevice
Control

CalibrationVerification
Prescriptionuseonly
Contents/contains
Expectedvalues
Biologicalrisks
Thaw
Transferpipette
Mean
Standarddeviation
CODECHIPmodule
PrinterPaper
Peelopenhere

Addsamplehere
UseEDTAplasmasampleonly
Useurinesampleonly
UseEDTAwholebloodorplasma
sampleonly
Addsampleimmediatelyafter
openingfoilpouch
Page9of86
Warnings,PrecautionsandLimitations
TheQuidelTriageMeterPromustbeusedwithintheoperatingtemperaturerangerequiredbythe
specifictestdeviceproductbeingused.Refertothetestdevicepackageinsertfordetails.
OperatetheQuidelTriageMeterProonalevel,drysurfaceawayfromdirectsunlight.
TheQCDevice
islightͲsensitiveandshould bestoredinitsblackopaquecasewhennotinuse.
DoNOTwriteorplacealabelanywhereontheQCdevice,asitmayinterferewiththedevice
functionality.
DonotmovetheQuidelTriageMeterProwhileatestis
inprogress.
Ensureallsamplefluidshaveabsorbedintothetestdevicepriortorunningthetesttopreventinternal
contaminationofthemeter.
UseonlytheAC/DCpoweradapterprovidedwiththeQuidelTriageMeterPro.Anidentifyingtaghas
beenplacedonthecordfor
quickidentification.
DonotdropQuidel TriageMeterPro.
DonotplaceobjectsontheQuidelTriageMeterPro.
DonotimmersetheQuidelTriageMeterProinwateroranyliquids.
Page10of86
Introduction
HowtoUseThisGuide
Thismanualcontains:
InstructionsfortheoperationandmaintenanceoftheQuidelTriageMeterPro,and;
Basicinstructionsfortestingsamples
TheQuidelTriageMeterProsoftwarehasbeendesignedtoprovidetheuserstepbystepinstructions.The
meterdisplaystestresultsandmenuoptionsintheupperhalf
ofthemeterscreen.Insmallerlettersatthe
bottomofeachscreen,themeterdisplaysinstructionspertainingtothetaskbeingperformed.
8:23AM01•20•18
RUNTEST
RECALLRESULTS
INSTALLNEWCODECHIP
SELECTWITHvOR^ANDPRESSENTER
PRESSEXITTOEJECTDEVICE
TheQuidelTriageMeterProUserManualisdividedintosixsections.
Section1 GeneralInformation:Contactinformation,MenuStructurefortheQuidelTriageMeterPro,
Symbols,WarningsandPrecautions
Section2Introduction:ProvidesanoverviewoftheQuidelTriageMeterPro.
Section3 Installation:Guidestheuserthrough
firsttimesetup.
Section4 Operation:Guidestheuserthroughrunningtests,recalling resultsanddeletingresults.
Section5 ServiceandMaintenanceProcedures:PaperReplacement,Cleaning,Batteries,TotalQuality
Assurance,andWarranty.
Section6 Appendix:Specifications,Troubleshooting,ReturnandDisposal,Glossary,SampleLogSheets
Description
WhatistheQuidelTriageMeterPro?
TheQuidelTriageMeterProisaportablefluorescenceinstrumentusedtomeasuretheresultsoftests
manufacturedbyQuidelCardiovascularInc.TheQuidelTriageMeterProcanbeusedinalaboratoryorina
pointͲofͲcaresetting.
TheQuidelTriageMeterProusesalaserasalightsource.Light
fromthelaserhitsatestdevicethathas
beeninsertedinthemeter.Thiscauses thefluorescentdyeinthetestdevicetogiveoffenergy.Themore
energythefluorescentdyegivesoff,thestrongerthesignal.
NOTE:Torunspecifictests,suchastheQuidelTriageCardiacPanel,youalsowillneeddetailed
informationcontainedintheproductͲspecificPackageInsert,includedinthetestkit.
Page11of86
Overview:RunningaTest
Afteratestsample(forexample,blood)fromapatientisaddedtothetestdevice,thetestdeviceisinserted
intheQuidelTriageMeter P ro.Themetermeasureshowmuchofthesubstance(aparticularprotein
marker)ispresent,basedonstandardsthathavebeenpreͲprogra mmedintothemeter.
Thepatient
identification,theUserI.D.andthetestresultscanbeprintedoutonapapertape.
Inadditiontotestingsamplesfrompatients,theQuidelTriageMeterProcontainspreͲprogrammedQuality
Control(QC)functionsthatallowthepersonoperatingthemetertomonitorqualitycontrol.

Unpacking
TheQuidelTriageMeterProandtheitemsthatcomewithitareprovidedinasinglebox.
QuidelTriagetestdevicesarepackagedseparatelyandincludeinstructionsforrunningspecifictests.Ifyou
haveanyquestions,contactQuidel(refertotheAssistancesection).

TriageTestDevice.Atestdeviceholdsthe
sampleandalsocontainsthereagentsand
othermaterialsneededforthetest.
NOTE:TheQuidelTriageMeterProacceptstestdevicesthataredesignedspecificallyforusewiththe
QuidelTriageMeterPro.PleaserefertoproductͲspecificPackageInsertsincludedineachtestkitformore
information.
Page12of86
Contents
QuidelTriageMeterPro
QCDeviceandCODECHIPmodule(intheQCDevicebox)
AC/DCPowerConverter
AABatteries
RollsofPrinterPaper(additionalrollsaresuppliedineachboxoftestdevices)
SupervisorCODECHIPmoduleandCODECHIPmoduleBox

Printerpaper
locatedbelow
QCDevicebox.
Page13of86
PartsoftheQuidelTriageMeterPro
Page14of86
On/OffKey
(Powerbutton)
UsedtoturntheQuidelTriageMeterProonandoff.
PrintKey
Usedtoprintthetestresultsordisplayscreen.
DeleteKey
Usedtodeletedata.
EnterKey
Allowstheoperatortoselectmenuitemsandacknowledgealarm
conditions.
ExitKey
Usedtoexitthedisplayedmenuorejectthetestdevice.
BrightnessKey
Usedtosetscreenbrightness
Keypad
Usedtoenteridentificationnumbers,letters,andspecialsymbols.
ArrowKeys
Usedtotogglethroughmenuitems.
Up
Down
Left
Right
ShiftKey
SymbolKey
UsedwithShiftKey
ThermalPrinter
Printsthetestresultonpaper.
DataPort
Connectionpointforthe(optional)BarCodeScanner,allowingthe
operatortoscanpatientoruseridentificationdatainsteadofmanually
enteringnumbersorlettersonthekeypad.Mayalsobeusedtoconnect
toadatamanagementsystem/LIS.
PaperFeedKey
Usedtofeedpaper
CODECHIP
ModulePort
InsertionpointforCODECHIPmoduleswhichcontainlot/devicespecific
dataforuseinprovidingtestresults.
PowerSupplyPort ConnectionpointforthesuppliedAC/ DCPowerConverter.
PrinterCover Thecover,whichwhenpulledstraightup,revealsthepaperroll.
LCDScreen
TheLCD(LiquidCrystalDisplay)Screenshowsthemenuofpossibletests
andtasksandpromptstheoperatortotakethenextstep.
ScreencommandsandoptionsareindicatedbyaBOLDEDFONT.
Page15of86
Installation
UnpackMeter
a. Removethemeterfromtheboxandprotectiveplasticbag.
PowerMeterOn
Batteries
a. Turnthemeterover.
b. Removebatterycover.
c. Install4AAbatteries,payingattentiontobatteryorientationinthe
batterycompartment.
d. Replacebatterycoverandturnmeterrightsideup.
e. Powerthemeteronbypressingthekey.Pressthekey
torunselfͲtest.Whenthe
testiscompleted,thescreenmeter
displaywillcometorestatthemainmenu.
LowPowerIndicator
 8:23AM01•20•18
Timeanddate
Menuitems
UserInstructions
RUN
TEST
RECALLRESULTS
INSTALLNEWCODECHIP
SELECTWITHvOR^ANDPRESSENTER
PRESSEXITTOEJECTDEVICE
TheMainMenu
NOTE:TheLowPowerIndicatormonitorstheactivepowersource.Toverifyadequatebatterpower:
NavigatetotheMainMenu
RemovetheACpowersupply
PresstheEXITkey(themeterattemptstoejectthedevice)
ObservetheMainMenuonthemeter’sdisplay.
Replacethebatteriesifthepowericonisvisible.
Page16of86
ACPowerSupply
a. Removethepowersupplyfromthebox.
b. PlugoneendintoanACoutlet.
c. Plugtheoppositeendintotheroundholeinthebackofthemeter.
InstallPaper
a. Removetheprintercoverbypullinguponthecoverasindicatedbythearrowonthebackofthecover.
FormoredetailsaboutInstallingPaperseetheCareandMaintenancesection.
b. Tearorcutaclean,straightedgetofeedintotheprinter.Donot
cutpaperatanangle,astheprinter
mustsensetheedgeofthepaperalongthefeedpath.
c. Positionthepapersuchthatthepaperwillfeedfromundertheroll(asopposedtooverthetopofthe
roll,seepicture).


NOTE:Theprintercontainsapapersensorandwillfeedthenewpaperrollonlywhenpaperwithaclean
straightlineispressedintothepaperroller.
Page17of86
d. Insertthepaperedgeunderthepaperroller(platen)untilitfirmlyseatsorresistanceisfelt.
e. Pressthekeytwice,paperwillfeedthroughtheprinterandextendoutthemeter.
f. Placetherollofpaperintothepapercompartment.
g. Replacetheprintercoverandcontinue
operation.
InsertSupervisorCodeChipModule
a. RemovetheSupervisorCODECHIPmoduleBoxfromthemeterbox.
b. RemovetheSupervisorCODECHIPmodulefromitsbox.
c. InserttheSupervisorCODECHIPmoduleintotheCODECHIPmodulePort.Theportislocatedontheleft
sideofthemeter,towardsthebottomfrontcorner.
NotetheorientationoftheCODECHIPmodule in
thedrawingbelow.
d. OncetheSupervisorCODECHIPmoduleisinstalled,verifythattwoadditionalmenuitemsappearonthe
displayscreen.
8:23AM01•20•18
RUNTEST
RECALLRESULTS
INSTALLNEWCODECHIP
Twoadditionalmenuoptions
accessiblewhenSupervisor
SETPARAMETERS
CODECHIPmoduleisinserted.
DELETERESULTS
SELECTWITHvOR^ANDPRESSENTER
PRESSEXITTOEJECTDEVICE
TheSupervisor’sMainMenu
FormoredetailsaboutCODECHIPmoduleseepage22.
ToChangeMeterSettings
a. SelectSETPARAMETERSusingthe

keysandpressthekey.
b. SelectMETERSETTINGSusingthekeysandpressthekey.
Page18of86
c. SelectIDSETTINGS,DISPLAYSETTINGSorCOMMSETTINGSusing
the
keysandpressthekey.
d. Selectthedesiredsettingusingthe
keysandchangethevalueofthesetting
byusingthe
keys.
IDSettings Default Options NewSettings
CHARACTERSUSERID Min1
Max16
1Ͳ16
1Ͳ16
CHARACTERSPAT.ID Min1
Max20
1Ͳ20
1Ͳ20
CHARACTERSAUXID Min1
Max12
1Ͳ12
1Ͳ12
CHARACTERSMISCID
Min1
Max20
1Ͳ20
1Ͳ20
AUX.IDENABLE/DISABLE Disabled Disabled,Enabled
Forreference,recordyoursettingsinthespaceprovided.
e. PressthekeytosavechangestoIDSe ttings.
DisplaySettings Default Options NewSettings
LANGUAGE English English,French,Italian,
German,Spanish,
Portuguese,Greek,
Danish,Swedish,
Japanese
PRINTMODE Automatic Automatic,Manual
AUTOPOWERͲOFF 2hour 1/2hour,1hour,2
hour,4hour,None
DISPLAYCONTRAST 4 0Ͳ8
(8=Heaviestcontrast)
PRINTERCONTRAST 4 08
(8=heaviestcontrast)
Forreference,recordyoursettingsinthespaceprovided.

Page19of86
f. PresstosavechangestoDisplaySettings.
COMMSettings Default Options NewSettings
BAUDRATE
 
PAT.RESULTAPPROVAL
'LVDEOHG (QDEOHG'LVDEOHG
LISENABLE/DISABLE
'LVDEOHG 'LVDEOHG(QDEOHG
AUTOUPLOAD
'LVDEOHG 'LVDEOHG(QDEOHG
LISPASSWORD

g. PressthekeytosavechangestoCommunicationSettings.
h. PressthekeytosavechangestoMeterSettings.
FormoredetailsaboutCommunications,includingentryofaQuidelprovided,meterspecificpassword,see
page29.
ChangeClock
a. SelectSETPARAMETERSusingthe
keysandpressthekey.
b. SelectCLOCKusingthe
keysandpressthekey.
c. SelecttheHOUR:MINUTEareausingthe
keys.Typeinthecorrecttimeusingthenumber
keys.
d. PressthekeytomovetoAM
orPM.
e. SelectAMorPMusingthe
keys.
f. PressthekeytomovetoTIMEFORMAT.
g. SelectAM/PMor24HR.usingthe
keys.Ifselecting24HR.,thepreviouslyenteredHOUR:
MINUTEwillchangetoa24ͲhourformatandtheAM
orPMwilldisappear.
h. PressthekeytomovetotheDATE.
i. Typeinthe6ͲdigitdateaccordingtotheDATEFORMAT
belowit.
j. PressthekeytomovetotheDATEFORMAT.
k. SelectMMͲDDͲYY,DDͲMMͲYY
orYYͲMMͲDDusingthe
keys.ThepreviouslyenteredDATEchangesas
theDATEFORMATchanges.
l. Pressthekeytosavechanges.
Page20of86
Setting Default Options NewSettings
TIMEFORMAT AM/PM AM/PM;24HR.
DATEFORMAT MMͲDDͲYY MMͲDDͲYY,
DD.MM.YY,
YYͲMMͲDD
SetUserID
TheQuidelTriageMeterPronormallyrequiresaUserIDtobeenteredpriortorunningapatienttest,QC
test,orMisctest,andpriortorecallingpatientresults.Refertopage31forinstructions.Alternately,this
featuremaybebypassed.Seepage21.
ChangeReferenceRanges/ThresholdsforTests:
a. SelectSETPARAMETERSusingthe
keysandpressthekey.
b. SelectRANGESusingthe
keysandpressthekey.
c. Selectthetestpaneltypeusingthe
keys.
d. Selecttheanalyteandlevelusingthe
keys.TypeinthenewvaluefortheHIGH
orLOWcutoff
asappropriate.
e. Pressthekeytosavechanges.
FormoredetailsaboutRangesSettingsseepage33.
ChangeTestSettings
a. SelectSETPARAMETERSusingthe
keysandpressthekey.
b. SelectTESTSETTINGSusingthe
keysandpressthekey.
c. Selectthetestpaneltypeusingthe
keys.
d. Selecttheanalyteusingthe
keys.
e. Selectthetestsetting(ACTIV,INACT,USRA,orUSRI)usingthe
keys.
f. Pressthekeytosavechanges.
FormoredetailsaboutTestSettingsseepage34.
NOTE:ThelowestvalueoftheRangeorThresholdmaybedisabledforsometestpanels.Refertothe
appropriateProductInsertforspecificinformation.
NOTE:Theabilitytochangetestsettingsmaybedisabledforsometestsinspecificpanels.Refertothe
appropriateProductInsertforspecificinformation. 
Page21of86
ChangeQCParameters
a. SelectSETPARAMETERSusingthe
keysandpress .
b. SelectQCPARAMETERSusingthe
keysandpress .
c. Selectthedesiredsettingusingthe
keysandchangethevaluetotherightofthesettingby
usingthe
keys.
Setting Default Options NewSettings
QCDEVICEFREQ. Daily None,8HR,Daily,
Weekly,Monthly
QCSAMPLEFREQ. Monthly None,8HR,Daily,
Weekly,Monthly
NUMBEROFCONTROLS 2 1or2
NUMQUANTSTDDEVS 3 2or3
d. PressthekeytosavechangestoQCPARAMETERS.
FormoredetailsaboutQCParametersSettingsseepage35.
ChangeBypassSettings
a. SelectSETPARAMETERSusingthe
keysandpressthekey.
b. SelectBYPASSusingthe
keysandpressthekey.
c. ForUSERIDselectONorOFFusingthe
keys.
d. PressthekeytosavechangestotheBypasssetting.
Formoredetailsabout
BypassSettingsseepage
36.
Priortocommencingpatienttesting,removetheSupervisorCODECHIPmodulefromthemeter,returnit
tothestorageboxandplaceinasafeplaceforfutureuse.
ThemeterisnowreadytorunQCTestsinpreparationforPatientTesting.
QCTestingshouldbeperformedinthe
followingmanner:
1. RunQCDEVICE(seepage40).
2. RunQCSAMPLEforeachlotofdevices(seepage42).
3. Ifappropriate,runCALIBRATIONVERIFICATIONSETasaMiscTestsampleforeachtestpaneltypetobe
used.(RefertotheapplicableProcedureManualandCalibrationVerification
SetPackageInsertfor
detailedinstructions).
NOTE:

Ifthebypassis
ON,
theuserIDfeatureisdisabledand anyonemayrunatest.Ifthebypassis
OFF,
theuserID
featureisactiveandonlyauthorizedusersmayrunatest.ToInstallUserID's,seepage31.
Page22of86
Location
TheQuidelTriageMeterProshouldbeplacedonadry,clean,flat,horizontalsurfaceawayfromdirect
sunlight.
Power
TheQuidelTriageMeterProrequires6ͲvoltsDCsuppliedbyeitherACorDC.Themeterwill run
approximately100testsonbatteries.Itisrecommendedtousebatteriesasabackupintheeventofa
poweroutage.
CAUTION:Useonlythepowersupplyprovidedwiththemeter,failuretodosomaycausepermanentdamageto
themeter.
SelfͲTest
TheQuidelTriageMeterProchecksthesystemtobesureit isoperatingproperlybyrunningselfͲtestswhen
poweredonandpriortoeachtest.
PowerOnincludeslaseroperation,internalstandardrange,batterypower,andsoftwareverification.
EachTestverifiesthelaseroperation,internalstandard
rangeandbatterypower.
CODECHIPModules
MeterdataisupdatedviaadisposableCODECHIPmodule.TheCODECHIPmodulecontainsmicrochip
circuitryembeddedintoaplastichousing.WheninsertedintothemeterCODECHIPmoduleportand
activated,theinformationistransferredintotheQuidelTriageMeterPro’smemory.ACODECHIPmodule
typicallyneedstobe
installedonlyonceandremainsinthemetermemory.ACODECHIPmoduledoesnot
needtoremaininthemeterwhileperf ormingtests.
TherearefivetypesofCODECHIPmoduleswhichcanbedistinguishedbyshapeorcolor.
StandardCODECHIP
module
ReagentTestCODECHIPmoduleincludedineach boxoftestdevices;containscalibration,expirationdate
andotherdataaboutthedevicelot.(Colorvariesbytesttype.)
QCSampleCODECHIPmoduleincludedineachboxofQCSample;containsexpirationdateandother
data
abouttheQCSamplelotincludingacceptablerange.(White.)
QCDeviceCODECHIPmoduleincludedintheblackQCDevicebox(comeswiththeQuidelTriage
MeterPro);containsdataabouthowthemetershouldreaditsmatchingQCDevice.(Darkgray.)
Supervisor AccessCODECHIPmoduleincluded
witheachQuidelTriageMeterPro;containscodeto
activateSupervisorfunctions.(Lightgray.)
NOTE:TheQuidelTriageMeterProwillautomaticallyturnoffifleftunusedfor2hours,unless
programmedotherwisebythesupervisor.Optionalsettingsallowautomatedturnoffafter30minutes,
1hour,2hours,4hoursorNONE.
Page23of86
SoftwareUpgradeCODECHIPmoduleprovidedwhenQuidelTriageMeterProsoftwareupgradesare
available;containsnewsoftwareforthemeter.
SetParameters
Purpose
TheSETPARAMETERSfunctionallowsthesupervisortoselectanumberofsettings.Accesstothesesettings
iscontrolledusingtheSupervisorCODECHIPmodule.
ListofProgrammableParameters
Asupervisorcansetthefollowingparameters:
Heading Parameters
IDSETTINGS NumberofCharactersinUserID,NumberofCharactersinPatient
ID,NumberofCharactersinAuxiliaryID,EnableorDisable
AuxiliaryID,NumberofCharactersinMiscTestID
DISPLAYSETTINGS Language,PrinterMode,AutoPowerͲOff,DisplayContrast,
PrinterContrast
COMMUNICATIONS BaudRate,EnableorDisableLIS,EnableorDisableAutoUpload,
EnableorDisablePatientResultApproval
CLOCK
Time,Date,andDisplayFormat
USERID
AddNewUserID,UpdateUserID,DeleteUserIDorUserIDList
RANGES
TestCutoffs
TESTSETTINGS
BlockAnalytes
QCPARAMETERS MinimumFrequencyforQCTests,NumberofControls,Number
ofStandardDeviations(QuantitativeTests)
BYPASS
DisableUserIDrequirement
SupervisorAccess
ASupervisorCODECHIPmoduleisshippedwitheachmeter.Installing
theCODECHIPmodulegivesthesupervisoraccesstotheSET
PARAMETERSandDELETERESULTSfunction.
Page24of86
ToinserttheSupervisorCODECHIPmoduleintotheQuidelTriageMeterPro,slidethechiplabeled"SPR"
intothemeter’sCODECHIPmoduleportasshowninthepicture.
HowtoSetParameters:BasicInstructions
Allinstructionsassume:
Themeterison.
Themeterscreenisdisplayingthemainmenu.
TheSupervisorCODECHIPmoduleisinstalled.
1. SelectSETPARAMETERSusingthe
keys.
2. Pressthekey.
3. Youwillseealistofprogrammableparameters.
METERSETTINGS
CLOCK
USERID
RANGES
TESTSETTINGS
Q
CPARAMETERS
BYPASS
SELECTWITHvOR^ANDPRESSENTER
ORPRESSEXI
T
TheSetParametersMenu
4. Usethe
keystoselecttheparameteryouwanttoprogram:
METERSETTINGS,CLOCK,USERID,RANGES,TESTSETTINGS,QCPARAMETERS,
orBYPASS.
5. Pressthekey.
6. Followthedirectionsforsettingtheselectedparameter(seethenextsection).


NOTE:Tosavechanges,presstheENTERkey.Tocancelchanges,presstheEXITkey.Wheneitherthe
ENTERortheEXITkeysarepressed,themeterperformstheappropriatefunctionandreturnstothe
previousmenu.
Page25of86
MeterSettings
Thismenuwillallowyoutosetbasicoperationalcriteriaforthemeter:language,
displaycontrast,maximumlengthoftheuser,patient, auxiliaryandmisc.ID's,
printermodeandautomaticpoweroff.
IDSettings
CharactersUserID
Thissettingwillsettheminimumandmaximumnumberofcharactersin
theUserID.
1. Fromalistofprogrammableparameters,chooseINSTRUMENTSETTINGSusingthe
keys.
2. Pressthekey.
3. Usethe
keystoselectIDSETTINGS.
4. Pressthekey.
5. Usethe

keystoselectthenumbertotherightofCHARACTERSUSERIDundertheMIN
column.
6. Usethe
keystochoosethesmallestnumberofcharactersthatwillappearintheUserID.The
minimumnumberofcharacterspossibleis1.
7. UsethekeytoselectthenumbertotherightofCHARACTERSUSERIDundertheMAXcolumn.
8. Usethe
keystochoosethelargestnumberofcharactersthatwillappearintheUserID.The
maximumnumberofcharacterspossibleis16.
9. Pressthekeytosavechanges.
CharactersPat.ID
ThissettingwillsettheminimumandmaximumnumberofcharactersinthePatientID.
1. Fromalistofprogrammableparameters,chooseINSTRUMENTSETTINGSusingthe

keys.
2. Pressthekey.
3. Usethe
keystoselectIDSETTINGS.
4. Pressthekey.
5. Usethe
keystoselectthenumbertotherightofCHARACTERSPAT.IDundertheMINcolumn.
6. Usethe
keystochoosethesmallestnumberofcharactersthatwillappearinthePatientID.
Theminimumnumberofcharacterspossibleis1.
7. UsethekeytoselectthenumbertotherightofCHARACTERSPAT.IDundertheMAXcolumn.
8. Usethe
keystochoosethelargestnumberofcharactersthatwillappearinthePatientID.The
maximumnumberofcharacterspossibleis20.
MainMenu
SetParameters
MeterSettings
IDSettings
MainMenu
SetParameters
Meter Settings
IDSettings
Page26of86
MainMenu
SetParameters
MeterSettings
IDSettings
9. Pressthekeytosavechanges.
CharactersAux.ID
ThissettingwillsettheminimumandmaximumnumberofcharactersintheAuxID.
IfthesettingofA
UX.IDENABLE/DISABLEisENABLED,anAuxiliaryIDismandatoryfor
eachpatienttest.
1. Fromalistofprogrammableparameters,chooseMETERSETTINGSusingthe
keys.
2. Pressthekey.
3. UsethekeystoselectIDSETTINGS
4. Pressthekey.
5. UsethekeystoselectthenumbertotherightofCHARACTERSAUX.IDundertheMIN
column.
6. Usethe keystochoosethesmallestnumberofcharactersthatwillappearinthe AuxiliaryID.
Theminimumnumberofcharacterspossibleis1.
7. UsethekeytoselectthenumbertotherightofCHARACTERSAUX.IDundertheMAXcolumn.
8. Usethe keystochoosethelargestnumberofcharactersthatwillappearintheAuxiliaryID.
Themaximumnumberofcharacterspo ssibleis12.
9. Pressthekeytosavechanges.
Aux.IDEnable/Disable
ThissettingwillturnonorturnofftheAuxiliaryIDfunction.TheAuxiliaryIDmaybe
usedforrecordingatestordernumber,physicianIDorotherspe cializedidentifying
information.
1. Fromalistofprogrammableparameters,chooseMETERSETTINGSusingthekeys.
2. Pressthekey.
3. UsethekeystoselectIDSETTINGS.
4. Pressthekey.
 Usethe
keystoselectthewordtotherightofAUX.IDENABLE/DISABLE.
 Usethe
keystochooseDISABLEDorENABLED.IfDISABLEDisselected,theAuxiliaryID
promptscreenswillneverappear.IfENABLEDisselected, anAuxiliaryIDismandatoryforeachpatient
test.
7. Pressthekeytosavechanges.
NOTE:
TheAuxiliaryIDis shownonly ontheNewResultdisplayorprintout.Itisnotvisibleinrecalled
results.Trackingmaybemanagedthroughtheoptionalsoftwarepackage,QuidelTriageCensus.
MainMenu
SetParameters
MeterSettings
IDSettings
Page27of86
CharactersMisc.ID
ThissettingwillsettheminimumandmaximumnumberofcharactersintheMisc.TestID.

Fromalistofprogrammableparameters,chooseINSTRUMENTSETTINGSusingthekeys.
2. Pressthekey.
3. UsethekeystoselectIDSETTINGS.
4. Pressthekey.
5. UsethekeystoselectthenumbertotherightofCHARACTERSMISC.IDundertheMin
column.
6. UsethekeystochoosethesmallestnumberofcharactersthatwillappearintheMISC.TEST
ID.Theminimumnumberofcharacterspossibleis1.
7. UsethekeytoselectthenumbertotherightofCHARACTERSMISC.IDundertheMAXcolumn.
8. UsethekeystochoosethelargestnumberofcharactersthatwillappearintheMISC.TESTID.
Themaximumnumberofcharacterspossibleis20.
9. Pressthekeytosavechanges.
DisplaySettings
Language
Thissettingwilldeterminewhichla nguagethemeterdisplaysandprints.
1. Fromalistofprogrammableparameters,chooseMETERSETTINGSusingthekeys.
2. Pressthekey.
3. UsethekeystoselectDISPLAYSETTINGS.
4. Pressthekey.
5. Usethekeystohighlightthenameto
therightofLANGUAGE.
6. Usethekeystochange thelanguageoftheQuidelTriageMeterP ro’sscreenandprintouts.
7. Pressthekeytosavechanges.
PrintMode
Thissettingwilldetermineifthemeterprintsautomaticallyoronlywhenrequested.
1. Fromalistofprogrammableparameters,
chooseMETERSETTINGSusingthekeys.
2. Pressthekey.
MainMenu
SetParameters
Meter Settings
DisplaySettings
Page28of86
3. UsethekeystoselectDISPLAYSETTINGS.
4. Pressthekey.
5. Usethekeystoselecthighlightthewo rdtotherightofPRINTMODE.
6. UsethekeystochooseAUTOMATICorMANUAL.IfAUTOMATICisselected,themeterwill
immediatelyprintouttheresults after
eachtest(patient,QCorMiscTest).IfMANUALisselected,the
usermustusethebuttononthekeypadtoprintoutresults.
7. Pressthekeytosavechanges.
AutoPowerOff
Thissettingwilldeterminethelengthoftimebeforeaninactivemeter
willpoweroff.
1. Fromalistofprogrammableparameters,chooseMETERSETTINGSusingthekeys.
2. Pressthekey.
3. UsethekeystoselectDISPLAYSETTINGS.
4. Pressthekey.
5. Usethekeystoselectthevaluetothe right ofAuto PowerͲOff.
6. Usethekeystochoose
thedesiredtimersettingforthemetertopoweritselfoffwhennotin
use.Thechoicesare:1/2HOUR,1HOUR,2HOURS,4HOURS orNONE.IfNONEisselected,themeter
mustbemanuallyshutoffusingtheON/OFFkey.
7. Pressthekeytosavechanges.
.
DisplayContrast
Thissettingwilladjustthedisplaycontrast.
1. Fromalistofprogrammableparameters,chooseMETERSETTINGSusingthekeys.
2. Pressthekey.
3. UsethekeystoselectDISPLAYSETTINGS.
4. Pressthekey.
5. UsethekeystoselectthenumbertotherightofContrast.
6.
Usethekeystochange thecontrastofthemeter’sscreen.Therangeisfrom08(8beingthe
greatestcontrast).
7. Pressthekeytosavechanges.
NOTE:
Whenpoweredonlybybatteries,select1/2hourtoconservebatterylife.Certainscreens,whendisplayed,
willnotautomaticallypoweroff.TheNewResultsscreenisanexample.
MainMenu
SetParameters
Meter Settings
DisplaySettings
Page29of86
PrinterContrast
Thissettingwilladjusttheprintercontrastformeterswithaserialnumber
29901orgreater.
1. Fromalistofprogrammableparameters,chooseMETERSETTINGSusingthekeys.
2. Pressthekey.
3. UsethekeystoselectPRINTERCONTRAST.
4. Pressthekey.
5.
UsethekeystoselectthenumbertotherightofContrast.
6. Usethekeystochangethecontrastofthemeter’sscreen.Therangeisfrom0Ͳ8(8beingthe
greatestcontrast).
7. Pressthekeytosavechanges.
CommunicationSettings
PriortoactivatingtheLISfeature,obtainthemeter’sserialnumberlocated
underneaththemeterandcontactQuidel(RefertotheAssistancesection).
TheQuidelRepresentativewillprovideyouwithapasswordthat,whenentered,
willallowdatatobetransmitted.
BaudRate
1.
1.
1.
1. Froma
listofprogrammableparameters,chooseMETERSETTINGSusingthekeys.
2. Pressthekey.
3. UsethekeystoselectCOMMSETTINGS.
4. Pressthekey.
5. UsethekeystoselectBAUDRATE.
6. Usethekeystochoose9600or38400.
7. Pressthekey.
PatientResult
Approval
Someinstitutionsmaywishforpatienttestresultstobeverifiedbytheuser
beforeprintingorsendingtotheLIS.Activationofthisfeaturedisplaystheresult,
thenrequirestheusertoacceptorrejecttheresultbeforetherecordisprinted
andsaved.Ifthetestisrejected,the
meterwillpermanentlyflagitasrejected
bytheuser.
1. Fromalistofprogrammableparameters,chooseMETERSETTINGSusingthe

keys.
2. Pressthekey.
MainMenu
SetParameters
Meter Settings
CommSettings
MainMenu
SetParameters
Meter Settings
DisplaySettings
MainMenu
SetParameters
MeterSettings
CommSettings
Page30of86
3. Usethe
keystoselectCOMMSETTINGS.
4. Pressthekey.
5. Usethe
keystoselectPAT.RESULTAPPROVAL.
6. Usethe
keystochooseDISABLEDorENABLED.
7. Pressthekey.
Whenenabled,afterapatient resultisavailableanddisplayed on the
meterscreen,theusermustacceptor
rejecttheresultbeforeprinting,uploadingorperforminganyothermeterfunctions.
ThemeterpromptstheusertoPRESS 1TOACCEPTOR0TOREJECT.Oncethetesthasbeenacceptedor
rejected,themeteroperatesnormally.
Ifthetesthasbeen
rejected,thedisplayandprintoutwillbe flaggedwithPAT.RESULTREJECTEDor
RESULTSREJECTEDBYUSER.Re calledresultswilldisplayanRbeforethepatientIDintherejectedrecord.
LISEnable
1. Fromalistofprogrammableparameters,chooseMETERSETTINGS
usingthe
keys.
2. Pressthekey.
3. Usethe
keystoselectCOMMSETTINGS.
4. Pressthekey.
5. Usethe
keystoselectLIS.
6. Usethe
keystochooseDISABLEDorENABLED.
7. Pressthekey.
8. IfDISABLEDwasselected,ascreenshowingaLISDISABLED
willappear.Press
thekey .
9. IfENABLEDwasselected,ascreenshowingaENTERLISPASSWORD willappear.Enterthepassword
providedbytheQuidelRepresentative.Pressthekey.
10.AcknowledgetheLISENABLEDmessagebypressingthekey.
AutoUpload
(ThisfeatureisvisibleaftertheLIShas
beenenabled)
1. Fromalistofprogrammableparameters,chooseMETERSETTINGSusingthe
keys.
2. Pressthekey.
3. Usethe
keystoselectCOMMSETTINGS.
4. Pressthekey.
5. Usethe
keystoselectAUTOUPLOAD.
6. Usethe
keystochooseENABLED
orDISABLED.
7. Pressthekey.
MainMenu
SetParameters
MeterSettings
CommSettings
Page31of86
Clock
Theclockmenusetsthetimeanddateofthemeterandtheformatfordisplaying
both.Ifthemeterhaslostpower,thetimeanddatewillblinkontheMainMenu.
1. Fromthelistofprogrammableparameters,chooseCLOCKusingthe
keys.
2. Pressthekey.
3. Usethe
keystoselecttheHOUR:MINUTEarea.Usingthekeypad,typeinthecorrecttime.
Thenumbersyoutypewillappearfromtherightandmoveleftasyoucontinuetotype.
4. PresstomovetoAMorPM.Ifthe24HR.formathasbeenpreviouslyselected,theAMorPMwill
notbevisableandtheTIMEFORMATwillbehighlighted.Proceedtostep7below.
5. SelectAMorPMusing the
keys.
6. PresstomovetoTIMEFO RMAT .
7. Usethe
keystoselectAM/PMor24HR.Ifselecting24HR.,thepreviouslyenteredHOUR:
MINUTEwillchangetoa24Ͳhourformatand theAMorPMwilldisappear.
8. PresstomovetotheDATE.
9. Usingthekeypad,typeinthe6ͲdigitdateaccordingtotheDATEFORMATbelowit.
10. PresstomovetotheDATEFORMAT.
11. Usethe
keystoselectMMͲDDͲYY,DDͲMMͲYYorYYͲMMͲDD.ThepreviouslyenteredDAT E
changesastheDATE
FORMATchanges.
12. Presstosavechanges.
UserID
Upto600UserID'swith116charactersmaybeprogrammedintothemeter.
TheUserIDmayalternatelybeenteredusingtheexternalBarCodeScanner
(optional).
AnadditionalsecurityfeatureoftheUserIDisthepartialmaskingoftheIDondisplaysandprintouts.This
prevents
unauthorized individualsfromdetectingandusinganexistingUserID.WhentheSupervisorCODE
CHIPmoduleisinstalledinthemeter,thisfeatureisoverriddenandallUserIDcharactersaredisplayed.
NewUserID
1. Fromthelistofprogrammableparameters,chooseUSERIDusingthe
keys.
2. Pressthekey.
3. SelectNEWUSERIDusingthe
keys.
4. Pressthekey.
MainMenu
SetParameters
UserID
MainMenu
SetParameters
Clock
NOTE:
IfyouhaveselectedAM/PM,AMorPMappearsafterthetime,asappropriate.
Page32of86
5. TypeinaUserIDwithanumberofcharacterswithinMINCHARACTERS USERIDandMAXCHARACTERS
USERID.Tocorrect,presstocleartheentireID,orselecttheincorrectcharacterbyusingthe
keys.Thentypeinthecorrect ID.
6. Pressthekey.
7.
Onthenextscreen,usethe
keystoselectanexpirationdatetotherightofUSERIDSTATUS:6
MONTHS,12MONTHS,VALIDorEXPIRED.
8. Pressthekeytosavechanges.
TheQuidelTriageMeterPr orecognizes0’sasauniquecharacter.Therefore,‘4341’will
berecognized
separatelyfrom‘04341’.
Usingthebarcodescanneroralphamode(shiftkey)willallow the QuidelTriageMeterProtoacceptalpha
charactersandthefollowingsymbols:#()*Ͳ./\
UpdateUserID
1. Fromthelistofprogrammableparameters,chooseUSERIDusingthe
keys.
2. Pressthekey.
3. SelectUPDATEUSERIDusingthe
keys.
4. Pressthekey.
5. TypeintheuserIDyouwanttoupdate.Tocorrect,presstocleartheentireID,or
selectthe
incorrectcharacterbyusingthe
keys.ThentypeinthecorrectID.
6. Pressthekey.
7. TheuserIDisdisplayedwiththeexpirationdate.Youcanchangetheexpirationdateto6MONTHS,12
MONTHS,EXPIREDorVALIDbyusingthe
keys.VALIDreconfirmsthe
previouslyselected
interval.
8. Pressthekeytosavechanges.
DeleteUserID
 Fromthelistofprogrammableparameters,chooseUSERIDusingthe
keys.
 Pressthekey.
MainMenu
SetParameters
UserID
NOTE:
UserID9999999999(10Ͳ9’s)isadefaultsettingdisplayedwhentheUserIDBypassisonorwhena
testisrunwiththeSupervisorCODECHIPmoduleinstalled.
NOTE:
IfyoutypeinaUserIDthathasalreadybeenassigned,themessageonthescreenstates:
USER###ALREADYEXISTS.PressENTERtogobackandtypeinadifferentID
NOTE:
IfyoutypeinaUserIDthatdoesnotexist,themessageonthescreenstates:USERID###DOES
NOTEXIST.
Ifthishappens,press
ENTER
toreturntothepreviousscreenandcorrecttheID,orpresstheEXITkey.
Page33of86
 SelectDELETEUSERIDusingthekeys.
 Pressthekey.
 Type intheuserIDtobedeleted.Tocorrect,press tocleartheentireID,orselecttheincorrect
characterbyusingthekeys.ThentypeinthecorrectID.
 Pressthekey.Themessageonthescreenstates:USERID###WILLBEDELETED.
 Pressthekeytodelete.Amessageonthescreentel lsyouthattheIDhasbeendeleted:USERID
###DELETED.
UserIDList
1. Fromthelistofprogrammableparameters,chooseUSERIDusing
the
keys.
2. Pressthekey.
3. SelectUSERIDLISTusingthe
keys.
4. Pressthekey.
5. Alistisdisplayed.
6. Usethe
keystoselectthetypeofIDlistyouwant:USERID,FROMEXPIRATIONDATE,TILL
EXPIRATIONDATE.
Fora listofallID's,selectUSERID.
Fora singleID,selectUSERIDandtypeintheID.
Tocorrect,presstocleartheentireID,or
selecttheincorrectcharacterbyusing the
keys.ThentypeinthecorrectID.
Fora range ofID's,entertheFRO M and/ortheTOdates.Thedatesareenteredbytypingthemonth
number,thedaynumberand
theyearnumberintheformatyouhavechosen.Thenumbersyou
typewillappearfromtherightandmoveleftasyoucontinuetotype.Tocorrect,pr ess
toclear
theentiredate.Thentypeinthecorrectdate.
7. Pressthekey.
8. Toprintthelistyou
havechosen,press .
Ranges
Dependingonthepaneltype,therangeonthedisplayandprintoutwillbe
titledREFERENCERANGESorTHRESHOLDS.Thelowend,whenthetestdisplays
thresholds,issetto‘0’andisunchangeable.
Someanalytesaredesignedforthreedistinctranges.Themeterdistinguishesbetweentworangeandthree
range
analytesinthesetupscreenbydifferentiatingthesymbolbetween theloweranduppervalues.
AnalyteswithadashTworanges.Thenormalrangeisanyvaluebetweenthetwonumbersdisplayed.
Dependingontheproduct,thenormalrangemaybeinclusiveorexclusiveoftheupper
number.Consult
NOTE
:

IfyoutypeinaUserIDthatdoesnotexist,themessageonthes creenstates:USERID###DOESNOT
EXIST.
Ifthishappens,press
ENTER
toreturntothepreviousscreenandcorrecttheID, orpresstheEXITkey.
MainMenu
SetParameters
UserID
MainMenu
SetParameters
Page34of86
theproductpackageinsertorcontactQuidelforclarification.Thesevalueswillbeinreversevideoon
thepatientresults.
AnalyteswithacommaThreerangespossible.Ifthelowernumberis0.0,themetertreatsthe
analyteashavingonlytworanges.If the lowernu mberisnot
0.0,thenormalrangeisanyvalueless
thanthelowernumber.Thefirstabnormalrangeisanyvaluebetweenthetwonumbersandisinclusive
ofthesenumbers.Thesevalueswillbeboxedonpatientresults.Thesecondabnormalrangeisanyvalue
greaterthantheuppernumber.These
valueswillbeinreversevideoonpatientresults.Consultthe
productpackageinsertorcontactQuidelforclarification.
1. Fromthelistofprogrammableparameters,chooseRANGESusingthe
keys.
2. Pressthekey.
3. Selectthetestpaneltypeusingthe
keys.
4. Selecttheanalyte
andlevelusingthe
keys.TheHIGHandLOWvalueforeachoftheanalytes
canbechangedusingthekeypadtotypeinthevalues.Tocorrect,presstocleartheentirenumber.
Thentypeinthecorrectnumber.
5. Pressthekeytosavechanges.
TestSettings
Incertaincases,thelabsupervisormaywishtocontro lwhichtestsareavailable
toberunatthetimeofpatienttesting.Theoptionsare:
1. ACTIVE(ACTIVONSCREEN):
Thetestisalwaysrunandcannotbedeselected.
2. INACTIVE(INACTONSCREEN):
Thetestisneverrunandisnotdisplayed.
3. USERINACTIVE(USRIONSCREEN):
Thetestcanbeselectedbytheuseratthetimeofrunningthetest.
4. USERACTIVE(USRAONSCREEN):
Thetestcanbedeselectedbytheuseratthetimeofrunningthetest
Iftheseoptionsarenotavailableforatest,themeterskipsoverthesettingasthesupervisorscrollsthrough
theoptions.Seepage45formoredetails.
1.
1.
1.
1.
1.
1.
1.
1.
1.
1.
1.
1.
1.
1.
1.
1. Fromthelistofprogrammableparameters,chooseTESTSETTINGSusingthe
keys.
2. Pressthekey.
3. Selectthetestpaneltypeusingthe
keys.
4. Selecttheanalyteusingthe
keys.
5. Select
thedesiredmodeusingthe
keys.
ACTIVE(ACTI VON
SCREEN)Thetestwillrunoneverypatientspecimen.
INACTIVE(INACTONSCREEN)Thetestwillneverrunonapatientspecimenandisdroppedfrom
alldisplaysandprintouts.
USERINACTIVE(USRIONSCREEN)
Atthetimeofpatienttesting,theuserhastheabilitytoselect
thetest.
NOTE:
Thefieldwillbedisabledifthepanelsettingisfixed.Consultthespecificpanelpackageinsertto
determineifthesettingmaybechanged.
MainMenu
SetParameters
Page35of86
USERACTIVE(USRAONSCREEN)Atthetimeofpatienttesting,theuserhastheabilityto
deselectthetest.
6. Pressthekeytosavechanges.
QCTestsWithTestSettingsActivated
AnytestwhichtheSupervisorhassettoACTIVE,USERACTIVEorUSERINACTIVEwillrunontheQCSample
panel.Themeterwilldisplayandprintthetest’sQCsampleresultandupdatetheQCSampletimerforthe
test.IftheSupervisorhassetthetesttoINACTIVE,
thetestwillhavetheQCSampleturnedOFF.TheQC
Sampleresultwon’tshowforthattestandtheQCSampletimerwon’tbeupdatedforit.
IftheSupervisorchangesthetestfromINACTIVEtoACTIVE,toUSERACTIVEortoUSERINACTIVEthattest
maybeout
ofQCSampledate.Asalwaysforatestwithanexpir edQCSample,apassingQCSampletestwill
needtorunbeforeapatientsamplecanreportaresultforthattest.
QCParameters
TheQCParametersmenuprovidestheusertheoptionsfordeterminingwhatthe
maximumfrequencyisforrunningtheQCDeviceandQCSample,whether1or2
levelsofcontrolarerequiredfortheQCSample,andwhethertheallowedQC
SampleRangeofquantitativetestsis2or3
standarddeviations.
QCDeviceFrequency
TheQCDEVICEFREQUENCYsettingdeterminesthemaximumintervalbetweenQCDeviceteststhatauser
isallowedtorunpatienttests.Whentheintervalhaslapsed,allusersarelockedoutoftheRunTestmenu
untilaQCDevicehasbeensuccessfullyrun.
ItisrecommendedthattheQCDeviceberundailywhenperformingpatienttesting.
1. Fromthelistofprogrammableparameters,chooseQCPARAMETERSusingthe
keys.
2. Pressthekey.
3. Usethe
keystoselecttheinformationtotherightofQCDEVICEFREQ.
4. Usingthe
keys,selectoneoftheseoptions:NONE,8HOURS,DAILY,WEEKLY,MONTHLY.
5. Pressthekeytosavechanges.
QCSampleFrequency
TheQCSAMPLEFREQUENCYsettingdeterminesthemaximumintervalbetweenQCSampleteststhatauser
isallowedtorunpatienttestsonapar ticulardevicelotnumber.Whentheintervalhaslapsed,allusersare
lockedoutoftheRUNPAT IENT TESTmenuuntilaQC
Samplehasbeensuccess fullyrun.Toviewwhena
particulardevicelot’sQCSamplewillexpire,usetheRECALLREAGENTLOTSQCfeature.
ItisrecommendedthattheQCSampleberunwitheverynewshipment,newlotormonthly,whicheveris
lesswhenperformingpatient testing.
NOTE:
Whentheanalyteisdeselected,thevalueis notcalculatednorstored
MainMenu
SetParameters
QCParameters
Page36of86
1. Fromthelistofprogrammableparameters,chooseQCPARAMETERSusingthe
keys.
2.
Pressthekey.
3. Usethe
keystoselecttheinformationtotherightofQCSAMPLEFREQ.
4. Usingthe
keys,selectoneoftheseoptions:NONE,8HOURS,DAILY,WEEKLY,MONTHLY.
5.Pressthekeytosavechanges.
NumberofControls
 Fromthelistofprogrammableparameters,chooseQCPARAMETERS
usingthe
keys.
2.
Pressthekey.
3. UsethekeystoselecttheinformationtotherightofNumberofControls.
4. Usingthekeys,selectoneoftheseoptions:1or2.
5. Pressthekeytosavechanges.
NumberofControlStandardDeviationsforQuantitativeTests
1. Fromthelistofprogrammableparameters,chooseQCPARAMETERS
usingthe
keys.
2. Pressthekey.
3. Usethe
keystoselecttheinformationtotherightofNUMQUANTSTDDEVS.
4. Usingthe
keys,selectoneoftheseoptions:2or3.
5. Pressthekeytosavechanges.
BypassUserID
Thisfunctionallowsthesupervisortose lectively bypasstherequirementtoenter
auserIDbeforeperforminganytest.
1. Fromthelistofprogrammableparameters,chooseBYPASSusingthe
keys.
2. Pressthekey.
3. SelectUSERIDusingthe
keys.
4. Usethe
keystoselectfromtwooptions:ONorOFF.
5. Pressthekey.
MainMenu
SetParameters
QCParameters
NOTE:
IfthebypassisON,theUserIDrequirementis disabled andthereforenotrequired.Whenthe
bypassisOFF,theUserIDrequirementoperatesasnormalandisrequired.
MainMenu
SetParameters
Bypass
Page37of86
InstallCodeChip
Purpose
TheQuidelTriageMeterPr owillprompt theusertoinstallaCODECHIPmodule whenrequired.However,
theINSTALLCODECHIPfunctionmaybeusedasanalternatemethodtotransferinformationfromaCODE
CHIPmoduleintothemeter’smemory.ThemeterwilldirecttheusertoinstallaCODE
CHIPmoduleifitis
attemptingtorunatestanddoesnothavethedatainitsmemory.
TypesofCODECHIPmodules
TherearefivetypesofCODECHIPmodules
ReagentCODECHIPmoduleincludedineachboxoftest
devices.Lotnumberbeginswitha W.
QC SampleCODECHIPmoduleincludedineachboxof
controls.LotnumberbeginswithaC.
QC Device
CODECHIPmoduleincludedintheblackQC
Devicebox(comeswiththeQuidelTriageMeterPro)andis
labeledwiththeserialnumberoftheQCDevice.
SupervisorAccessCODECHIPmoduleincludedwitheach
QuidelTriageMeterPro.LotNumberbeginswithSPR.
SoftwareUpgradeCODE
CHIPmodulesentseparately
whenQuidelTriageMeterProsoftwareupgradesare
available.
CAUTION:RunningtestswiththeSupervisorCODECHIPmoduleinstalleddisablesallQC
Lockouts.RemovetheSupervisorCODECHIPmodulepriortorunningtests.
InstallationofReagent,QCSampleorQCDeviceCODECHIPModules
InstallCodeChip
1. FromtheMainMenuselectINSTALLCODECHIPusingthe
keys.
2. Pressthekey.
MainMenu
Page38of86
3. SlidetheCODECHIPmoduleintotheCODECHIPmodulePortintheQuidelTriageMeterPro,asshownin
thepicture.
4. Pressthekey.
5. ThemeterwilldisplayaconfirmationmessagethattheinformationwasinstalledintotheQuidelTriage
MeterPro’smemory.
6. Presstoreturntoacknowledgethemessage.
7. RemovetheCODECHIPmodulefromtheCODECHIPmodulePort.
8. PlacetheCODECHIPmodulebackinto
itsoriginalcontainerforstorage.
INSTALLATIONOFSUPERVISORCODECHIPMODULE
TheSupervisorCODECHIPmoduleneedsonlytobeinsertedintotheCODECHIPmodulePorttoactivatethe
supervisorfunctions.ThisisanautomaticfunctionandstandardCODECHIPmoduleinstallationisnot
necessary.
Page39of86
Operation
AccesstotheQuidelTriageMeterPro
Function Use Restrictedto
RUNTEST PatientTesting OperatorswithvalidUserID's*
QualityControlTesting OperatorswithvalidUserID's*
Misc.TestTesting OperatorswithvalidUserID's*
RECALLRESULTS LastRecord Iflasttestrunisapatientsample:
OperatorswithvalidUserID's*
Otherwise:Norestriction
PatientTestResults OperatorswithvalidUserID's*
QualityControlResults NoRestriction
Misc.TestResults NoRestriction
INSTALLCODECHIPS TypesofCODECHIP
modules
IfCODECHIPmoduleissoftware
upgrade:
Supervisor
InstallationProcedure Otherwise:Norestriction
SETPARAMETERS AccessControl Supervisor
ProgrammableSettings Supervisor
DELETERESULTS PatientTestResults Supervisor
*UnlessUserIDisbypassedortheSupervisorCODECHIPmoduleisinstalled.
IMPORTANT:Thelasttwofunctions,SETPARAMETERS,andDELETERESULTS,appearonthemain
menuonlywhentheSupervisorCODECHIPmoduleisinsertedintotheCODECHIPmodulePort.The
SupervisorCODECHIPmoduleshouldberemovedfromthemeterduringroutinepatient,quality
control,andMisc.testing.
RunTest
Purpose
RunTestisusedfortheanalysisofatest.
Options
TheRuntestfunctionallowsyoutotest:
QCDevice
NOTE:Accesstovariouspartsofthesoftwareisuser(institution)defined.
Page40of86
QC Sample
PatientSample
Misc.TestSample
Inordertorunapatientsample,themeterrequires:
1. AvalidUserIDhasbeenentered(unless thisfeatureisbypassedoraSupervisorCODECHIPmoduleis
installed).Seepage31.
2. AQCDevicehaspassedwithinthespecifiedtimeperiod.Seepage41.
3. AQCSa mpleforthetestdevicelothaspassedwithinthe
specifiedtimeperiod.Seepage 42.
Ifanyoftheseconditionsdonotexist,themeterwillnotcontinueandwill prompttheusertoperformthe
requiredsteps.
IMPORTANT:
TheQuidelTriageMeterProwillalerttheuserifQChasnotbeenperformedforthe
devicelotbeingtested.
QCDevice
TheQCDeviceshouldberunoneachdayofpatienttesting.
ToruntheQCDevicetest:
1. PressthekeytopowertheQuidelTriageMeterProon.
2.
SelectRUNTESTusingthe keys.
3. Pressthekey.
4. Typeinyouridentification(UserID).Tocorrect,pressthekeytocleartheentireID,orselectthe
incorrectcharacterbyusingthekeys.ThentypeinthecorrectID.
8:23AM01•20•18
RUNTEST
RECALLRESULTS
INSTALLNEWCODECHIP
SELECTWITHvOR^ANDPRESSENTER
PRESSEXITTOEJECTDEVICE
TheMainMenu
NOTE:
TheSupervisorCODECHIPmoduleprovidesaccesstofunctionsthatmayoverridetheseLockouts.
NOTE:
IftheQuidelTriageMeterPro’sinternalqualitycontrol(QC)resultsareunacceptable,exclamation
points(!)willbedisplayedonthescreenintheplaceofthepatientresults.Toobtainresultsrepeatthe
testusinganewdevice.
NOTE:
IftheUserIDbypa ss ison,themeterwillnotaskyoutoenteraUserID.
MainMenu
Run Test
Page41of86
5. Pressthekey.
6. SelectQCDeviceusingthekeys.
“EXP”indicatesthatthe
QCDevicereͲtest
frequencyhasexpired.
PATIENTSAMPLE
QCDEVICE EXP
QCSAMPLE EXP
MISC.TEST
SELECTWITHvOR^ANDPRESSENTER
PRESSEXITTOEJECTDEVICE
MenuscreenasitlookswhenoperatorselectsQCDevice.
 Pressthekey.
8. GentlyinserttheQCDeviceintotheQuidelTriageMeterPrountilyoufeeltheQCDevicecatchonthe
pinandhearanaudible‘click.’
 Pressthekey.
10.TheQuidelTriageMeterPropullsintheQCDeviceandscansit.Thetest
devicemaypartiallymoveinandoutofthemeterseveraltimes.
11.Whenthetestiscomplete,themeterwillbeep,ejectthedeviceanddisplaytheresultsonthe
meter’sscreen.
12.Press
thekeytomakeaprintedcopyoftheresults.
13.RemovetheQCDevicefromtheMeterandplaceintheQCDeviceBox.DONOTDISCARDTHE
QCDEVICE.
IMPORTANT:Importantinstructionsforalltests:
a. GentlyinsertthetestdeviceintotheQuidelTriageMeterPrountilyoufeelthedevicecatchonthe
pin.Youwillhearanaudible‘click’thattellsyouthetestdevicehasbeeninsertedproperly.
b. Pressthekeytostartthetest.
c.
TheQuidelTriageMeterPropullsinthetestdeviceandscansit.
d. Whenthetestiscomplete,themeterwillbeepanddisplaytheresultsonthemeter’sscreen.
e. Pressthekeytomakeaprintedcopyoftheresults.
CAUTION:Afterthetestdevicehasbeeninserted,donotpushthedeviceinfurtherorattempt
topullitout.ThedevicemaybeejectedbyreturningtotheMainMenuandpressingtheEXIT
key.
NOTE:ThemeterwillprompttheusertoinstalltheQCDeviceCODECHIPmoduleiftheQCDevicehasnot
beenrunbefore.
Page42of86

IMPORTANT:
ImportantInstructionsforQCDevicetests:
Dust,lint,fibersandothersmallparticlesmayinterferewiththeQCDevice.K eeptheQCDevicefreeof
contaminants.
TheQCDeviceislightͲsensitiveandshouldbestoredinitsblackopaquecasewhennotinuse.
Ifthe
QCDevicetestsfail,wipetheQCDevicecleanwithalintfreeclothtoremoveanyoils,dust,
fibers,orfingerprints.DonotapplyanyliquidtotheQCDevice.Aftercleaningthedevice,repeatthe
QCDevicetest.
IftheQCDevicetestfailsafteryouhavecleaned
thedevice,contactQuidel.
QCSample
ToruntheQualityControl sample:
1. PressthekeytopowertheQuidelTriageMeterProon.
2. SelectRUNTESTusingthe
keys.
8:23AM01•20•18
RUNTEST
RECALLRESULTS
INSTALLNEWCODECHIP
SELECTWITHvOR^ANDPRESSENTER
PRESSEXITTOEJECTDEVICE
TheMainMenu
3. Pressthekey.
4. Typeinyouridentification(UserID).Tocorrect,pressthekeytocleartheentireID,orselectthe
incorrectcharacterbyusingthe
keys.ThentypeinthecorrectID.
5. Pressthekey.
6. SelectQCSampleusingthe
keys.
PATIENTSAMPLE
QCDEVICE
QCSAMPLE
MISC.TEST
SELECTWITHvOR^ANDPRESSENTER
ORPRESSEXIT TOEJECTDEVICE
MenuscreenasitlookswhenoperatorselectsQCSample.
NOTE:
IftheUserIDbypa ss isON,themeterwillnotaskyoutoenteraUserID.
MainMenu
Run Test
Page43of86
7. Presstheke y.
8. EntertheQClotnumberfromthelabelonthesideofthebottlevialcontaining theQCSample.
9. Tocorrect,pressthe keytocleartheentirenumber.
Thentypeinthecorrectnumber.
10.Pressthekey.
11.Performthetestaccordingtotheinstructionsprovided
inthepackageoftestdevicesyouareusing.
IMPORTANT:Importantinstructionsforalltests:
a. GentlyinsertthetestdeviceintotheQuidelTriageMeterPrountilyoufeelthedevicecatchonthe
pin.Youwillhearanaudible‘click’thattellsyouthetestdevicehasbeeninsertedproperly.
b. Pressthe keytostartthetest.
c.
TheQuidelTriageMeterPropullsinthetestdeviceandscansit.
d. Whenthetestiscomplete,themeterwillbeepanddisplaytheresultsonthemeter’sscreen.
e. Pressthe keytomakeaprintedcopyoftheresults.
CAUTION:Afterthetestdevicehasbeeninserted,donotpushthedeviceinfurtherorattempt
topullitout.ThedevicemaybeejectedbyreturningtotheMainMenuandpressingEXIT.
12.Repeatstepsforeachqualitycontrolsample.
IMPORTANT:ImportantinstructionsforQCSampleTests
D Ifthetestdevice’sinternalqualitycontrol(QC)resultsareunacceptable,theresultsforthe
affectedanalyte(s)willnotappearonthescreen(insteadwillbeflaggedwitha!symbol).
b. IfanyoftheQCSampleresultsareoutofrange,theresultsforthatparticularanalytewill
bein
reversevideo(lighttextonadarkbackgr ound).
c. Apanelwithmultipleanalyteswillstillbeabletoreportpatientresultsonthoseanalyteswhich
passedQC.AnanalytethatfailedQCwillnotbereportedonpatienttests(insteadwillbeflagged
witha#symbol).
NOTE:OnlyenterthefourͲdigitnumericvalueoftheQClotnumber.Donotentertheprecedingalpha
character.
NOTE:IfyouenteraQClotnumberforwhichthereisnodataintheQuidelTriageMeterPro’smemory,you
willseeamessageonthemeter’sscreen:NOQCSampledatainmemory.
Tocorrectthis:InstalltheQCSampleCODECHIP(seepictureonpage37).
Press
ENTERtocontinue.
Page44of86
d.Ifanindividualanalyteisoutsidethespecifiedrange,theQuidelTriageMeterProallowstheusertorerunonlythefailed
analyteonthenextQCSampleTest.Whenallanalyteshavepassed,theQCSampletimerissettothedateofthefirst
passinganalyte.
PatientSample
1. PressthekeytopowertheQuidelTriageMeterProon.
2. SelectRUNTESTusingthekeys.
8:23AM01•20•18
RUNTEST
RECALLRESULTS
INSTALLNEWCODECHIP
SELECTWITHvOR^ANDPRESSENTER
PRESSEXITTOEJECTDEVICE
3. Pressthekey.
4. Typeorscaninyouridentification(UserID).Tocorrect,presstocleartheentireID,orselectthe
incorrectcharacterbyusingthe
keys.ThentypeinthecorrectID.
5. Pressthekey.
6. SelectPATIENTSAMPLEusingthekeys.
PATIENTSAMPLE
QCDEVICE
QCSAMPLE
MISC.TEST
SELECTWITHvOR^ANDPRESSENTER
ORPRESSEXIT TOEJECTDEVICE
7. Pressthekey.
8. Typeorscanthepatient’sidentification(PatientID).Tocorrect,presstocleartheentireID,or
selecttheincorrectcharacterusingthe
keys.ThentypeinthecorrectID.
9. Pressthekey.
MainMenu
Run Test
NOTE
:IftheUserIDbypassisON,themeterwillnotaskyoutoenteraUserID.
NOTE
:IftheBarCodeScannerisbeingused,simplypointthescanneratthebarcode,press thebuttonon
thescannerhandleandwa itforthebeep.TheIDwillappearonthedisplay
Page45of86
PatientSample
10.Iftheinformationis correct,pressthekeytoconfirmthePatientID.
Iftheinformationisincorrect:
SelectCORRECTPATIENTIDusingthe
keys.
Presstheke y.
PressthekeytocleartheentireID,orselecttheincorrectcharacterusingthe
keys.

ThentypeinthecorrectID.
Presstheke y.
11.IftheAUXILIARYIDfeatureisactivated,typeorscanintheID.Tocorrect,pressthekeytoclearthe
entireID,orselecttheincorrectcharacterusingthe

keys.Thentypein thecorrectID.
12.Press
thekey.
13.Preparethetestsampleandperformthetestinaccordancetotheinstructions providedinthepackage
oftestdevicesyouareusing.
14.Ifthetestingfacilityhasauthorizedananalytetobeselectedordeselectedatthetimeofrunninga
patienttest,theuser
willbepromptedtoselectdesiredanalytesordeselectundesiredanalytes.After
thetestdevicehasbeenpulledintothemeter,themeterwilldisplayamessagesimilartothefollowing:
SELECTTESTS
CARDIAC
CKMB
2
MYO
3
TNI
PRESS^TOPAUSEOR0 Ͳ9FORTEST
PRESSENTERTOCONTINUEOREXI
T
Acheckmarkindicatestestswhichareselected.Example:TNI
Theabsenceofacheckmarkindicateswhichtestshavenotbeenselected.Example:MYO
Anumbertotheleftofthetestindicatestestswhichcanbeselectedordeselected.
Example:2MYOindicatesthat
MYOiscurrentlynots elected,butcanbeselectedbypressing2onthe
numerickeypadtodisplay2MYO.
Example:3TNIindicatesthatTNIiscurrentlyselected,butcanbedeselectedbypressing 3 onthe
numerickeypadtodisplay3TNI.
Pressingthenumber
againtogglesthetestbetweenselectanddeselect.
Nonumbertotheleftofthetestindicatestestswhichcannotbealtered.
ExampleCKMBindicatesthatCKMBisalwaysselected,and cannotbedeselected.
f Pressthekey.
f Iftheinformationiscorrect,pressthekeyto
continue.
MainMenu
Run Test
NOTE
:Atleastoneanalytemustbeselectedinordertocontinuetesting.
Page46of86
f Iftheinformationisincorrect,selectCORRECTTESTusingthe
keysandpressthekey
toreturntothelistsoftests.
IMPORTANT:
Importantinstructionsforalltests:
a. GentlyinsertthetestdeviceintotheQuidelTriageMeteruntilyoufeelthedevicecatchonthe
pin.Youwillhearanaudible‘click’thattellsyouthetestdevi cehasbeeninsertedproperly.
b. Pressthekeytostartthetest.
c. The
QuidelTriageMeterPropullsinthetestdeviceandscansit.
d. Whenthetestiscomplete,themeterwillbeepanddisplaytheresultsonthemeter’sscreen.
e. Pressthe keytomakeaprintedcopyoftheresults.
CAUTION
:
Afterthetestdevicehasbeeninserted,donotpushthedeviceinfurtherorattempt
topullitout.ThedevicemaybeejectedbyreturningtotheMainMenuandpressingEXIT.
IMPORTANT:Importantinstructionsforpatienttests:
a. Ifthetestdevice’sInternalQualityControl(QC)resultsareunacceptable,theresultsforthe
affectedanalyte(s)willnotappearonthescreen(insteadwillbeflaggedwitha!symbol).
b. IfanyoftheQCSampleresultsforthelotbeingtestedwere
outofrange,theresultsforthat
particularanalytewillnotappearonthescreen(insteadwillbeflaggedwitha#symbol).
PAT.ID89547326

PAT.ID89547326
BNP 15 pg/mL BNP 650 pg/mL
PAT.RESULTABNORMAL
PRESS1TOACCEPTOR0TOREJECT PRESS1TOACCEPTOR0TOREJECT
Screendisplayingpatientsresult
belowthecutofflevelandacceptor
rejectinstructions.
Screendisplayingpatientsresultabovethe
cutofflevels.
NOTE:IfnokeyispressedontheSELECTTESTSscreen,themeterwaits30secondsthenproceeds
usingonlytheselecteddefaulttests.Then,ifnokeyispressedontheCONFIRMATIONSCREEN,the
testproceedsafterwaiting30seconds.
Page47of86
Misc.TestSample
1. PressthekeytopowertheQuidelTriageMeterProon.
2. SelectRUNTESTusingthe
keys.
8:23AM01•20•18
RUNTEST
RECALLRESULTS
INSTALLNEWCODE CHIP
SELECTWITHvOR^ANDPRESSENTER
PRESSEXITTOEJECTDEVICE
3. Pressthekey.
4. Typeorscaninyouridentification(UserID).Tocorrect,presstocleartheentireID,orselectthe
incorrectcharacterbyusingthe
keys.ThentypeinthecorrectID.
5. Pressthekey.
6. SelectMISC.TEST
usingthe
keys.
PATIENTSAMPLE
QCDEVICE
QCSAMPLE
MISCTEST
SELECTWITHvOR^ANDPRESSENTER
PRESSEXITTOEJECTDEVICE
7. Pressthekey.
8. TypeorscantheMisc.Testidentification(Misc.TestID).Tocorrect,presstocleartheentireID,or
selecttheincorrectcharacterusingthe
keys.ThentypeinthecorrectID.
9. Pressthekey.
10.Iftheinformationis correct,pressthekey
toconfirmtheMisc.TestID.
Iftheinformationisincorrect:
SelectCORRECTMISC.TESTIDusingthekeys.
Pressthekey.
MainMenu
Run Test
NOTE
:IftheUserIDbypassisON,themeterwillnotaskyoutoenteraUserID.
NOTE
:IftheBarCodeScannerisbeingused,simplypointthescanneratthebarcode,pressthebutton
onthescannerhandleandwaitforthebeep.TheIDwillappearonthedisplay.
Page48of86
PressthekeytocleartheentireID,orselecttheincorrectcharacterusingthekeys.
ThentypeinthecorrectID.
Pressthekey.
11.Preparethetestsampleandperformthetestinaccordancetotheinstructions
providedinthepackageoftestdevicesyouareusing.
IMPORTANT:

Importantinstructionsforalltests:
a. GentlyinsertthetestdeviceintotheQuidelTriageMeteruntilyoufeelthedevicecatchonthepin.
Youwillhearanaudible‘click’thattellsyouthetestdevicehasbeeninserted properly.
b. Pressthekeytostartthetest.
c. The
QuidelTriageMeterPropullsinthetestdeviceandscansit.
d. Whenthetestiscomplete,themeterwillbeepanddisplaytheresultsonthemeter’sscreen.
e. Pressthe keytomakeaprintedcopyoftheresults.
CAUTION
:
Afterthetestdevicehasbeeninserted,donotpushthedeviceinfurtherorattempt
topullitout.ThedevicemaybeejectedbyreturningtotheMainMenuandpressingEXIT.
IMPORTANT:Importantinstructionsformisc.tests:
a. Ifthetestdevice’sInternalQualityControl(QC)resultsareunacceptable,theresultsforthe
affectedanalyte(s)willnotappearonthescreen(insteadwillbeflaggedwitha!symbol).
Page49of86
RecallResults
Purpose
RecallResultsisusedtoretrieveresultsstoredintheQuidelTriageMeterPro’smemory.
Function
TheRecallResultsfunctioncanretrievetheseresults:
LastRecord
PatientResults
f PatientID
f PanelType
f TestDeviceLotNumber(L/N)
f UserID
f FromDate
f TillDate
QCResul ts
f QCSampleResults
f QCDeviceResults
f ReagentLots
QC
f ArchiveLastTest
Misc.TestResults
PrintallResults
f PatientResults
f QCSampleResults
f QCDeviceResults
f Misc.TestResults
UploadResultstoLIS
IMPORTANT:
TosafeguardtheUserIDLockout,displayedandprintedUserID'sarepartially
masked.TheSupervisorCODECHIPmo duleisrequiredtobeinstalledtoviewthefullUser ID.
LastRecord
Purpose:
LastRecordisusedtoretrievethetestresultsforthelasttestrun.
Toretrievethelastrecord:
1. PressthekeytopowertheQuidelTriageMeterProon.
2. FromtheMAINMENUselect RECALLRESULTSusingthe
keys.
3. Pressthekey.
4. SelectLASTRECORD
usingthe
keys.
5. Pressthekey.
6. Ifthelasttestrunisapatientr esult,typeorscaninyouridentification(UserID).Tocorrect,press
tocleartheentireID,orselecttheincor rectcharacterbyusingthe
keys.Thentypeinthe
correctID.Press
thekey.
7. Theresultsfromthelasttestrunappearonthescreen.
8. Usethe
keystoviewtheentirerecord.
9. Pressthekeytomakeaprintedcopy.
MainMenu
RecallResults
NOTE
:IftheUserIDbypassisON, themeterwillnotaskyoutoenteraUserID
NOTE
:Ifaresultwasrejectedbytheuser,an“R”will appearnexttotherecalledresult.
Page50of86
PatientResults
Options:
Patientresultscanberetrievedbasedonavarietyofcriteria:
PatientID
TestDeviceLotNumber(L/N)
UserID
FromDate
TillDate
PatientID
ToretrieveresultsbasedonthepatientID:
1. PressthekeytopowertheQuidelTriageMeterProon.
2. FromtheMAINMENUselect RECALLRESULTSusingthe
keys.
3. Pressthekey.
4. SelectPATIENTRESULTSusingthe
keys.
5. Pressthekey.
6. Typeorscan
inyouridentification(UserID).Tocorrect,presstocleartheentireID,orselectthe
incorrectcharacterbyusingthe
keys.ThentypeinthecorrectID.
7. Pressthekey.
8. SelectPANELTYPEusingthe
keys.
9. Usethe
keystoselectthedesired
panel.
10. Choosetheres u ltsyouwanttosee:
Ifyouwanttoseeallpatientresults,pressthekey.Allthepa tientresultsintheQuidelTriage
MeterPro’smemoryfortheselectedpanelwillbedisplayed.
Forindividual patientresults:
a. SelectPatientIDusingthe
keys.
b. TypeorscaninthepatientID.Tocorrect,pressthekeytocleartheentireID,orselectthe
incorrectcharacterbyusingthe
keys.ThentypeinthecorrectID.
c. ToselectapatientID withalphacharacters,usethekeytoenteralphacharacters.
d. Pressthekey.AllresultsoftherequestedpatientIDfortheselectedpanelwillbe
displayed.
11. Usethe
keystoviewtheentirerecord.
12.Pressthekeytomakeaprintedcopy.
13.PressthekeytouploadtheselectedresultstoLIS.
MainMenu
RecallResults
MainMenu
RecallResults
Patient Results
NOTE
:IftheUserIDbypassisON, themeterwillnotaskyoutoenteraUserID
NOTE
:Youmaychoosemorethanoneoftheoptionsforrecallingresultsinordertodefinemoreprecisely
theinformationyouwanttorecallfromtheQuidelTriageMeterPro’smemory.Whenyouchoosemultiple
options,firstinputvaluesinallthecategoriesyouhavechosen.ThenpresstheENTERkey.
Page51of86
TestDeviceLotNum b er(L/N)
Toretrieveresultsbasedonthelotnumberofthetestdevicesused:
1. PressthekeytopowertheQuidelTriageMeterProon.
2. FromtheMAINMENUselectRECALLRESULTSusingthe
keys.
3. Pressthekey.
4. SelectPATIENTRESULTSusingthe
keys.
5.
Pressthekey.
6. Typeorscaninyouridentification(UserID).Tocorrect,presstocleartheentireID,orselectthe
incorrectcharacterbyusingthe
keys.ThentypeinthecorrectID.
7. Pressthekey.
8. SelectPANELTYPEusingthe
keys.
9. Usethe
keystoselectthedesiredpanel.
10.SelectDEVICEL/Nusingthe
keys.
11.Todisplayallresults fortheselectedpanel,pressthekey.
12.Todisplayresultsspecifictoonelot,typeinthelotnumberofthetestdevice.Tocorrect,pressthe
keyto
cleartheentirenumber.Thentypeinthecorrectnumber.
13.Presstheke y.
14.Allresultsfromtestsrunontherequestedlotnumberfortheselectedpanelaredisplayed.
15.Usethe
keystoviewtheentirerecord.
16.Pressthekeytomakeaprintedcopy.
17.PressthekeytouploadtheselectedresultstoLIS.
MainMenu
RecallResults
Patient Results
NOTE
:IftheUserIDbypassisON, themeterwillnotaskyoutoenteraUserID.
NOTE
:OnlyenterthefiveͲdigitnumericvalueofthelotnumber.
Page52of86
UserID
ToretrieveresultsfromtestsrunontheQuidelTriageMeterProstartingwith
theuseryouselect:
1. PressthekeytopowertheQuidelTriageMeterProon.
2. FromtheMAINMENUselect RECALLRESULTSusingthe
keys.
3. Pressthekey.
4. SelectPATIENTRESULTSusingthe
keys.
5. Pressthekey.
6. Typeorscaninyouridentification(UserID).Tocorrect,presstocleartheentireID,orselectthe
incorrectcharacterbyusingthe
keys.ThentypeinthecorrectID.
7. Pressthekey.
8. Select
3$1(/7<3(usingthe
keys.
9. Usethe
keystoselectthedesiredpanel.
10. Select
86(5,'usingthe
keys.
11. Todisplayallresultsfortheselectedpanel,pressthekey.
12. TodisplayresultsspecifictooneUserID,typeintheUserID.Tocorrect,presstocleartheentireID,
orselecttheincorrectcharacterbyusingthekeys.Thentypein
thecorrectID.
13. Pressthekey.
14. Allresultsfromtestsrunbytherequesteduserfortheselected panelaredisplayed.
15. Usethe
keystoviewtheentirerecord.
16. Pressthekeytomakeaprintedcopy.
17. Pressthekeytouploadtheselected
resultstoLIS.
FromDate
ToretrieveresultsfromtestsrunontheQuidelTriageMeterProstartingwith
adateyouselect:
1. PressthekeytopowertheQuidelTriageMeterProon.
2. FromtheMAINMENUselect RECALLRESULTSusingthe
keys.
3. Pressthekey.
MainMenu
RecallResults
Patient Results
NOTE
:IftheUserIDbypassisON, themeterwillnotaskyoutoenteraUserID
MainMenu
RecallResults
Patient Results
Page53of86
4. SelectPATIENTRESULTSusingthe
keys.
5. Pressthekey.
6. Typeorscaninyouridentification(UserID).Tocorrect,presstocleartheentireID,orselectthe
incorrectcharacterbyusingthe
keys.ThentypeinthecorrectID.
7. Pressthekey.
8.
SelectPANELTYPEusingthe
keys.
9. Usethe
keystoselectthedesiredpanel.
10. SelectFROMDATEusingthe
keys.
11. Todisplayallresultsfortheselectedpanel,pressthekey.
12. Todisplayresultsspecifictoonedate,typeinthedate.Tocorre ct,
pressthekeytoclear theentire
number.Thentypeinthecorrectnumber.
13. Pressthekey.
14. Allresultsfromtestsrunfromthedateselectedfortheselectedpanelaredisplayed.
15. Usethe
keystoviewtheentirerecord.
16. Pressthekeytomake
aprintedcopy.
17. PressthekeytouploadtheselectedresultstoLIS.
TillDate
ToretrieveresultsfromtestsrunontheQuidelTriageMeterProbeforeandincluding
adateyouselect:
1. PressthekeytopowertheQuidelTriageMeterProon.
2. FromtheMAINMENUselect RECALLRESULTSusingthe
keys.
3. Pressthekey.
4. SelectPATIENTRESULTSusing
the
keys.
5. Pressthekey.
6. Typeorscaninyouridentification(UserID).Tocorrect,presstocleartheentireID,orselectthe
incorrectcharacterbyusingthe
keys.ThentypeinthecorrectID.
7. Pressthekey.
NOTE
:IftheUserIDbypassisON, themeterwillnotaskyoutoenteraUserID.
MainMenu
RecallResults
Patient Results
NOTE
:IftheUserIDbypassisON, themeterwillnotaskyoutoenteraUserID
Page54of86
8. Select
3$1(/7<3(usingthe
keys.
9. Usethe
keystoselectthedesiredpanel.
10. Select
7,//'$7(usingthe
keys.
11. Todisplayallresultsfortheselectedpanel,pressthekey.
12. Todisplayresultsspecifictoonedate,typeinthedate.Tocorre ct,pressthekeytocleartheentire
number.Thentypeinthecorrectnumber.
13. Pressthekey.
14. All
resultsfromtestsrunbeforeandincludingthedateyouselectedfortheselectedpanelare
displayed.
15. Usethe
keystoviewtheentirerecord.
16. Pressthekeytomakeaprintedcopy.
17. PressthekeytouploadtheselectedresultstoLIS.
QCResults
Options:
Theresultsofthesequalitycontrol(QC)testscan beretrievedfromthe
QuidelTriageMeterPro’smemory:
QualityControl(QCtest) Resultscanberetrievedby:
QCSampleResults PanelType
QCSampleLotNumber(L/N)
DeviceLotNumber(L/N)
UserID
FromDate
TillDate
ViewAll(Default)
QCDeviceResults UserID
FromDate
TillDate
ViewAll(Default)
ReagentLotsQC (DefaultstoViewAll)
ToretrieveQCResultsfortheQCSampleandtheQCDevice,selecttheoptiondesiredandfollowthe
directions.
MainMenu
RecallResults
NOTE
:Youmaychoosemorethanoneoftheoptionsforrecallingresultsinordertomoreprecisely
definetheinformation you wanttorecallfromtheQuidelTriageMeterPro’smemory.Whenyouchoose
multipleoptions,firstmakeselectionsandinputvaluesinallthecategoriesyouhavechosen.The npress
theENTERkey.
Page55of86
QCSampleLotNumber(L/N)
ToretrieveresultsbasedonthelotnumberoftheQCSampleused:
1. PressthekeytopowertheQuidelTriageMeterProon.
2. FromtheMAINMENUselect RECALLRESULTSusingthekeys.
3. Pressthekey.
4. SelectQCRESULTSusingthe
keys.
5. Pressthekey.
6. SelectQCSAMPLERESULTSusingthe
keys.
7. Pressthekey.
8. SelectPANELTYPEusingthe
keys.
9. Usethe
keystoselectthedesiredpanel.
10. SelectQCSAMPLE
L/Nusingthe
keys.
11. TypeintheQCSamplelotnumberofthetestdevice.Tocorrect,presstocleartheentirenumber.
Thentypeinthecorrectnumber.
12. Pressthekey.
13. AllresultsfromtestrunontherequestedQCSamplelotnumberforthe
selectedpanelaredisplayed.
14. Usethe
keystoviewtheentirerecord.
15. Pressthekeytomakeaprinted copy.
16. PressthekeytouploadtheselectedresultstoLIS.
DeviceLotNumber(L/N)
ToretrieveQCSampleresultsbasedonthelotnumberofthetestdevicesused:
1. PressthekeytopowertheQuidelTriageMeterProon.
2. FromtheMAINMENUSELECTRECA LLRESULTSusingthe
keys.
MainMenu
RecallResults
QCResults
QCSam
p
leResults
NOTE
:Todisplayallresultsfortheselectedpanel,leavetheQCSampleL/Nblank.
MainMenu
RecallResults
QCResults
QCSampleResults
Page56of86
3. Pressthekey.
4. SelectQCRESULTSusingthe
keys.
5. Pressthekey.
6. SelectQCSAMPLERESULTSusingthe
keys.
7. Pressthekey.
8. SelectPANELTYPEusingthe
keys.
9. Usethe
keystoselectthedesiredpanel.
10. SelectDEVICEL/N
usingthe
keys.
11. Typeinthedevicelotnumberfromthesideoftheboxoftestdevices.Tocorrect,pressthekeyto
cleartheentirenumber.Thentypeinthecorrectnumber.
12. Pressthekey.
13. AllQCresultsthathavebeenrunonthe
requestedlotnumberfortheselectedpanelaredisplayed.
14. Usethe
keystoviewtheentirerecord.
15. Pressthekeytomakeaprintedcopy.
16. PressthekeytouploadtheselectedresultstoLIS.
UserID
ToretrieveQCresultsfromtestsrunontheQuidelTriageMeterProstarting
withtheuseryouselect:
1. PressthekeytopowertheQuidelTriageMeterProon.

FromtheMAINMENUselectRECALLRESULTSusingthe
keys.
3. Pressthekey.
4. SelectQCRESULTSusingthe
keys.
5. Pressthekey.
6. Selecttheappropriatemenu,QCSAMPLERESULTSorQCDEVICERESULTSusingthe
keys.
7. Pressthekey.
8. IfQCSAMPLERESULTS,selectPANELTYPEusingthe
keys.
9. Usethe
keystoselectthedesiredpanel.
10.SelectUSERID
usingthe
keys.
11.Presstheke y.
12.TypeintheuserID.Tocor rect,pressthekeytocleartheentireID,orselecttheincorrectcharacter
byusingthe

keys.ThentypeinthecorrectID.
13.Presstheke y.
14.AllQCSampleorQCDevice
resultsfromtestsrunbytherequesteduser,fortheselectedpanelifQC
Sample,aredisplayed.
MainMenu
RecallResults
QCResults
QCSampleResults
QCDeviceResults
Page57of86
15.Usethe
keystoviewtheentirerecord.
16.Pressthekeytomakeaprintedcopy.
17.PressthekeytouploadtheselectedresultstoLIS.
FromDate
ToretrieveQCresultsfromtestsrunontheQuidelTriageMeterProstarting
withadateyouselect:
1. PressthekeytopowertheQuidelTriageMeterProon.
2. FromtheMAINMENUselect RECALLRESULTSusingthe
keys.
3. Pressthekey.
4. SelectQCRESULTSusingthe
keys.
5. Pressthekey.
6. Selecttheappropriatemenu,QCSAMPLERESULTS
orQCDEVICERESULTSusingthe
keys.
7. Pressthekey.
8. IfQCSAMPLERESULTS,selectPANELTYPEusingthe
keys.Usethe
keystoselectthe
desiredpanel.
9. SelectFROMDATEusingthe
keys.
10. Pressthekey.
11. Typeinthedate.
Tocorrect,pressthekeytocleartheentirenumber.Thentypeinthecorrect
number.
12. Pressthekey.
13. Allresultsfromtestsrunfromthedateselected,fortheselectedpanelifQCSample,aredisplayed.
14. Usethe
keystoviewtheentirerecord.
15. Press
thekeytomakeaprintedcopy.
16. PressthekeytouploadtheselectedresultstoLIS.
TillDate
ToretrieveQCresultsfromtestsrunontheQuidelTriageMeterProbeforeand
includingadateyouselect:
1. PressthekeytopowertheQuidelTriageMeterProon.
2. FromtheMAINMENUselect RECALLRESULTSusingthe
keys.
3. Pressthekey.
4. SelectQCRESULTSusing
the
keys.
5. Pressthekey.
6. Selecttheappropriatemenu,QCSAMPLERESULTSorQCDEVICERESULTSusingthe
keys.
7. Pressthekey.
8. IfQCSAMPLERESULTS,selectPANELTYPEusingthe
keys.Usethe
keystoselectthe
desiredpanel.
MainMenu
RecallResults
QCResults
QCSampleResults
QCDeviceResults
MainMenu
RecallResults
QCResults
QCSampleResults
QCDeviceResults
Page58of86
9. SelectTILLDATEusingthe
keys.
10. Pressthekey.
11. Typeinthedate.Tocorrect,pressthekeytocleartheentirenumber.Thentypeinthecorrect
number.
12. Pressthekey.
13. Allresultsfromtestsrunbeforeandincludingthedateselected,fortheselectedpanelifQCSample,are
displayed.
14. Usethe
keystoviewtheentirerecord.
15. Pressthekeytomakeaprintedcopy.
16. PressthekeytouploadtheselectedresultstoLIS.
ReagentLotsQC
AlistofReagentCODECHIPmodulesinmemory,alongwithcorresponding
expirationdatesforaQCSamplecanberecalledfrommemory.
Thelistwillcontain:LotNumber,PanelType,ExpirationDate
IftheQCexpirationdatehaspassed(expired),thedatewillbeinreversevideo(lighttextonadark
background).
IftheQCforthelothasfailed,thewordFAILEDwilla ppearinplaceofanexpirationdate.
If
aCODECHIPhasbeeninstalledforanewlot,butQChasyettobeestablished,thewordsNOTRUN
willappearinplaceoftheexpirationdate.
IfQC frequencyissettonone,thereagentlotexpirationdatewillbedisplayed.
REAGENTLOTSͲQC
DLN PANEL EXP
47067 CARDIAC 12Ͳ16Ͳ17
48741 CARDIAC 09Ͳ04Ͳ18
48197 BNP NOTRUN
47827 DRUGSCREEN FAILED
PRESSPRINTORPRESSENTER
TheReagentLotsQCScreen
ArchiveLastTest
ThisfunctionwillsavedetailedinformationregardingthelastperformedtestforanalysisbyQuidel.Usethis
functionwhenrequestedbyQuidel.
Thisfunctionshouldbe usedwhenaveryunusualtestresultisreceivedandwhenthelaboratorywishesto
havethemeterevaluatedfurther.ARCHIVELASTTESTwillonly
beabletosavedatafromthemostrecent
testrun.
MainMenu
RecallResults
QCResults
ReagentLotsQC
ArchiveLastTest
Page59of86
Misc.TestResults
Options:
Misc.Testresultscanberetrievedbasedonavarietyofcriteria:
Misc.TestID
TestDeviceLotNumber(L/N)
UserID
FromDate
TillDate
Misc.TestID
ToretrieveresultsbasedontheMisc.TestID:
1. PressthekeytopowertheQuidelTriageMeterProon.
2. FromtheMAINMENUselect RECALLRESULTSusingthe
keys.
3. Pressthekey.
4. SelectMISC.TESTRESULTSusingthe
keys.
5. Pressthekey.
6. SelectPANELTYPEusingthe
keys.
7. Usethe
keystoselectthedesiredpanel.
8. Choosetheresultsyouwanttosee:
IfyouwanttoseeallMisc.Testresults,pressthekey.AlltheMisc.TestresultsintheQuidelTriage
MeterPro’smemoryfortheselectedpanelwillbedisplayed.
Forindividual Misc.Testresults:
a. SelectMISC.TESTIDusingthe
keys.
b. Typeorscanin
theMisc.TestID.Tocorrect,pressthekeytocleartheentireID,orselectthe
incorrectcharacterbyusingthe
keys.ThentypeinthecorrectID.
c. ToselectaMisc.TestIDwithalphacharacters,usethekeytoenteralphacharactersorusethe
keyinplaceofthealphacharacter.
d. Pressthekey.Allresults oftherequestedMisc.TestIDfortheselectedpanelwillbedisplayed.
9. Usethe
keystoviewtheentirerecord.
10. Pressthekeytomakeaprinted copy.
11. Pressthekeyto
uploadtheselectedresultstoLIS.
MainMenu
RecallResults
MainMenu
RecallResults
MiscTestResults
NOTE
:Youmaychoosemorethanoneoftheoptionsforrecallingresultsinordertodefinemoreprecisely
theinformationyouwanttorecallfromtheQuidelTriageMeterPro’smemory.Whenyouchoosemultiple
options,firstinputvaluesinallthecategoriesyouhavechosen.ThenpresstheENTERkey.
Page60of86
TestDeviceLotNum b er(L/N)
Toretrieveresultsbasedonthelotnumberofthetestdevicesused:
 Pressthekeytopower theQuidelTriageMeterProon.
2. FromtheMAINMENUselect RECALLRESULTSusingthe
keys.
3. Pressthekey.
4. SelectMISC.TESTRESULTSusingthe
keys.
5. Pressthekey.
6. SelectPANELTYPEusingthe
keys.
7. Usethe
keystoselectthedesiredpanel.
8. SelectDEVICEL/N
usingthe
keys.
9. Todisplayallresultsfortheselectedpanel,pressthekey.
10. Todisplayresultsspecifictoonelot,typeinthelotnumber ofthetestdevice.Tocorrect,pressthe
keytocleartheentirenumber.Thentypeinthecorrectnumber.
11.
Pressthekey.
12. Allresultsfromtestsrunontherequestedlotnumberfortheselectedpanelaredisplayed.
13. Usethe
keystoviewtheentirerecord.
14. Pressthekeytomakeaprinted copy.
15. PressthekeytouploadtheselectedresultstoLIS.
MainMenu
RecallResults
MiscTestResults
NOTE
:OnlyenterthefiveͲdigitnumericvalueoftheQClotnumber.
Page61of86
UserID
ToretrieveresultsfromtestrunontheQuidelTriageMeterProstartingwith
theuseryouselect:
1. PressthekeytopowertheQuidelTriageMeterProon.
2. FromtheMAINMENUselect RECALLRESULTSusingthe
keys.
3. Pressthekey.
4. SelectMISC.TESTRESULTSusingthe
keys.
5. Pressthekey.
6. SelectPANELTYPEusingthe
keys.
7. Usethe
keystoselectthedesiredpanel.
8. SelectUSERIDusingthe
keys.
9. Todisplayallresultsfortheselectedpanel,pressthekey .
10. Todisplayresultsspecificto
oneUserID,typeintheUserID.Tocorrect,presstocleartheentireID,
orselecttheincorrectcharacterbyusingthe
keys.ThentypeinthecorrectID.
11. Pressthekey.
12. Allresultsfromtestsrunbytherequesteduserfortheselectedpanelare
displayed.
13. Usethe
keystoviewtheentirerecord.
14. Pressthekeytomakeaprinted copy.
15. PressthekeytouploadtheselectedresultstoLIS.
FromDate
ToretrieveresultsfromtestrunontheQuidelTriageMeterProstartingwithadate
youselect:
 Pressthekeytopower theQuidelTriageMeterProon.
2. FromtheMAINMENUselect RECALLRESULTSusingthe
keys.
3. Pressthekey.
4. SelectMISC.TESTRESULTSusingthe
keys.
5. Pressthekey.
6. SelectPANELTYPEusingthe
keys.
7. Usethe
keystoselectthedesiredpanel.
8. SelectFROMDATE
usingthe
keys.
MainMenu
RecallResults
MiscTestResults
MainMenu
RecallResults
MiscTestResults
Page62of86
9. Todisplayallresultsfortheselectedpanel,pressthekey.
10. Todisplayresultsspecifictoonedate,typeinthedate.Tocorre ct,pressthekeytocleartheentire
number.Thentypeinthecorrectnumber.
11. Pressthekey.
12. Allresultsfromtestsrunfrom
thedateselectedfortheselectedpanelaredisplayed.
13. Usethe
keystoviewtheentirerecord.
14. Pressthekeytomakeaprintedcopy.
15. PressthekeytouploadtheselectedresultstoLIS.
TillDate
ToretrieveresultsfromtestrunontheQuidelTriageMeterProbefoeandincluding
adateyouselect:
1. PressthekeytopowertheQuidelTriageMeterProon.
2. FromtheMAINMENUselectRECALLRESULTSusingthe
keys.
3. Pressthekey
4. SelectMISC.TESTRESULTSusingthe
keys.
5. Pressthekey.
6. SelectPANELTYPEusingthe
keys.
7. Usethe
keystoselectthedesiredpanel.
8. SelectTILLDATEusingthe
keys.
9. Todisplayallresultsfortheselected
panel,pressthekey.
10. Todisplayresultsspecifictoonedate,typeinthedate.Tocorre ct,pressthekeytocleartheentire
number.Thentypeinthecorrectnumber.
11. Pressthekey.
12. Allresultsfromtestsrunbeforeandincludingthedateyouselectedfor
theselectedpanelare
displayed.
13. Usethe
keystoviewtheentirerecord.
14. Pressthekeytomakeaprintedcopy.
15. PressthekeytouploadtheselectedresultstoLIS.
MainMenu
RecallResults
MiscTestResults
Page63of86
PrintAllResults
Options:
AllstoredresultscanbeprintedfromtheQuidelTriageMeterPro’s
memory.
PatientResults
QC SampleResults
QC DeviceResults
Misc.TestResults
PatientResults
ToprintallPatientResults:
1. PressthekeytopowertheQuidelTriageMeterProon.
2. FromtheMAINMENUselect RECALLRESULTSusingthe
keys.
3. Pressthekey.
4. SelectPRINTALLRESULTSusingthe
keys.
5. Pressthekey.
6. SelectPATIENTRESULTSusingthe
keys.
7. Pressthekey.
8. Typeorscaninyouridentification(UserID).Tocorrect,presstocleartheentireID,orselectthe
incorrectcharacterbyusingthe
keys.ThentypeinthecorrectID.
9. Pressthekeytomakeaprintedcopy.
QCSampleResults
ToprintallQCSampleResults:
1. PressthekeytopowertheQuidelTriageMeterProon.
2. FromtheMAINMENUselect RECALLRESULTSusingthe
keys.
3. Pressthekey.
4. SelectPRINTALLRESULTSusingthe
keys.
5. Pressthekey.
6. SelectQCSAMPLERESULTSusingthe
keys.
7. Pressthekeytomakeaprintedcopy.
MainMenu
RecallResults
MiscTestResults
MainMenu
RecallResults
PrintAllResults
NOTE
:Whenprintinglargequantities ofpatientresults,itisrecommendedtofirstinsertafullrollof
paper.
NOTE
:IftheUserIDbypassisON, themeterwillnotaskyoutoenteraUserID.
Page64of86
QCDeviceResults
ToprintallQCDeviceResults:
1. PressthekeytopowertheQuidelTriageMeterProon.
2. FromtheMAINMENUselect RECALLRESULTS
usingthe
keys.
3. Pressthekey.
4. SelectPRINTALLRESULTSusingthekeys.
5. Pressthekey.
6. SelectQCDEVICERESULTSusingthe
keys.
7. Pressthekeytomakeaprintedcopy.
Misc.TestResults
ToprintallMisc.TestResults:
1. PressthekeytopowertheQuidelTriageMeterProon.
2. FromtheMAINMENUselect RECALLRESULTS
usingthe
keys.
3. Pressthekey.
4. SelectPRINTALLRESULTSusingthekeys.
5. Pressthekey.
6. SelectMISC.TESTRESULTSusingthe
keys.
7. Pressthekeytomakeaprintedcopy.
UploadResultstoLIS
WheninterfaceddirectlytoanLIS,selectingthisfeaturewillsendalltestresults
totheLIS.Priortothefirstupload,theLISfeaturemustbeactivatedfromthe
SetParametermenuitemusingtheSupervisorCODECHIP
module.
DeleteResults
Purpose
TheDeletefunctionisusedtoremoveallpatienttestinformationfromtheQuidelTriageMeterPromemory.
MainMenu
RecallResults
UploadResultstoLIS
MainMenu
RecallResults
PrintAllResults
NOTE
:Thedeletefunction isavailableonlywhentheSupervisorCODECHIPmoduleisinsertedinthe
QuidelTriageMeterProCODECHIPmodulePort.
Page65of86
T
HE
QuidelTriageMeterProMemory
WithintheQuidelTriageMeterPro,PatientResultmemorycanhold750datasets,QCSamplecanhold200
datasets,QCDevicecanhold70datasets,andMisc.Testcanhold250datasets.Whenthememoryisfull,
theoldestresultisautomatically overwrittenwhenanewresultis
added.Themeterdisplayalertstheuser
asthepatientmemorybecomesfull.
WARNING
PATIENTMEMORYHAS
SPACEFORXXXMORE
RECORDS
PRESSENTER
TheMemoryAlertScreen
IMPORTANT:
TheDeletefunctionpermanentlyremovesallpatientdatafromthemetermemory.
Beforecontinuing,youmaywishtofirstprintresults(oruploadtoyourdatamanagement
system/LIS).QCorMisc.TestResultsca nnotbemanuallydeleted.
DeletePatientResult
Thisfunctionwilldeleteallpatientresultsinmetermemory.Sincethisactionisnot
reversible,themeterwillrequiretwoconfirmations ofintenttodeleteresults.
1. InserttheSupervisorCODECHIPmoduleintheCODECHIPmodulePort
2. FromtheMAINMENU,selectDELETERESULTSusing
thekeys.

MainMenu
DeleteResults
Page66of86
8:23AM01•20•18
RUNTEST
RECALLRESULTS
INSTALLNEWCODECHIP
SETPARAMETERS
DELETERESULTS
SELECTWITHvOR^ANDPRESSENTER
PRESSEXITTOEJECTDEVICE
TheSupervisor’sMainMenu
3. Pressthekey.
4. Awarningmessageonthescreenstates:DELETEPATIENTRESULT.
5. Ifthisisacceptable,pressthekey.(Ifthisisnotacceptable,pressthekey).
6. Amessageonthescreenstates:ALLPATIENTRESULTSWILLBEDELETED.
7.Pressthekeytodeleteallpatientresults.(Ifyoudonotwanttodeleteallpatientresults,
pressthekey).
DisplayBacklight
TheQuidelTriageMeterProhasthecapabilityfortheoperatortobacklightthedisplayforproviding
improvedclaritydependingonambientlightingconditions.Thebacklightfeaturecanbecontrolledbythe
backlightbuttononthekeypad.Whenthebacklightbuttonispressed,thebacklightinitially
comesonatmaximumbrightness.Subsequent
pressescyclethebacklightfrombrighttodimandthento
off.Tosavepower,thebacklightturnsoffifnokeyshavebeenpressedfor15seconds.Pressinganykeywill
restorethebacklighttothepreviouslevel.

NOTE:PresstheEXITkeytocancelatanytimebeforestep7.
Page67of86
AlphaNumericMode
TheQuidelTriageMeterProhasthecapabilityfortheoperatortoenterbothnumericandalphacharacters
fromthekeypad.Thekeypadoperatesinnumericmodeuntilalphamodeisactivatedbypressingthe
key.Themeterstaysinalphamode,indicatedbytheletters"ABC"shownatthetopof
thedisplay,untilthe
keyispressedagain.Wheninalphamode,pressinganykeycausesthealphacharactertobe
displayed.
1
Ͳ/#()*.\
7
PQRS
2
ABC
8
TUV
3
DEF
9
WXYZ
4
GHI
0 noalphacharacterisdispla yed
5
JKL
. noalphacharacterisdispla yed
6
MNO
Thefirstpressofabuttoncausesthefirstlettertobedisplayed.Thesecondpressofthesamebuttonwithin
threesecondscausesthesecondlettertobedis played.Thethirdpressofthesamebuttonwithinthree
secondscausesthethirdlettertobedisplayed.
Afterthe
lastcharacterinasequenceisdisplayed,thenextbuttonpressofthesamebuttonwithin3
secondsdisplaysthefirstcharacterandthesequenceisrepeated.
Ifmorethanthreesecondspasswithouttheoperatorpressingthebutton,thenextpressofthatbutton
causesthefirstcharacterto
bedisplayed.Ifanewbuttonispressed,thefirstcharacterofthenewbuttonis
displayedwithsubsequentpressesrepeatingthesequence.


Page68of86
ServiceandMaintenanceProcedures
Ifanyserviceormaintenanceisrequired,theQuidelTriageMeterProshouldbesenttothemanufactur er.
Nomaintenanceotherthanpaper/batteryreplacementandperiodicexternalcleaningisrequiredofthe
operator.
PaperReplacement
ChangingPaperBetweenFunctions
1. TearoffanyexcesspaperstickingoutoftheQuidelTriageMeterPro.
2. Removethepapercompartmentcoverbypullinguponthecoverasindicatebythearrowontheback
ofthecover.
3. Removeunusedpaperortheemptypaperspindlefrompapercompartment.
4. Tearorcutaclean,straightedgetofeedintotheprinter.Donotcutpaperatanangle,astheprinter
mustsensetheedgeofthepaperalongthefeedpath.
5. Insertthenewrollofpaperintothepapercompartment.
6. Positionthepapersuchthatthepaperwillfeedfromundertheroll(asopposedtooverthetopofthe
roll,seepicture,above).
7. Insertthepaperedgeunderthepaperroller(platen)untilitfirmlyseatsorresistanceisfelt.
8. Pressthekey.
9.
Replacethecoveroftheprinterandcontinueoperation.
NOTE:Theprintercontainsapapersensorandwillfeedthenewpaperrollonlywhenpaperwithaclean
straightlineispressedintothepaperroller.
NOTE:IfPAPERFEEDbuttonispressedbeforetheprintersensesthepaper,theprinterwillnotrespond
tothecommand.
NOTE:Theprinterutilizesthermalpaper,iftheprintedpaperisblankafterreplacingaroll,verifythatthe
paperrollisnotupsidedown.
Page69of86
ChangingPaperWhilePrinting
IfthemeterrunsoutofpaperwhileinthemiddleofaprintingtaskandisdisplayingthemessagePRINTER
FAILURE,CHECKPRINTERORREPLACEPAPER,thepapermaybechangedandtheprintjobcontinued
withoutlossofdataorreprintingfromthebeginning.Performthefollowingstepspriortopressing anynew
keys:
1. Removethepapercompartmentcoverbypullinguponthecoverasindicatebythearrowontheback
ofthecover.
2. Gentlypullremainingpaperslowlyupthrough theprinter.
3. Removepaperspindlefromthepapercompartment.
4. Tearorcutaclean,straightedgetofeedintotheprinter.Donotcutpaperatanangle,astheprinter
mustsensetheedgeof
thepaperalongthefeedpath.
5. Insertthenewrollofpaperintothepapercompartment.
6. Positionthepapersuchthatthepaperwillfeedfromundertheroll(asopposedtooverthetopofthe
roll,seepicture,above).
7. Insertthepaperedgeunderthe
paperroller(platen)untilitfirmlyseatsoraresistanceisfelt.
8. Presstheke y.
9. Whentheprinterbeginstopullthepaperin,releasethepaper.TheprinterwillfeedafewlinesandreͲ
commencetheprintjob.Toensurenodataislost,the
printerwillreprintuptotheprevious5lines.
10.Afterthepaperhasappearedabovetheprinter,replacetheprintercover,takingcaretoensurethe
paperdoesnotgetcaughtinsidethemeter.
NOTE:Theprintercontainsapapersensorandwillfeedthenewpaperrollonlywhenpaperwithaclean
straightlineispressedintothepaperroller.
NOTE:IfPAPERFEEDbuttonispressedbeforetheprintersensesthepaper,theprinterwillnotrespond
tothecommand.
NOTE:Theprinterutilizesthermalpaper,iftheprintedpaperisblankafterreplacingaroll,verifythatthe
paperrollisnotupsidedown.
Page70of86
Cleaning
TheQuidelTriageMeterProrequiresminimalmaintenance.Occasionalcleaningoftheexteriorwithmild
soapandwatersolutionissufficient.Afterusingadamp,notwet,spongeorclothtoapplyamildsoapand
watersolutionontheoutsideofthemeter,wipethemeterdryusingasoftcloth
orabsorbenttissue.Donot
allowwatertoseepintotheprinter.Donotimmersethemeterinwaterorotherliquids.
Ifbloodorotherfluidsarenotallowedenoughtimetofullyabsorbintotestdevices,thedevicetrackdoor
mayoccasionallyrequirecleaning.Usingacottonswab
dampenedwithisopropylalcoholandapairof
tweezers,carefullyliftthedoorandcleanbothfrontandbackofthedoor.
CheckBatteryLevel
TheQuidelTriageMeterProcheckstheactivepowersourceafterejectingadeviceorprinting.Tocheckthe
batterypower,navigatetotheMAINMENU,disconnecttheACpowersupply,pressthekey(themeter
attemptstoejectthedevice),andlookfortheLowPowerIndicatorontheMain
Menuofthemeterdisplay.
Iftheflashingindicatorappears,replacethebatteries.
BatteryReplacement
Turnoffthemeter.
Removethebatterycoverfromthebottomofthemeterbyslidingthecoverinthedirectionofthe
arrow.
NOTE:Neverinsertanytools,swabs,orcleaningmaterialsintothedevicetrack.
NOTE:Themeterisdesi gnedtoperformandprintatleast100testsbeforenewbatteriesarerequiredtobe
replaced.Whenthemeterisnotinuse,itshouldbeturnedoff.Topreservebatterylife,ensuretheAUTO
POWERͲOFFfunctionissetto1/2HOUR.
Page71of86
Removethebatteries.Somecountriesrequirethatbatteriesbedisposedofinaccordancewithspecific
governmentalregulations(Directive2006/66/EC).Disposeofthebatteriesincompliancewithlocal
regulations.
InsertfoursizeAA1.5Ͳvoltbatteriesintothebatterycompartment.Besurethattheplus(+)andminus
(Ͳ)
signsofthebatteriesmatchthesamesignsembossedinthebatterycompartment
Replacethecovertothebatterycompartment.
TotalQualityAssurance
Introduction
Qualityassuranceprogramscontrolandmonitoranalyticalmethods.Thequalitycontrolofanalytical
methodsachievestwobasicgoals:
1. Ensuresthatthereagentsandinstrumentareworkingproperly
2. Ensuresthattheprocedurewasperformedcorrectly
Themechanismbywhichanalyticalmethodsaretraditionallycontrolledisthroughtheuseof
liquidcontrol
specimenscontainingknownamountsofanalytes.
Overthelast15yearsadvancesintechnologyhavechangedthemannerinwhichthediagnosticreagents
areconfigured.AssaysystemsforthePointͲofͲCarearegenerallyunitizedtestingdevicesthataredistinct
fromtraditionallaboratoryͲbasedanalyzers.TheQuidelTriagetests
areunitizedteststhatdonotrequire
NOTE:Ifthemeterdoesnotpoweronafterreplacingbatteries,verifythebatteriesarelinedupaccording
tothesymbolsinthebatterycompartment.
NOTE:Rechargeablebatteriesareauthorizedforuse,howeverduetothenatureofthesetypesof
batteries,itisexpectedtheywillrequirerechargingmorefrequently.TheQuidelTriageMeterProdoesnot
containabuiltͲinbatterycharger,rechargeablebatteriesmustberechargedoutsideofthemeter.
NOTE:Formoredetailedinformation,contactQuidel.
Page72of86
additionalreagents.Eachdevicecontainsafullcompleme ntof reagentsthatarerequiredtoperformthe
immunoassay.
TheQuidelTriagetestshavebeendesignedtoincludeassaycontrolsineachtestdevicethatmonitorthe
assayprocedureandthereagentintegrity.Theassaycontrolspresentauniqueandvaluableextensionto
thetraditionalapproachtoqualitycontrolusingliquidreagents.Iftheproductisstoredunderspecifie d
conditions,
properfunctioningoftheassaycontrolsindicatesthatthetestdeviceisvalidandproper
adherencetoassayprotocolhasbeenfollowed.TheassaycontrolsintheQuidelTriagetestsrequirethe
needtoreͲthinkthefrequencywithwhichexternalqualitycontrolspecimensshouldbetested.
TheQuidelTriage
testshavebeendesignedtoyieldlong Ͳtermreagentstability.Inaddition, thesystem
providesqualitycontrolparametersinthreeseparateanddistinctcategories:
Internaldevicecontrols
Meterelectroniccontrols
Softwarecontrols.
Thecombinationofthesequalitycontrolfeaturesprovides moreinformationaboutthevalidityofeach
assaythancanbeprovidedbythemoretraditionalmethodsofreagentandassayprocedurevalidation.
QualityAssuranceFunctions
1. InternalDeviceQualityAssurance
InternalPositiveControlZone
BaselineControlZone
TimingControlZone
ZoneQualityRequirements(TraceQC)
2. MeterQualityAssurance
SelfͲTestMode
InternalCalibrationChip
ElectronicQCDevice
BarCode
3. SoftwareQualityAssurance
SupervisorCODECHIP
module
UserID#
PatientID#
Resultmemorystorage
QCFrequency
QCLockouts
4. QualityControlSamples
InternalDeviceQualityAssurance
TheInternalControlZonesarebuiltͲin functionalimmunoassays.Thiscontrolverifiesthatthedevice
functionedproperlyandthattheassaywascorrectlyperformed.TheInternalPositiveControlZoneensures
thattheantibodies,thereagentreconstitution,thetimegateandthedeviceflowarefunctioning properly.
Themanufacturersetsacceptablerangesfor
thecontrolzonesforeachlotofreagents.
TheBaselineControlZonemonitorsthepresenceofinterferingsubstancesinthepatientsamplethat may
altertheimmunoassayprocess.Ifaspecimencontainsasubstancethathasamajoreffectonthehighand
Page73of86
lowcontrolsandthebaselinecontrolandtherebyontheimmunoassay,theresultsarenotdisplayedand
erroneousresultsarenotreported.
TheTimingControlZonemonitorsassaycompletion.Ifaninadequateamountofspecimenisaddedtothe
testdeviceorifaspecimenclotsinthedevice,this
controlwillpreventthedisplayoftheresults.
ATraceQCalgorithmhasbeenprogrammedwhichevaluates the size,locationandsignaltonoiseratioof
eachzoneonthedevice.Aberrationsinzonequalityexceedingpresetlimitswillcauserejectionofoneor
moreoftheassayresults.
MeterQualityAssurance
TheSelfͲTestModeisinitiatedeachtimethemeteristurnedON.IntheselfͲtestmodethemeterscansan
InternalCalibrationChip.Eachcalibration chipscanisusedtovalidateandadjust,ifnecessary,themeter
calibration.IntheselfͲtestmode,theavailablememoryforpatientresults
ischecked.Whenthecapacityis
belowapreͲsetlimit,theremainingcapacityisrelatedviaascreenmessage.Thesoftwareprogramis
evaluatedandisterminatedifcorrupted.SoftwaretestresultsandCODECHIPmoduledataareverifiedfor
integritybeforeeachuse.
AQCDeviceisprovided
witheachmetertoallowtheusertofurtherchecktheintegrityofthemeter.The
simulatorcontainssixfluorescentzonesofvaryingintensitythataremeasuredbythemeter.TherearepreͲ
programmedacceptablerangesforthesixzones.Inaddition,thezonesallowthemetertocheckfor
horizontalandverticalalignmentofthezonemeasurementtoverifylaseralignmentanddevicetransport.
TheQCDevicealsoconfirmslaserstability.
ABarCodeisprintedonthebottomofeverytestdevicetoidentifythereagentlotnumberandtoprevent
theuseofexpiredreagents.
SoftwareQualityAssurance
ThesoftwareQAiscontrolledthroughavarietyofsoftwarefeatures,someofwhichcanbeoptional
dependingontheenvironmentinwhichtheassayisperformed.InaPointͲofͲCarelocation,adesignated
supervisorcanselectprogramparameterstoensureadherencetoqualitystandards.Inacentrallaboratory
location,
selectedparameterscanbebypassedifdesired.
ASupervisorCODECHIPmoduleisprovidedwitheachmeter.ThisCODECHIPmoduleallowsaccessto
meterfunctionsnotavailabletotheroutineuser.Theadditionalfunctionsareselectionofparameter
settingsanddownloadingandde letionofresultsfromthemetermemory.
AUserIDcanbeassignedinthesoftwaretolimitaccesstothepatienttestingmodetoonlythose
individualswhohavebeenproperlytrainedandhavedemonstratedtestingproficiency.AUserIDand
expirationdateareprogrammedintothemetermemoryandaresubsequentlyassociatedwithallpatient
resultsgeneratedbytheuser.Thisallowsthesupervisortoreviewusers’testingfrequency,andintheevent
offrequenttestfailures,toidentifypotentialproblemsintraining.
AnadditionalsecurityfeatureoftheUserIDisthepartialmaskingoftheIDondisplaysandprintouts.This
preventsunauthorized
individualsfromdetectingandusinganexistingUserID.Thisfeatureisoverridden
whentheSupervisorCODECHIPmoduleisinstalled.
Page74of86
APatientIDisrequiredforallpatienttestsperformed.All QC sampleandpatientresultsarestoredinthe
meter’smemoryandcanbedownloadedtoalaboratoryinformationsystem(LIS)oranexternaldata
managementsoftwareprogramforfurtherviewingandprintingofqualitycontrolreports.
TheSupervisor
canselecttheQCFreque ncyforperformingexternalliquidcontrolsamples.Whenitistime
toperformexternalQCsamples,theusermusttesttheexternalliquidcontrolsandobtainacceptable
resultsbeforepatientsamplescanbetested.WhenQCsampleshavebeensuccessfullyperformed,theuser
isallowedtoperform
testsonpatientsamples.IfQCsamples arenotwithinanacceptablerange,aQC
Lockoutfunctionpreventstestingofpatientsamples.Ifdesired,thesupervisorinacentrallabsettingcan
bypassthisfunction.Iftestdevicecontrolsareoutofrange,aQCLockoutfunctionisactivatedand
patient
testresultsareblockedfromviewingandprinting.Thispreventsunacceptableresultstobecommunicated
toandusedbythephysician.
Additionally,iftestdevicesareexpired,aQCLockoutfunctionisactivatedandpreventstestingusing
expiredtestdevices.
QualityControlSamples
Itisstillvaluabletoapplythetraditionalapproachtoqualitycontrolbytestingqualitycontrolsamples.
Thesecontrolswillcheckthetotalintegrityofthesystem.Theintervalforanalyzingthesecontrols,
however,canbeextendedduetothemanyother TotalQualityAssurancefeaturesinherentintheQuidel
Triage
tests.
TheQuidelTriagetestshavebeendesignedtomaximizeTotalQualityAssuranceinany testing environment.
ThecombinationoftheQCfeaturesoftheQuidelTriagetestsreducetheimpactofproceduralerrors,
ensurereagentintegrity,andassurethatpatientresultsareaccurateeachandeverytimeatest
is
performed.BasedonthecurrentCLIAguidelinesandotherregulatingbodies,Quidelmakesthefollowing
QCrecommendations:
RuntwolevelsorPOSandNEGexternalliquidcontrolsamplesasa ppropriate witheachnewlotof
reagentsandonceeverythirtydayswithcontinueduseofthesamereagentlot
number
RuntheQCDevicedaily.
NotesRegardingSoftwareControls
1. UserIDAccess
Topreventuntrainedpers ons fromperformingatest,avalidUserID(1Ͳ16charactersinlength)
mustbeenteredbeforeaccessisgiventotheRUN
TESTfunctioniftheUserIDbypassisOFF.Ifa
UserIDBypassisON,thenanyonecananalyzeatest.
Topreventunauthorizedpersonsfromaccessingpatientinformation,avalidUserIDmustbe
enteredbeforeaccessisgiventotheRECALL
PATIENTRESULTSfunctionsiftheUserIDbypassis
OFF.IfaUserIDBypassisON,thenanyonecanrecallpatientdata.RECALL
PATIENTRESULTS
functionsincludere callingpatientresults,printingpatientresults,andrecallingthelasttestrunon
themeterwhenthelasttestisapatientsample,
TheUserIDBypassiscontrolledbytheSET
PARAMETERSfunction,whichrequiresaSupervisor
CODECHIPmoduleforaccess.
AccessisavailabletoanyonewantingtoRECALL
NONͲPATIENTRESULTSfrommemoryorINSTALL
CODECHIP.
2.OutͲofͲRangeFlags
Page75of86
Ifapatientisoutsidethenormalrangeforanyoftheanalytes,thentheresultwillbeflaggedby
reversevideo,thatis,thebackgroundoftheresultwillbeblackandthevalueinwhitenumbers.A
warningwillbedisplayedbelowtheresults:
PATIENTRESULTABNORMAL
IftheInternalQCzonesforthedeviceoraspecificanalyteareunacceptable,theanalyte(s)in
questiondisplayanexclamationpoint(!)inplaceofavalueandawarningisdisplayed:
!
INTERNALQCOUTOFRANGE.
IftheQCSampleforaspecificanalyteonthatdevicelotisunacceptable,theanalyteinquestionwill
displayapoundsign(#)inplaceofavalueandawar ningisdisplayed:
#
QCSAMPLEOUTOFRANGE.
IfbothQCSampleandInternalQCerrorsarepresentforaspecificanalyte,bothapoundsignandan
exclamationpoint(#
!)willbedisplayedinplaceofavalueandbothwarningsaredisplayed:
!
INTERNALQCOUTOFRANGEand#QCSAMPLEOUTOFRANGE.
3. BlockedResults
WhenQCerrorconditionsexistonatestdevice,whetherforaspecificanalyteortheentiretestpanel,
theanalytesinquestionareblockedfromviewandreplacedwithanOutofRangeFlag(!and/or#).The
errorconditionmustberemediedandthe
testrepeatedtoobtainaresult.
4. QCSampleFrequency
Howoftenanexternalliquidcontrol, QCSample,shouldberuncanbeselectedunderSET
PARAMETERS.Ifthetimeintervalhasexpired,QCSamplesmustberunandacceptableresults
obtainedbeforeapatientsamplecanbetested.
AlistoftestdevicelotsandthedatethattheQCSampleexpirescanbeviewedandprintedby
selectingREAGENTLOTSQCunder
theRECALLRESULTSmenu.
5. SupervisorAccess
Topreventmisuse,theSupervisorCODECHIPmoduleisrequiredforaccesstotheSETPARAMETERSand
DELETERESULTSfunctions ofthesoftware.TheSupervisorCODECHIPisuniversalandmaybeusedonany
QuidelTriageMeterPro.
LimitedWarranty.FORTHEAPPLICABLEWARRANTYPERIOD,
QUIDELWARRAN TSTHATEACHPRODUCT
SHALLBE(I)OFGOODQUALITYANDFREEOFMATERIAL DEFECTS,(II)FUNCTIONINACCORDANCEWITHTHE
MATERIALSPECIFICATIONSREFERENCEDINTHEPRODUCTMANUAL,AND(III)APPROVEDBYTHEPROPER
GOVERNMENTALAGENCIESREQUIREDFORTHESALEOFPRODUCTSFORTHEIRINTENDEDUSE
(the“LIMITEDWARRANTY”).IF
THEPRODUCTFAILSTOMEETTHEREQUIREMENTSOFTHELIMITED
WARRANTY,THENASCUSTOMER’SSOLEREMEDY,QUIDELSHALLEITHERREPAIRORREPLACE,ATQUIDEL'S
DISCRETION,THEPRODUCT.EXCEPTFORTHELIMITEDWARRANTYSTATEDINTHISSECTION,QUIDEL
DISCLAIMSANYANDALLWARRANTIES,EXPRESSORIMPLIED,INCLUDINGBUTNOTLIMITEDTO,
ANY
WARRANTYOFMERCHANTABILITY,FITNESSFORAPARTICULARPURPOSEANDNONͲINFRINGEMENT
REGARDINGTHEPROD UCT.QUIDEL'SMAXIMUMLIABILITYWITHANYCUSTOMERCLAIMSHALLNOT
EXCEEDTHENETPRODUCTPRICEPAIDBYCUSTOMER.NEITHERPARTYSHALLBELIABLETOTHEOTHER
PARTYFORSPECIAL,INCIDENTALORCONSEQUENTIALDAMAGES,INCLUDING,WITHOUTLIMITATION,LOSS
OFBUSINESS,PROFITS,DATAORREVENUE,EVENIFAPARTYRECEIVESNOTICEINADVANCETHATTHESE
KINDSOFDAMAGESMIGHTRESULT.

TheLimitedWarrantyaboveshallnotapplyiftheCustomerhassubjectedtheProduct tophysicalabuse,
misuse,abnormaluse,useinconsistentwiththeProductManualorInsert,fraud,
tampering,unusual
physicalstress,negligenceoraccide nts.AnywarrantyclaimbyCustomerpursuanttotheLimitedWarranty
shallbemadeinwritingwithintheapplicableLimitedWarrantyperiod.
Page76of86
Appendices
AppendixASpecifications
Physical
Size 8.5”x6.25“x2.75”(22.5cmx19cmx7cm)Dx Wx H
Weight 1.5pounds(0.7kg)withoutbatteries
Electrical 6vDCat1ampsuppliedvia4AAbatteriesorAC/DCConverter
Keypad Numericwithspecialfunctionkeys(22total)
Environmental
Temperature 15ºC30ºC
Humidity 10%85%
Location Dry,clean,flathorizontalsurfaceawayfromdirectsunlight
Optical
LightSource LaserDiode<1milliwatt
LaserClassification 1
Detector SiliconPhotodiode
MemoryCapacity
750PatientRecords 32ReagentLotCODECHIPmodules
200QCSampleResults 200QCSampleCODECHIPmodules
70QCDeviceResults 4QCDeviceCODECHIPmodules
250MiscTestResults
600UserID's
Miscellaneous
RSͲ232computerinterfaceport
ThermalPrinter
LCDscreendisplay
Options
ExternalBarCodeReader(rapidentryofUser,Patient,AuxiliaryorMiscTes tID's).
QuidelTriageCensusDataManagementSoftware.
QuidelTriageCensusDataManagementSoftwarewithLISConnectivity.
TestDeviceSpecifications
Testdevicespecificinformationisprovidedintheapplicableproductinsert.Theinstrumentanalyzes
immunoassaytestdevicesmanufacturedbyQuidelCardiovascularInc.
Page77of86
AppendixBTroubleshooting
MeterSymptoms
Symptom ProbableCause CorrectiveAction
InvalidResults
a. Inappropriate bodyfluidor
improperlyanticoagulatedsample.
a. Retestsampleusinganewtestdevice
withaproperspecimen.
b. Samplesizeincorrect—toolittle
sample.Testdevice’sinternalQC
controlswillbeoutofrange.
b. Retestsampleonanewtestdevice.
Makesuretheentiretubeisfull.Be
certainthelowerbulbonthepipette
tubecontainsasmallamountfluid
beforeyoupressthetop
bulb.
c. Meterwaspickeduporcarried
verticallywhiletestwasrunning.
Thetestdevice’sinternalQC
controlswillbeoutofrange.
c. Iflessthan30minuteshaspassedsince
preparingthetestdevice,placemeter
onflatsurfaceandreinsertthetest
device.Ifmorethan30
minuteshas
passed,retestsampleonnewtest
device.
Meterwillnotpull
testdevicein.
Hookthatpullstestdeviceintothe
metermaybebroken.
Ifmeterrepeatedlyfailstopullthedevice
in,themetermayneedtobereplaced.
ContactQuidel.
CODECHIPmodule
willnotfitinmeter.
a. CODECHIPmodulemayhavebeen
insertedupsidedown.
a. ReͲinsertCODECHIPmodulerightside
up.
b. CODECHIPmoduleorMeter
contactscouldbebent.
b. TrynewCODECHIPmodule,if
available.Ifstillnotagoodfit,contact
Quidel.
Printerworksbutno
printappearson
paper.
Papermaybeinstalledbackwards. Installpaperreversetocurrentposition.
Printerjam.
Papermayhavebeenfedintothe
meterimproperly.
Carefullyremoveoldpaperoutofthe
printer.Donotinsertanytoolsintothe
printer.
Printerstops
printing
orskipsaline.
Batterypowerlow.
Replacebatteriesorverifypowercordis
installedinthemeterandwalloutlet.
Meterwillnot
poweronormeter
powersoffwhen
runningatest.
a. Batteriesnot installed,drainedor
incorrectlyaligned.
a. Verifybatteriesareinstalled and
correctlyaligned(+)and(Ͳ);replace
batteriesifnecessary.
b. Powersourcenotpluggedin.
b. Verifypowercordisinstalledinthe
meterandwalloutlet.
Blinkingbattery
icon.
Lowbatterypower. Replacebatteries.
Metercontainsno
batteriesor
batteriesarelow
whenfacilityhas
lossofpower.
AlldataissavedintheMeterexceptdate
andtime.Resetdateandtime.
Page78of86
AppendixBTroubleshooting
MeterMessageorSymptom
Message ProbableCause CorrectiveAction
>XXX.X
Patientsampleconcentrationisabove
thedynamicrange.
Seeinstructionsthatcameintheboxof
testdevicesorcontactQuidel.
<XX.X
Patientresultisbelowthedynamic
rangeorisaconcentrationbelowthe
statisticallydeterminedlowesttest
concentration.
Forsomeparameters,a0.0ng/mlpatient
resultmaybeappropriate.
BatteryLow. Batteriesneedreplacement.
Replacewith4ͲAAbatteriesorusetheAC
poweradapterasapowersource.
CannotReadBar
Code
Damagedbarcodeormeterhookmay
bebroken.(Themeterhookpullsthe
testdeviceintothemeter.)
Lookforparticlesonthebarcodeofthe
testdevice.wipeoffwithadrycloth.
Repeatthetest.
CannotRead
CodeChip.
a. CODECHIPmoduleerror.
a. EnsureCODECHIPmoduleis
completely inserted.
TryareagentCODECHIPmodulefrom
adifferentboxofdevicesofthesamelot
number.
IfanewCODECHIPmoduleeliminates
themessage,theoriginalCODECHIP
isat
fault,discardCODECHIPmodule.
b. Metererror
b. If thenewCODECHIPmoduledoesnot
work,installaknownworkingCODE
CHIPmoduletoverifyproperoperation.
IftheCODECHIPmodulefailstowork,
contactQuidel.
DetectorFailure
OpticsDetectordidnotproperlypower
on.
Powermeteroffandwait15seconds.
Powermeteron.Ifmessagefailstoclear,
contactQuidel.
Measurement
Failure.
a. Inappropriate bodyfluidor
improperlyanticoagulatedsample.
a. Retestsampleusingnewdevicewitha
properspecimen.
b. Samplesizeincorrecttoolittle
sample.
b. Retestsampleonnewdevice.Becertain
thelowerbulbonpipetcontainsasmall
amountoffluidandtheentiretubeisfull
beforedispensingcontents.
c. Meterpickeduporcarriedvertically
whiletestisrunning.
c. Retestsampleandleavemeteronbench
topwhileperformingassay.Ifmorethan
30minutessinceinoculation,useanew
device.
d. Samplerunningslowly
d. Possibly duetocolddevicesorsample,
highhematocritwithawholeblood
sample,ormechanicalissuewithdevice.
Havecustomerimmediatelyreinsert
deviceintometer(ifwithin30minutes
ofsampleaddition).
Page79of86
Message ProbableCause CorrectiveAction
MotorFailure.
Motorceasestofunction.
Metermaysloworstopduetoobject
lodgedontheMetertrack.
Testdevicemaystickonthemeter
track.
Ejectandinspectthedeviceforanysticky
substances,labelsontopofthedeviceor
otheritemsthatmayinterferewith
the
meter.Ifproblempersists,contact
Quidel.
NoDeviceDatain
Memory.
ReagentCODECHIPmodulehasnotbeen
installed.
InstalltheCODECHIPmoduleincludedin
theboxoftestdevices(theReagent
CODECHIP
module).
OpticFailure.
a. Internalstandardisoutof
specification.
a. Powermeteroffthenonagain.If
OPTICFAILUREmessageclears,run
theQCDevicetoverifyproper
operation.Ifmessagefailstoclear,
contactQuidel.
b. Meterexposedtoextreme
temperatures.
b. Powermeteroff.Allowmetertositat
roomtemperaturefor1–2hours.
Powermeterbackon.
ProgramCheckOK.
PressENTERtostart
existingprogramor
pressEXITtoloada
newprogram.
MessageappearswhentheON/OFF
buttonispressedwhenthemeterhas
beenwithoutpower.
PresstheENTERkey.
QCDevice—
calibration,
alignmentorlaser
failed.
Messages:
CalibPass(orFail)
LaserPass(orFail)
AlignPass(orFail)
a. Maybeduetolintordustaffecting
reading.
a. CleanQCDevicewithalintͲfreecloth
orspraywithcannedairandrepeat
thetest.
b. QCDevicewasnotrunregularly.
b. QCDeviceshouldberunregularly
evenifpatientsamplesarenotrun.
c. Ifproblempersists,contactQuidel.
Time/DateisblinkingMetertemporarilylostallpower.
Resettimeanddateifrequired.Check
theAC/DCpowerconverterplugand
batteries.
Warning:InternalQC
OutofRange
Warningindicatesthattheinternal
qualitycontrolonthetestdevicewasout
ofrange.
Thismaybesamplespecific.
Thismaybedevicerelated.
Repeatsampleusinganewtestdevice.If
problempersists,contactQuidel.
Warning:Patient
MemoryhasSpace
for###More
Records.
Messageisdisplayedwhenmeteris
turnedonifthereisspaceleftforless
than100patientrecordsorspaceleftfor
lessthan20patientrecordswhen
runningpatientsample.
Options:
a. Printallresults.
b. Deletepatientresults.
c.
No actionisrequiredasthemeter
willdeletetheoldestresultoncethe
memoryisfull.
Page80of86
AppendixCReturnandDisposalProcedures
ReturnProcedure
Shouldamalfunctionoccur,contactQuidel.IfQuideldeterminesthatthemetershouldbereturned,youwill
beassignedareturnauthorizationnumberandbeprovidedshippinginstructions.
Immediatelyafterthereturnhasbeenauthorized,QuidelwillsendoutareplacementQuidelTriage
MeterPro.Returnthemalfunctioning metertoQuidel.
Notethereturnauthorizationnumberonboththe
shippingboxandtheairbillandsendthemeterbacktoQuidelassoonaspossiblefollowing receiptofthe
replacementmeter.
DisposalProcedure
Theproductmaycomeintocontactwithbloodduringtesting.Usedproductsthereforecarryariskof
infection.Whendisposingofameterthathasbeenutilized, pleasedosoinaccordancetotheregulations
applicableinyourcountry.Forinformationaboutcorrectdisposal,pleasecontactyourlocalcouncilor
authority.
TheproductfallsoutsidethescopeoftheEuropeanDirective2002/96/EC(directiveonWaste
ElectricalandElectronicEquipment(WEEE)).
AppendixDGlossary
Analyte:Chemicalsubstancebeingmeasured.
Assay:Test.
Calibration:Comparisonofresulttoareferencestandard.
Character:Onepartofalongerstringoflettersandnumbers;forexample,inthestring"ABC",theA,B,
andCareeachacharacter.
CODECHIPModule:Anyofseveralkindsofchipsbearinginformationthatcanbedownloadedinthe
QuidelTriageMeterPro,includingReagent,QCSample,QCDevice,SupervisorandProgram.
CODECHIPModulePort:ThesmallslotontheunderneathsideoftheQuidel TriageMeterProwherea
CODECHIPmodulecanbeinserted.
Dateformats:DDͲMMͲYY:DayͲMonthͲYear
MMͲDDͲYY:MonthͲDayͲYear
YYͲMMͲDD:YearͲMonthͲDay
DeviceL/N:Lotnumberofatestdevicecanbefoundonthesideoftheboxthatcontainsthetestdevices
andoneachtestdevice.ThedeviceL/N
isafiveͲdigitnumberthatmaybeprecededbyoneormoreletters.
Themeterdisregardsallletters.
Digit:Onepartofalongernumber;forexample,inthenumber231,the2,3,and1areeachadigit.
Fluorescence:Thecharacteristicofachemicalsubstancethatenablesit
togiveofflightwhenstimulated.
NOTE:AllpartsofdatesshouldbegivenastwoͲdigitnumbers.Forexample,themonthofMayis05.
Page81of86
ID:Identification.
Immunoassay:Atestthatusesantibodiestomeasuresubstances.
InternalQC:QualitycontrolsinplaceaspartoftheQuidelTriageMeterPro’ssoftwareorzonesbuiltinto
thetestdevice.
Invitro:Madetooccurinalaboratoryvesselorothercontrolledexperimentalenvironmentratherthanin
a
livingorganism(literally,inglass).
Invitrodiagnosticuse:Foruseinacontrolledsetting.
Laserdiode:Light sourceusedintheQuidelTriageMeterPro.
LCD:Liquidcrystaldisplay;thescreenon theQuidelTriageMeterPro.Showsthemenuofpossibletestsand
proceduresandpromptstheoperatortotakethe
nextstep.
Lotnumber:L/N;identificationnumbergiventoagroupoftestdevices.
L/N:Lotnumber;thenumberthatidentifiesabatchof
testdevices.
Menu:ListofchoicesdisplayedontheQuidelTriage
MeterPro’sscreen.
MMX:TheMultiMarkerIndexvalue,isanumber
calculatedfromthemeasured
valuesoftheanalytesin
aTestDevice.NotallQuidelTriagetestpanelshave
theMMXfeature.
Parameters:Options; functions
Photodiode:DetectsthelightemittedfromthefluorescentdyeinthetestdevicesusedintheQuidelTriage
MeterPro.
PointofCare(POC):Notinalaboratory;attheplace
wherecareisbeinggiventoapatient.
QC:QualityControl.
QCSample:Externalcontrols;aliquidsolutioncontainingchemicalsthat,whendispensedintoatest
device,reactwiththetestreagents.Usedtoverifyproperperformanceoftestdevicesaftershipmentor
longperiodsofstorage.
QCSampleCODE
CHIPmodule:CODECHIPmodulethatcanbedownloadedintotheQuidelTriage
MeterPromemorytogivethemeterinformationneededaboutthespecificlotofQCSample.
QCDevice(previouslycalledQCSimulator):
Testdevicecontaining6independentzonesthatverifiesproperoperationofthe
QuidelTriageMeterPro’sabilityto
properlyreadpatienttestsacrossvaryinganalyte
concentrations.
QCDeviceCODECHIP:CODECHIPmodulethatcanbedownloadedintotheQuidel
TriageMeterPro’smemorytogivethemeterinformationneededaboutthespecificQCDevice’s
performanceintheassignedmeter.
Page82of86
Quantify,quantifying:Togivetheresultsofatestasnumber.
Reagent:Substancethatcauseschemicalreactions;usedinanalysis.
ReagentCODECHIPmodule:CODECHIP
TM
moduleloadedwithinformationaboutaspecifictestandlot
number.
Retrieve:Find,bringback.
ReverseVideo:Lighttextdisplayedonadarkbackground.
Screen:ThepartoftheQuidelTriageMeterProthatdisplayswordsandnumbers.
Scrollto:Usearrowkeys


togototheinformationyouwant
onthescreen.
SetParameters:AfunctionoftheQuidelTriageMeterPro;requirestheuseoftheSupervisorCODECHIP
module.
SoftwareUpgradeCODECHIP:
CODECHIPmoduleusedtodownloadnewsoftwareontotheQuidelTriageMeterPro.
SupervisorCODECHIPmodule:TheCODECHIPmodulethatallowsasupervisor
toset
parametersanddeleteresults.
Testdevice:Holdsthesample(for example,blood)fromthepatientsothatitcanbeinserted
intotheQuidelTriageMeterProfortesting.Containsreagentzonesthatcausereactionsto
occurthatallowthemetertodetermineanalytelevelsinthepatient.
Testpanel:
AlternatenameforTestDevice,usuallyindicativeofaTestDevicewithmultipleanalytes.
AppendixESampleLogSheets
Thefollowinglogsheetsareprovidedasalternatemethodsoftrackingtestresults.
TheQuidelTriageMeterProiscapableoftrackingandprintingouttesthistoryandprovidingQCLockouts
topreventpatienttestingwhenoutsideQCrequirements.

Page83of86
Page84of86
Page85of86
Page86of86

55070QuidelTriageMeterPro
55071QuidelTriageMeterPro

MDSSGmBH
Schiffgraben41
30175Hannover,
Germany
QuidelCardiovascularInc.
9975SummersRidgeDrive
SanDiego,CA92121USA
quidel.com
ENSRC26611enA
PN:26611enRev.A2018/03
RevisionChanges:
InitialreleaseforQuidelCardiovascularInc.