Mem
C / F
PRO 6000
Ear thermometer
Directions for use
English
1
This manual applies to 901054 Ear Thermometer, 901009 Accessory, Thermometry
and 901010 Accessory, Thermometry
06000-200 06000-300
06000-100 06000-150
06000-125 104894
These instructions for use/directions for use (IFU/DFU) may contain information about products
that may or may not be approved for use by a relevant regulatory authority in any particular
country or region of the world. Customers and/or End-Users are requested to contact their
local sales representative for further information regarding regulatory registration status and
availability of products.
Welch Allyn Limited
Navan business Park
Dublin road,
Navan, County Meath
Republic of Ireland
C15 AW22
For information about any Welch Allyn product, contact:
Welch Allyn Technical Support:
www.welchallyn.com/support
visit locations:
www.welchallyn.com/about/company/locations.htm
Replacement parts
Probe covers: 06000-005, 06000-801, 06000-800
For a complete list of parts, go to www.welchallyn.com
For patent information, visit
www.welchallyn.com/patents
© 2019 All rights reserved.
Made in Mexico
This product is manufactured under license to the ‘Braun’ trademark.
‘Braun’ is a registered trademark of Braun GmbH, Kronberg, Germany.
ThermoScan and ExacTemp, are trademarks of Helen of Troy Limited and/or its affiliates.
Duracell is a registered trademark.
ThermoScan
®
PRO 6000 Ear thermometer
Manufactured by:
Welch Allyn, Inc.
4341 State Street Road
Skaneateles Falls, NY 13153
USA
107983 (CD)
DIR 80024839 Rev C
Revision date: 2019-
10
Use only
Hillrom
probe covers
English
2
1. Braun Thermoscan
®
PRO 6000 Ear thermometer
Mem
C / F
15
16
MemC / F
MemC / F
1
12
11
13
14
2
3
4
10
6
7
9
8
5
Mem
C / F
Mem
C / F
18
18
19
18 a
17
20
English
3
2. Package components
Braun Thermoscan
®
PRO 6000 Ear thermometer
Cradle
Probe covers (1 or 2 probe cover boxes, depending on model)
CD containing Directions for use, Quick reference guide, Welch Allyn Service Tool software
(available via internet link) and Service Tool installation guide (available via internet link).
2 (AA) Duracell
®
alkaline batteries
3. Product description
(See Section 1. Braun Thermoscan
®
PRO 6000 Ear thermometer)
1
Probe lens window
2
Probe
3
ExacTemp
light
4
Measure button
5
Measure light
6
Display
7
C/F button
8
Memory button
9
Timer button
10
Tether mount (Tether sold separately)
11
Probe cover detect switch
12
Probe cover ejector
13
Battery door latch
14
Battery door
15
Small cradle - one storage box
16
Large cradle - two storage boxes
17
Probe cover
18
Probe cover box
18a
Probe cover box carrier
19
GTIN code
20
Temperature scale switch (inside battery
compartment)
English
4
4. Screen elements
Battery
Full battery—indicates battery
is between 100 % and 70 % of
usable battery capacity
Partial battery—indicates
battery is between 70 % and
30 % of usable battery capacity
Low battery—indicates battery
is between 30 % and 10 % of
usable battery capacity
Critically low battery—battery
is between 10 % and 1 % of
usable battery capacity.
When final segment flashes,
batteries have low power.
The thermometer will take a
proper measurement but
batteries must be
replaced soon. If rechargeable
batteries are in use, batteries
should be recharged.
Empty battery—battery has
1 % or less of usable battery
capacity. When battery outline
flashes, the thermometer will
not operate. Replace the
batteries. If rechargeable
batteries are in use, batteries
should be recharged. See 14.7
Maintenance and service,
Replacing the batteries
Probe cover icon
The icon animates in an upward
motion to remove probe cover.
The icon animates in a downward
motion to apply a probe cover.
See 9. Using the Braun Thermoscan®
PRO 6000 Ear thermometer
 
Timer icon
The Braun Thermoscan
®
PRO 6000 Ear
thermometer includes a 60 second
timer that features an audible
notification and visual indicator at 0,
15, 30, 45 and 60 seconds. The first
quadrant begins flashing when
timer starts and becomes solid at
15 seconds. This repeats for each
15 second interval. The timer turns
off automatically 5 seconds after
60 seconds is completed. See 10.3
Controls, Manual timer
Alert icon
Icon that appears with error message.
See 12. Errors and notifications
Memory indicator
Indicates the reading shown on the
display is the reading in memory.
See 10.1 Controls, Memory
C/F scale
Indicates default temperature scale.
Either °C or °F will display, depending
on setting. See 10.2 Controls, C/F
Celsius/Fahrenheit
English
5
Security unlock icon and countdown time
(Requires charging station or compatible Welch Allyn Vital Signs device,
sold separately.) If the security function is enabled, it requires the thermometer be
returned to the charging station within a pre-selected time interval. Countdown time
indicates the amount of time remaining until the thermometer will be locked if not
returned to the charging station. See 11.2 Settings, Advanced functions
Security lock icon
(Requires charging station or compatible Welch Allyn Vital Signs device,
sold separately.) Indicates the thermometer is locked. Return the thermometer to
the charging station to reset the countdown and resume normal operation. See 11.2
Settings, Advanced functions
5. About the Braun Thermoscan
®
PRO 6000 Ear thermometer
5.1 Intended Use
The Braun Thermoscan
®
PRO 6000 Ear thermometer is indicated for the intermittent
measurement of human body temperature for patients having ages ranging from normal
weight (full term) newborn to geriatric adults in a professional use environment. The probe
cover is used as a sanitary barrier between the infrared thermometer and the ear canal.
5.2 How does Braun ThermoScan work?
The Braun Thermoscan
®
PRO 6000 Ear thermometer technology
reads the infrared energy emitted by the tympanic membrane and
surrounding tissues to determine the patient’s temperature. To help
ensure accurate temperature measurements, the sensor itself is
warmed to a temperature close to that of the human body. When the
Braun ThermoScan is placed in the ear, it continuously monitors the
infrared energy until a temperature equilibrium has been reached and
an accurate measurement can be taken. The thermometer displays an
actual measured ear temperature or clinically accurate, oral equivalent
temperature that has been validated in clinical studies by comparing IR
measurements with the oral readings from afebrile and febrile patients
of various ages. Unadjusted operating mode ear temperature readings
are available via unadjusted mode, which can be accessed using the
Welch Allyn Service Tool.
English
6
5.3 PerfecTemp™ Sensor System
Speed and ease of access are two key advantages of ear
thermometry. Concerns regarding accuracy and reliability
have hindered adoption of this technology. In clinical studies,
the precision of ear temperature measurement has been shown
to be influenced by ear canal anatomy and variability in user
technique. Proper probe placement can also be a challenge,
particularly with young patients that move around during
measurement. Shallow probe placement, coupled with anatomical
variabilities such as ear canals with a small circumference and poor
visibility of the tympanic membrane, can result in readings that are
low compared to core temperature because the thermometer could
be focused on the cooler outer ear canal.
The Braun Thermoscan
®
PRO 6000 Ear thermometer incorporates a new proprietary sensor
system, PerfecTemp™, that overcomes the challenges presented by ear canal anatomy and
variations in technique among clinicians. The thermometer collects information about the
direction and depth of ear probe placement as it is placed into the ear canal and automatically
incorporates this information into the temperature calculation. Incorporating information
related to patient specific anatomy and exact ear probe placement in the ear canal increases
the accuracy of the measurement as compared to core temperature, especially when probe
positioning is not ideal.
5.4 ExacTemp™ technology
The Braun Thermoscan
®
PRO 6000 Ear thermometer also features ExacTemp™ technology
which supports temperature measurement reliability by detecting the stability of the probe
placement during the measurement. The ExacTemp light flashes during the measuring
process and remains illuminated when the measurement is complete, indicating consistent
placement of the probe during the measurement process. Consistent probe placement helps
support accurate temperature measurement.
5.5 Why measure in the ear?
Clinical studies have shown that the ear is an excellent site for
temperature measurement because temperatures taken in the ear
reflect the body’s core temperature
1
. Body temperature is regulated
by the hypothalamus
2
, which shares the same blood supply as
the tympanic membrane
3
. Changes in core body temperature are
usually seen sooner at the tympanic membrane than at other sites,
such as the rectum, mouth or under the arm. Advantages of taking
temperatures at the ear versus traditional sites:
Axillary temperature measurements reflect skin temperature
which may not reliably indicate the internal body temperature.
Rectal temperatures often lag significantly behind internal body temperature changes,
especially at times of rapidly changing temperatures. Also, there is a risk of cross
contamination.
Oral temperatures are often influenced by eating, drinking, thermometer placement,
breathing through the mouth, or the inability of the person to close their mouth completely.
1. Guyton A C, Textbook of medical physiology, W.B. Saunders, Philadelphia, 1996, p 919
2. Guyton A C, Textbook of medical physiology, W.B. Saunders, Philadelphia, 1996, p 754–5
3. Netter H F, Atlas of Human Anatomy, Novartis Medical Education, East Hanover, NJ, 1997, pp 63, 95.
Table 1: Ear canal
temperature gradient
English
7
5.6 Body temperature
Normal body temperature is a range. The following table shows that
this normal range also varies by site. Therefore, measurements from
different sites, even if taken at the same time, should not be directly
compared.
Normal ranges by site
1
:
Axillary
1, 2
: 95.6–99.4 °F 35.3–37.4 °C
Oral
1, 2
: 95.7–99.9 °F 35.4–37.7 °C
Rectal
1, 2
: 96.6–100.8 °F 35.9–38.2 °C
ThermoScan
1, 2
: 95.7–99.9 °F 35.4–37.7 °C
A person’s normal temperature range tends to vary with age.
The following table shows normal ThermoScan ranges by age.
Normal ThermoScan ranges by age
1, 2
:
< 3 months 96.4–99.4 °F 35.8–37.4 °C
3–36 months 95.7–99.6 °F 35.4–37.6 °C
> 36 months 95.7–99.9 °F 35.4–37.7 °C
The range of normal varies from person to person and can be influenced by many factors
such as time of day, level of activity, medications and gender.
1. Sund-Levander M, Forsberg C, Wahren LK. Normal oral, rectal, tympanic and axillary body temperature in adult men
and women: a systematic literature review. Scand J Caring Sci 2002 June;16(2):122–8.
2. Herzog L, Phillips SG. Addressing concerns about fever. Clin Pediatr (Phila) 2011 May;50(5):383–90.
English
8
6. Contraindications
None
6.1 What affects accuracy
Always use a new disposable probe cover for each measurement to maintain accuracy
and hygiene. The right ear measurement may differ from the measurement taken in the
left ear. Therefore, always take the temperature in the same ear. The ear must be free from
obstructions or excess earwax build-up to take an accurate reading.
External factors that may influence ear temperatures include:
Factor Affected Not affected
Used probe cover
Ambient temperature
Wet / dirty / damaged lens
Hearing aid
Lying on pillow
Moderate cerumen
(ear wax)
Otitis media (ear infections)
Tympanostomy tubes
In the event the patient is lying on a pillow or wearing ear plugs or a hearing aid, remove the
individual from the situation and wait 30 minutes prior to taking a temperature.
7. Warnings and precautions
WARNING This thermometer is for professional use only.
WARNING This thermometer must only be used with genuine Hillrom probe covers.
WARNING Do not use any cleaning agent other than isopropyl or ethyl alcohol for
cleaning the probe lens window and probe as specified in the cleaning section of this
manual.
CAUTION Do not use any cleaning agent other than those on the approved list of
cleaners for cleaning the body of the thermometer.
WARNING To avoid inaccurate measurements, always attach a new, clean probe cover
for each temperature measurement.
English
9
WARNING The probe lens window must be kept clean, dry and undamaged at all
times to ensure accurate measurements. To protect the probe lens window, always
keep the thermometer in the storage cradle while transporting or when not in use.
WARNING This thermometer is not intended for pre-term babies or small-for-
gestational age babies.
WARNING Do not modify this equipment without authorization of manufacturer.
CAUTION Never use the thermometer for purposes other than those it has been
intended for. Please follow the general safety precautions.
CAUTION Do not expose the thermometer to temperature extremes (below –25 ºC /
–13 ºF or over 55 ºC / 131 ºF) nor excessive humidity (> 95 % RH).
CAUTION This thermometer complies with current required standards for
electromagnetic interference and should not present a problem to other equipment
nor is it affected by other devices. As a precaution, avoid using this device in close
proximity to other equipment.
WARNING Do not use an ear thermometer if there is blood or drainage in the external
ear canal.
WARNING An ear thermometer should not be used on a patient who exhibits
symptoms of an acute or chronic inflammatory condition of the external ear canal.
WARNING Common situations like the presence of moderate amounts of cerumen
(ear wax) in the ear canal, otitis media and tympanostomy tubes do not significantly
impact temperature readings. However, complete ear canal occlusion due to cerumen
(ear wax) can result in lower temperature readings.
WARNING If prescription ear drops or other medications have been placed in the ear
canal, use the untreated ear to take a measurement.
WARNING Patients who have deformities of the face and/or ear may not be able to
have a temperature taken with an ear thermometer.
8. Setup
8.1 Battery installation
Your Braun Thermoscan
®
PRO 6000 Ear thermometer ships with two (AA) alkaline batteries.
See 14.7 Maintenance and service, Replacing the batteries.
The Braun Thermoscan
®
Charging station (sold separately) ships with one rechargeable
battery pack.
8.2 Mounting instructions (Large cradle only)
Mounting hardware not included.
English
10
The large cradle (2 box storage) can be mounted as an easily removable wall hanger or a
permanent wall mount. All mounting must be done into a wall stud. To mount the cradle the
following items are needed:
2 #8 wood or sheet metal, pan head screws, 3.2 cm (1.25") long
Ruler (or tape measure can be substituted)
Screwdriver to match drive of screws
Remove probe cover box carrier from the cradle
by rotating the probe cover box carrier forward.
Mount to wall:
Removable wall hanger:
Locate the stud on the wall. Mount the 1st screw in
position 1 and the 2nd screw in position 2a.
Permanent wall mounting:
Locate the stud on the wall. Mount the 1st screw in
position 1 and the 2nd screw in position 2b. Tighten
screws.
Note: Permanent mounting is not recommended if using
the charging station to access the security function, other
advanced functions or charge the rechargeable battery.
1
2a
2b
126 mm (4.95")
107 mm (4.21")
Place probe cover box carrier back in cradle by
lining up the brackets and pushing down.
Note: For a mounting template please reference the Braun
Thermoscan
®
PRO 6000 Ear thermometer CD.
English
11
8.3 Tether installation
A kit to tether the thermometer to the cradle is available separately.
Installation instructions are provided with the tethering kit. Contact Welch Allyn for details.
English
12
9. Using the Braun Thermoscan
®
PRO 6000 Ear thermometer
Temperature measurement
Remove thermometer from cradle by gripping
thermometer at base and pivoting up.
Thermometer will turn on automatically. Probe cover icon
will flash on display indicating a new probe cover is needed.
1a
1b
Attach new probe
cover by pushing probe tip
straight into box, then pulling
thermometer out.
Wait for the ready indication. Ring around measure
button turns green, thermometer will beep once and three
lines on display indicate thermometer is ready.
Place probe snugly in
ear canal and direct
toward opposite temple.
Keep thermometer probe steady
in the ear canal. Correct probe
placement is essential for accurate
measurements.
English
13
Press and release measure button.
The thermometer will beep once, running dashes will appear
on the display then the green ExacTemp light will flash,
indicating consistent probe position.
Note: Always press the measure button before taking a
measurement.
Temperature measurement. A long beep and
steady green ExacTemp light will signal the end of the
measuring process.
The temperature will show on the display.
If the thermometer is unstable or patient is moving during
the measuring process, the device will beep, the green
ExacTemp light will flash and POS (Position Error) will flash
on the display. Be sure the device is stable and restrict
patient movement for the next measurement. Change the
probe cover to reset.
* See 12. Errors and notifications
English
14
Remove used probe cover by pressing Probe Cover
Ejector button .
To achieve accurate measurements, use a new, clean probe
cover for each measurement.
To take another measurement, place a new, clean probe
cover on the thermometer. If no action is taken the
thermometer will enter sleep mode after 10 seconds or upon
return to the thermometer cradle or host device.
10. Controls
10.1 Memory
Press MEM (the memory button) to reveal the last completed
temperature. The temperature will show with a Mem indicator until
MEM (the memory button) is pushed again, a new probe cover is
applied, or the thermometer enters sleep state. Memory can also
be accessed from the thermometer sleep mode and will display for
5 seconds before returning to sleep.
MemC / F
10.2 C/F (Celsius/Fahrenheit)
Once the temperature scale has been set (See 11.1 Default temperature scale), the alternative
scale can be quickly referenced at any time while a temperature is displayed.
If the temperature scale is set to Celsius, press and release
C/F (the C/F button) to view the temperature in Fahrenheit.
If the temperature scale is set to Fahrenheit, press and
release C/F to view the temperature in Celsius.
Press and release C/F again to revert back to
default scale.
Note: If temperature conversion is disabled, refer to the
Service Manual for more information.
MemC / F
MemC / F
English
15
10.3 Manual timer
The Braun Thermoscan
®
PRO 6000 Ear thermometer includes a 60 second timer that features
an audible notification and visual indicator at 0, 15, 30, 45, and 60 seconds. The timer turns
off automatically 5 seconds after 60 seconds is completed. The timer can be stopped at any
time by pressing the Timer button or applying a probe cover. This feature can be used to time
pulse, respirations, etc. To use this feature:
Press and hold timer button for one second to activate
timer. A beep is heard at initiation of the timer.
The display will show timer counting up in seconds.
The display will show an icon that has four 15 second
quadrants.
The timer will beep at the completion of each 15 second
interval to provide audible notification. Then the current
segment will become solid and the next segment will flash.
At 60 seconds a long beep will sound, all quadrants will
appear solid ending the timer function. The thermometer will
exit the timer mode after an additional 5 seconds.
0 sec
“Beep”
15
“Beep”
30
“Beep”
45
“Beep”
60
“Beep”
To stop the timer at any time, press the timer button.
11. Settings
11.1 Default temperature scale
To set the default temperature scale:
Open battery door by sliding
spring latch to the right using
your finger or a pointed object
such as a pen. While holding the
latch in the open position, grasp
the battery door and remove it.
Remove batteries and set aside.
Once batteries are removed,
the C/F switch is accessible.
or
English
16
Slide switch to C or F using a pen or pointed object.
Place batteries back into thermometer. Snap the battery door
back into place and ensure the latch returns to the original
latched position. Celsius or Fahrenheit symbol will appear on
the display.
ºC ºF
11.2 Advanced functions
The Welch Allyn Service tool software is required to modify the thermometer’s configuration.
A charging station and rechargeable batteries or compatible Welch Allyn Device is required to
connect to the PC running the Welch Allyn Service Tool. (See 11.3 Advanced function settings
and 11.4 Service tools)
Item Description Settings
Default
setting
PerfecTemp™ Improves reading accuracy by
detecting placement of probe in
ear canal
On/Off On
C/F button Use the C/F button to view
measurements in the non-
default (alternate) temperature
scale. When off (disabled) only
the default scale is available.
On/Off On
Default C/F
manual switch
When on (enabled), the default
scale can be set using the
manual switch in the battery
compartment. When the control
is off (disabled), the radio
buttons for Celsius and
Fahrenheit are enabled allowing
the service tool to set the
default scale.
On/Off On
Security function Sets countdown time after
removal from charging dock
to lock
1 through 12 hours Off
Timer icon Displays an icon along with the
timer counter
On/Off On
Unadjusted
operating mode
Places thermometer in mode to
detect raw ear temperature only
Allows user to set
device to unadjusted
operating mode
Off
English
17
11.3 Advanced function settings
The Welch Allyn Service tool software is required to modify the thermometer’s configuration.
A charging station and rechargeable batteries or compatible Welch Allyn Device is required to
connect to the PC running the Welch Allyn Service Tool.
Follow the instructions to access Braun Thermoscan
®
PRO 6000 Ear thermometer advanced
settings using the Welch Allyn Service Tool.
Dock the Braun Thermoscan
®
PRO 6000 Ear thermometer in
the Charging Station
The recommended action is to use the USB cable that
connects to the wall adapter—unplug it from the wall adapter
and plug it into your computer.
a. Launch the Welch Allyn Service Tool.
b. If the startup screen with the Add new features”
and “Service” button appears click “Service”.
c. Log in as ADMIN with no password or use any
account you previously created. Note: If the log on
prompt does not appear click the Log on button.
You must be logged on to access the configuration
dialog.
English
18
Click the Braun Thermoscan
®
PRO 6000
Ear thermometer from the Device list
(Device list) to highlight, then click the
select button.
The device tab opens.
Click the Configure tab to the right of the
device information tab.
Click the Change button in the current
settings box. The configuration settings
dialog box opens.
English
19
Select the setting you wish to enable or
disable by clicking the check box next
to the setting. A check mark indicates
the setting will be enabled, an empty
check box indicates the setting will be
disabled. To select the Security function
click the drop down menu and click the
desired time or Off to disable. To restore
the factory default settings click Restore
Defaults. Once the desired settings are
selected click the Save button to send
the settings to the Braun Thermoscan
®
PRO 6000 Ear thermometer and close
the box.
To close the box without changing the settings click the Cancel button.
11.4 Service tools
For more information on the Service Tool and Service Tool installation guide
go to www.welchallyn.com and download the service tool found under the
Services & Support Tab/Service Centers/Download service tool.
11.5 Charging station for storage, charging and
security function (optional)
A charging station is available for the Braun Thermoscan
®
PRO 6000 Ear thermometer. The charging station automatically
recharges the thermometer when using the rechargeable battery
which is included. Use of alkaline batteries in the thermometer
while using the charging station is permitted but the alkaline
batteries will not be charged.
The charging station has an electronic, individually adjustable
security function which requires the thermometer to be returned
to the station within an individually pre-selected time or the
thermometer will be locked. The charging station serves as a
convenient storage base which can also be wall mounted.
Contact Welch Allyn for details.
Welch Allyn Technical Support:
www.welchallyn.com/support
visit locations:
www.welchallyn.com/about/company/locations.htm
English
20
12. Errors and notifications
Error message Situation Solution
RECHARGABLE: Battery icon is empty
and ashing, and lock icon displays to
prompt user to return to charge dock
ALKALINE: Battery icon
is empty and ashing
No probe cover is attached
(animates ON).
Attach new, clean probe cover.
RECHARGABLE: Battery icon is empty
and ashing, and lock icon displays to
prompt user to return to charge dock
ALKALINE: Battery icon
is empty and ashing
Used probe cover is attached
(animates OFF).
Discard probe cover that is
on and attach new, clean
probe cover if taking another
temperature measurement.
(POS = position error)
The infrared monitor cannot
find a temperature equilibrium
and allows no measurement.
Change the probe cover to reset.
Restrict patient movement and
ensure that the positioning
of the probe is correct and
remains stable while taking new
temperature.
Ambient temperature is not
within the allowed operating
range (10–40 °C or 50–104 °F)
or changing too rapidly.
Wait 20 sec. until thermometer
turns off automatically,
then turn on again. Ensure
thermometer and patient are in
an environment for 30 minutes
where the temperature is
between 10 ºC and 40 ºC or 50 ºF
and 104 ºF prior to measurement.
Temperature taken is not
within typical human
temperature range.
HI will be displayed when
temperature is higher than
42.2 °C (108 °F).
LO will be displayed when
temperature is lower than
20 °C (68 °F).
Change probe cover to reset.
Then, make sure thermometer is
properly inserted and take a new
temperature.
English
21
Error message Situation Solution
RECHARGABLE: Battery icon is empty
and ashing, and lock icon displays to
prompt user to return to charge dock
ALKALINE: Battery icon
is empty and ashing
RECHARGABLE: Battery icon is empty
and ashing, and lock icon displays to
prompt user to return to charge dock
ALKALINE: Battery icon
is empty and ashing
or
System error
(All icons display or display
is blank)
Wait 20 seconds until the
thermometer turns off
automatically, then turn on again.
If error persists, reset the thermometer by
removing the batteries and
putting them back in.
If error still persists, batteries are dead. Insert new
batteries.
If error still persists, contact local Welch
Allyn Service Center or
representative.
Battery is low,
but thermometer will still
operate correctly.
Insert new batteries.
Battery is too low to allow
temperature measurement.
Insert new batteries.
Do you have any further
questions?
contact local Welch
Allyn Service Center or
representative.
13. PerfecTemp™ status
Error message Situation Solution
PerfecTemp™ sensor system
is not functioning or disabled.
contact local Welch
Allyn Service Center or
representative.
U is “Unadjusted Operating
Mode”. Mode used for raw
temperature measurement.
Requires service tool access
to enable.
See 11.3 Advanced function
settings and adjust settings via
Service Tool or contact local
Welch Allyn Service Center or
representative.
English
22
14. Maintenance and service
14.1 Cleaning the probe lens window, probe and contacts
WARNING Only use Hillrom Thermometer disposable probe covers.
WARNING Do not use damaged, perforated, soiled or poor fitting probe covers. Do not reuse
probe covers.
WARNING Dirty probe lens window = lower readings. Fingerprints, cerumen, dust and
other soiling components reduce transparency of the tip and result in lower temperature
measurements.
If thermometer is placed in ear without a probe cover, clean immediately.
WARNING Do not damage probe lens window. Avoid touching probe lens window except
when cleaning. If probe lens window is damaged return to Welch Allyn for service.
CAUTION Do not modify, change or adjust probe lens window. These changes will affect
calibration and accuracy of the thermometer. If probe lens window is damaged return to
Welch Allyn for service.
CAUTION Do not use cleaning solution other than isopropyl or ethyl alcohol to clean probe
lens window and probe. Bleach and other cleaning agents will cause permanent damage to
the probe and probe lens window.
CAUTION Cleaning the probe lens window and probe
The thermometer probe and probe lens window should be cleaned when there are
fingerprints, cerumen, dust or other soiling components present, following the directions
below:
Remove probe cover and discard.
Moisten cotton swab or cloth slightly with isopropyl or ethyl
alcohol. Do not saturate.
English
23
Gently wipe surface of probe lens window with cotton
swab or cloth slightly moistened with isopropyl or ethyl
alcohol only.
Note: Use gentle pressure when cleaning the sensor to
prevent damaging the unit by accidently changing the
position of the sensor.
3.000SCALE
With probe facing down, wipe the probe with a damp cloth
or cleaning wipe moistened with isopropyl or ethyl alcohol.
Mem
C / F
Gently wipe with dry, clean cotton swab or cloth immediately.
Allow at least 5 minutes drying time before taking
temperature. Make sure probe lens window is clean and dry
prior to use.
Cleaning the contacts
Caution: Do not use bleach solutions of any kind when cleaning metal electrical contacts.
It will damage the device.
Slightly dampen a cotton swab with 70% isopropyl alcohol.
MINUTES
English
24
Remove the thermometer from the dock and clean the metal
electrical contacts on the thermometer.
Place the thermometer aside for 1 minute, allowing the
contacts to air dry.
1 MINUTE
Note: If any cleaning agent, other than isopropyl or ethyl alcohol, comes into contact with
the probe, probe lens window or contacts immediately wipe dry. Then clean the probe, probe
lens window and contacts with isopropyl or ethyl alcohol.
14.2 Cleaning the thermometer body and cradle
CAUTION Do not submerge the thermometer. Excess liquid may cause damage to the
thermometer.
Wipes should be moist, not saturated.
CAUTION Do not use any chemical other than those listed in the Approved Cleaning Solution
Table to clean the thermometer body and cradle. Other cleaning agents could cause damage
to the thermometer.
When cleaning the probe lens window or probe ONLY use isopropyl or ethyl alcohol.
CAUTION Do not use abrasives pads or cleaners.
English
25
Approved cleaning solutions
Family
Solution or
brand
Probe
lens
window
Probe Contacts
Thermometer
body & cradle
Tether
Chlorine
and
chlorine
compounds
10 %
Chlorine
bleach
solution
No No No Yes Yes
Quaternary
ammonium
compounds
CaviWipes™
Clinell
®
Universal
Wipes
SaniCloth
Metrex
No No No Yes Yes
Hydrogen
peroxide
Virox
Oxivir
No No No Yes Yes
Alcohol 70 %
isopropyl or
ethyl alcohol
Yes Ye s Yes Yes Yes
Additional cleaning agents may be periodically evaluated for compatibility. If your cleaning
agent is not listed, contact Welch Allyn to determine if additional cleaning agents are
approved for use.
As needed, clean the thermometer body and cradle following the below directions.
Note: If any cleaning agent, other than isopropyl or ethyl alcohol, comes into contact with
the probe, probe lens window or contacts immediately wipe dry. Then clean the probe, probe
lens window and contacts with isopropyl or ethyl alcohol.
For additional protection, we recommend placing a new
probe cover on the thermometer probe to protect this area
when cleaning the body of the thermometer.
English
26
Use a damp cloth or cleaning wipe with cleaning solution
from the Approved Cleaning Solutions Table. To clean
the body, ensure that the wipe is moist, not saturated.
Wipe the body with the display facing up.
Mem
C / F
Remove probe cover box carrier from the cradle by rotating
the probe cover box carrier forward. See 14.5 Removing and
installing probe cover box carrier.
Wipe the cradle and probe cover carrier with a damp cloth
or cleaning wipe with cleaning solution from the Approved
Cleaning Solutions Table.
Allow at least 5 minutes drying time before taking
temperature. Make sure probe, body and cradle are clean
and dry prior to use.
MINUTES
English
27
14.3 Cleaning the tether (sold separately)
To clean the tether, ensure that the wipe is moist,
not saturated. Wipe the tether with a damp cloth or cleaning
wipe with cleaning solution from the Approved Cleaning
Solutions Table.
14.4 Installing new probe cover box
Keep out of reach of children.
Remove empty probe cover box from probe cover box carrier
by pulling up.
Open the new probe cover box. Pull down on the perforated
strip. Discard perforated strips.
Insert new probe cover box into probe cover carrier by
placing inside brackets and pushing down.
English
28
14.5 Removing and installing probe cover box carrier
Remove thermometer from cradle by gripping thermometer
at base and pivoting up.
Remove probe cover box carrier from the cradle by rotating
the probe cover box carrier forward.
Place probe cover box carrier back in cradle by lining up the
brackets and pushing down.
14.6 Storage environment
Store thermometer and probe covers in a dry location (the thermometer is not protected
against ingress of water), free from dust and contamination and away from direct sunlight.
Storage temperature:
−20 to 50 °C (−4 to122 °F)
Storage humidity:
0 % to 85 % non condensing
Replacing the batteries
The thermometer is supplied with two 1.5 V batteries type AA (LR 6).
For best performance, Duracell
®
alkaline batteries are recommended.
Note: The battery life performance testing was based on the Duracell
®
alkaline batteries.
Batteries other than these are not guaranteed to provide the same life performance results.
English
29
Insert new batteries when the
battery symbol begins to flash on
the display (See 12. Errors and
notifications).
Open battery door by sliding
spring latch to the right using your
finger or a pointed object such
as a pen. While holding the latch
in the open position, grasp the
battery door and remove it.
Remove the batteries and replace with new batteries, making
sure the poles are in the right direction.
ThermoScan
Snap battery door back into place and ensure the latch
returns to original latch position.
This product contains batteries and recyclable electronic
waste. To protect the environment, do not dispose of it in
the trash, but take it to appropriate local collection points
according to national or local regulations.
14.7 Calibration testing
The thermometer is initially calibrated at the time of manufacture. If the thermometer is used
according to the directions for use, periodic readjustment is not required. However, Welch
Allyn recommends checking calibration on an annual basis or whenever clinical accuracy
of the thermometer is in question. Procedures for checking calibration are outlined in the
9600 Plus Calibration Tester (REF 01802-110) directions for use manual.
The above recommendations do not supersede the legal requirements. The user must
always comply with legal requirements for the control of the measurement, functionality,
and accuracy of the device which are required by the scope of relevant laws, directives or
ordinances where the device is used.
English
30
15. Specifications
Displayed temperature range: 20–42.2 °C (68–108 °F)
Operating ambient temperature range: 10–40 °C (50–104 °F)
Display resolution 0.1 °C or 0.1 °F
Accuracy for displayed temperature
range:
± 0.2 °C ± (0.4 °F) (35.0 °C–42 °C) (95 °F–107.6 °F)
± 0.3 °C ± (0.5 °F) (outside this temperature range)
Clinical bias: 0.09 °C (0.16 °F)
Limits of agreement: 0.58 °C (1.0 °F)
Clinical repeatability: 0.19 °C (0.34 °F) (Calculated per ASTM E:1965)
(less than the standard < 0.3 °C (0.57 °F)
per EN 12470-5)
Reference body site: Oral measuring
Site: Ear
Long term storage ranges
Storage temperature: −20 to 50 °C (−4 to122 °F)
Storage humidity: 0 % to 85 % non condensing
Shock: Withstands drop of 3 feet (91.44 cm)
Warm up time: Initial start-up time: 3–4 seconds
Measurement time: 2–3 seconds
Automatic power down: 10 seconds
Battery life: 6 months / 1000 measurements
Battery type: 2 × MN 1500 or 1.5 V AA (LR6)
Thermometer dimension: 6" × 1.7" × 1.3" (152 mm × 44 mm × 33 mm)
Thermometer weight: 3.6 oz (100 g) without batteries
Pressure: 700–1060 hPA (0.7–1.06 atm)
The thermometer is specified to operate at 0.7–
1.06 Atmospheric Pressure.
CAUTION: Do not use this device in the presence of electromagnetic or other
interference outside the normal range specified in IEC 60601-1-2.
English
31
Standards and compliance
This device conforms to the following Safety and Performance Standards:
This infrared thermometer meets requirements established in ASTM Standard E 1965–
98 (for the thermometer system [thermometer with probe cover]). Full responsibility for the
conformance of the product to the standard is assumed by Welch Allyn, Inc.
4341 State Street Road, Skaneateles Falls, NY, USA 13153.
ASTM laboratory accuracy requirements in the display range of 37 °C to 39 °C (98 °F to 102 °F)
for infrared thermometers is ± 0.2 °C (± 0.4 °F), whereas for mercury-in-glass and electronic
thermometers, the requirement per ASTM Standards E 667-86 and E 1112-86 is ± 0.1 °C
(± 0.2 °F).
This product conforms to the provisions of the EC directive 93/42/EEC (Medical Device
Directive).
A clinical summary is available upon request.
ANSI/AAMI STD ES60601-1, UL STD 60601-1, CAN/CSA STD C22.2 No. 60601.1, IEC 60601-1
and EN 60601-1; 2nd and 3.1 Editions.
Medical electrical equipment— Part 1: General requirements for basic safety and essential
performance
Meets CB Scheme
BS EN 60601-1-2:2015, IEC 60601-1-2:2014
Medical electrical equipment—Part 1–2: General requirements for basic safety and essential
performance—Collateral standard: Electromagnetic compatibility—Requirements and test
IEC/EN 62304:2006 +A1: 2015 Medical device software—Software life cycle processes
IEC/EN 62366-1:2015 (IEC 60601-1-6:2010+A1: 2013) Medical devices—Application of usability
engineering to medical devices
ISO 14971:2012 Medical devices—Application of risk management to medical devices
ISO 80601-2-56:2009 (EN 80601-2-56: 2012) Medical electrical equipment—Part 2–56: Particular
requirements for basic safety and essential performance of clinical thermometers for body
temperature measurement
ISO 10993-1:2009 Biological Evaluation of Medical Devices Part 1: Evaluation and Testing
(includes US FDA Blue book memo G95-1-100 Title)
GBT 21417.1:2008
MEDICAL ELECTRICAL EQUIPMENT needs special precautions regarding EMC. For detailed
description of EMC requirements please contact an authorized local Service Centre.
Portable and mobile RF communications equipment can affect MEDICAL ELECTRICAL
EQUIPMENT
Internally powered ME equipment.
Continuous operation.
Not protected against ingress of water.
IPX0
English
32
Symbol definitions:
Type BF Applied Part Measure icon
Caution
The caution statements in this
manual identify conditions or
practices that could result in
damage to the equipment or
other property, or loss of data.
Timer icon
Warning
The warning statements
in this manual identify
conditions or practices that
could lead to illness, injury,
or death. Warning symbols
will appear with a grey
background in a black and
white document.
Calibration date
Separate collection of
Electrical and Electronic
Equipment. Do not dispose as
unsorted municipal waste.
Consult directions for use
(DFU). A copy of the DFU is
available on this website. A
printed copy of the DFU can
be ordered from Welch Allyn
for delivery within 7 calendar
days.
-
25
-
13
55
131
Storage temperature
Storage humidity
50°C
122°F
–20°C
–4°F
85
0
English
16. Warranty
For Model Braun Thermoscan
®
PRO 6000 Ear thermometer
Welch Allyn (an affiliate of Hill-Rom®) warrants the product to be free of defects in material
and workmanship and to perform in accordance with manufacturers specifications for the
period of three years from the date of purchase from Welch Allyn or its authorized distributors
or agents.
The date of purchase is: 1) the invoiced ship date if the device was purchased directly from
Welch Allyn, 2) the date specified during product registration, or 3) the date of purchase of
the product from a Welch Allyn authorized distributor as documented from a receipt from said
distributor, whichever date is earlier.
This warranty does not cover damage caused by 1) handling during shipping, 2) use or
maintenance contrary to labeled instructions, 3) alteration or repair by anyone not authorized
by Welch Allyn, or 4) accidents. This warranty does not cover batteries, damage to the probe
window, or damage to the instrument caused by misuse, negligence or accident, and extends
to only to the first purchaser of the product. Replaced units under warranty will have the
remainder of the replaced unit’s warranty length. Additionally, this warranty becomes void if
the thermometer is operated with anything other than genuine Hillrom™ probe covers.
English
34
English
35
Guidance and manufacturer’s declaration – RF wireless communication equipment immunity
English
36
Guidance and manufacturer’s declaration – RF wireless communication equipment immunity